Ultrasensitive determination of mercury by ICP-OES coupled with a vapor generation approach based on solution cathode glow discharge

被引:27
|
作者
Cai, Zhaoqing [1 ,2 ]
Zou, Huijun [1 ]
Chen, Yirui [1 ]
Wang, Zheng [1 ,2 ]
机构
[1] Chinese Acad Sci, Shanghai Inst Ceram, Shanghai 200050, Peoples R China
[2] Univ Chinese Acad Sci, Ctr Mat Sci & Optoelect Engn, Beijing 100049, Peoples R China
基金
中国国家自然科学基金;
关键词
Solution cathode glow discharge; Vapor generation; ICP-OES; Mercury; Microplasma; ATOMIC-EMISSION-SPECTROMETRY; PERFORMANCE LIQUID-CHROMATOGRAPHY; PLASMA INDUCED VAPORIZATION; FLUORESCENCE SPECTROMETRY; SPECIATION; CADMIUM; WATER; HG; ENHANCEMENT; ELEMENTS;
D O I
10.1016/j.cclet.2021.09.107
中图分类号
O6 [化学];
学科分类号
0703 ;
摘要
We herein proposed a sample introduction technique based on solution cathode glow discharge (SCGD) of a portable design for inductively coupled plasma-optical emission spectrometry (ICP-OES) and its application in sensitive determination of mercury. The products from SCGD containing mercury vapor, were transported by an Ar flow to ICP spectrometer for detection. A gas liquid separator (GLS) and a dryer were used to condense and remove most of the accompanying moisture, which greatly improved both the stability and sensitivity of the signal. The detection limit (DL) acquired by this developed method was 0.22 mu g/L (194.1 nm), which was nearly 82 times lower than that obtained by pneumatic nebulization (PN). The relative standard deviation (RSD) was 1.4% (n = 14) for a 50 mu g/L standard. Blank solution (HNO3, pH 1) can effectively elute mercury residue. Its accuracy and practicality were also demonstrated by the determination of GBW10029 (fish) certified reference material, shrimp, crawfish, soil and human hair samples. The results showed good consistency with the certified values and the values obtained using inductively coupled plasma- mass spectrometry. (C) 2021 Published by Elsevier B.V. on behalf of Chinese Chemical Society and Institute of Materia Medica, Chinese Academy of Medical Sciences.
引用
收藏
页码:2692 / 2696
页数:5
相关论文
共 50 条
  • [21] Solution cathode glow discharge induced vapor generation of mercury and its application to mercury speciation by high performance liquid chromatography-atomic fluorescence spectrometry
    He, Qian
    Zhu, Zhenli
    Hu, Shenghong
    Jin, Lanlan
    JOURNAL OF CHROMATOGRAPHY A, 2011, 1218 (28) : 4462 - 4467
  • [22] Development of Environmental Analysis for Determination of Total Mercury in Fish Oil Pearls by Microwave Closed Vessels Digestion Coupled with ICP-OES
    Shabestari, Aida Badamchi
    Adergani, Behrouz Akbari
    Shekarchi, Maryam
    Mostafavi, Seyed Mojtaba
    EKOLOJI, 2018, 27 (106): : 1935 - 1943
  • [23] Ultratrace Pb determination in seawater by solution-cathode glow discharge-atomic emission spectrometry coupled with hydride generation
    Cheng, Junqi
    Li, Qing
    Zhao, Mingyue
    Wang, Zheng
    ANALYTICA CHIMICA ACTA, 2019, 1077 : 107 - 115
  • [24] Ultra-sensitive determination of antimony valence by solution cathode glow discharge optical emission spectrometry coupled with hydride generation
    Zhao, Mingyue
    Peng, Xiaoxu
    Yang, Bingcheng
    Wang, Zheng
    JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2020, 35 (06) : 1148 - 1155
  • [25] Determination of cadmium content in pyrotechnics used for fireworks and firecrackers based on inductively coupled plasma optical emission spectrometric approach (ICP-OES)
    Wu Jun-yi
    PROCEEDINGS OF THE 2017 5TH INTERNATIONAL CONFERENCE ON MACHINERY, MATERIALS AND COMPUTING TECHNOLOGY (ICMMCT 2017), 2017, 126 : 84 - 89
  • [26] Online mercury determination by laser-induced breakdown spectroscopy with the assistance of solution cathode glow discharge
    Zheng, Peichao
    Liu, Hongdi
    Wang, Jinmei
    Shi, Minjie
    Wang, Xiaomeng
    Zhang, Bin
    Yang, Rui
    JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2015, 30 (04) : 867 - 874
  • [27] Determination of lead content in pyrotechnics used for fireworks and firecrackers based on inductively coupled plasma optical emission spectrometric approach (ICP-OES)
    Wu Jun-yi
    PROCEEDINGS OF THE 2017 5TH INTERNATIONAL CONFERENCE ON FRONTIERS OF MANUFACTURING SCIENCE AND MEASURING TECHNOLOGY (FMSMT 2017), 2017, 130 : 374 - 379
  • [28] Online mercury determination by laser-induced breakdown spectroscopy with the assistance of solution cathode glow discharge
    Zheng, Peichao (zhengpc@cqupt.edu.cn), 1600, Royal Society of Chemistry (30):
  • [29] Determination of arsenic content in pyrotechnics used for fireworks and firecrackers based on inductively coupled plasma optical emission spectrometric approach (ICP-OES)
    Wu Jun-yi
    PROCEEDINGS OF THE 2017 5TH INTERNATIONAL CONFERENCE ON FRONTIERS OF MANUFACTURING SCIENCE AND MEASURING TECHNOLOGY (FMSMT 2017), 2017, 130 : 368 - 373
  • [30] Determination of chromium content in pyrotechnics used for fireworks and firecrackers based on inductively coupled plasma optical emission spectrometric approach (ICP-OES)
    Wu Jun-yi
    PROCEEDINGS OF THE 2017 5TH INTERNATIONAL CONFERENCE ON FRONTIERS OF MANUFACTURING SCIENCE AND MEASURING TECHNOLOGY (FMSMT 2017), 2017, 130 : 1206 - 1211