Intra- and intervariability in beam data commissioning among water phantom scanning systems

被引:18
作者
Akino, Yuichi [1 ]
Gibbons, John P. [2 ]
Neck, Daniel W. [2 ]
Chu, Connel [2 ]
Das, Indra J. [1 ]
机构
[1] Indiana Univ Sch Med, Dept Radiat Oncol, Indianapolis, IN 46202 USA
[2] Mary Bird Perkins Canc Ctr, Dept Med Phys, Baton Rouge, LA USA
关键词
quality assurance; scanning water phantom; beam data commissioning; IONIZATION CHAMBERS; DOSIMETRY; ELECTRON; AAPM;
D O I
10.1120/jacmp.v15i4.4850
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Accurate beam data acquisition during commissioning is essential for modeling the treatment planning system and dose calculation in radiotherapy. Although currently several commercial scanning systems are available, there is no report that compared the differences among the systems because most institutions do not acquire several scanning systems due to the high cost, storage space, and infrequent usage. In this report, we demonstrate the intra- and intervariability of beam profiles measured with four commercial scanning systems. During a recent educational and training workshop, four different vendors of beam scanning water phantoms were invited to demonstrate the operation and data collection of their systems. Systems were set up utilizing vendor-recommended protocols and were operated with a senior physicist, who was assigned as an instructor along with vendor. During the training sessions, each group was asked to measure beam parameters, and the intravariability in percent depth dose (PDD). At the end of the day, the profile of one linear accelerator was measured with each system to evaluate intervariability. Relatively very small (SD < 0.12%) intervariability in PDD was observed among four systems at a region deeper than peak (1.5 cm). All systems showed almost identical profiles. At the area within 80% of radiation field, the average, and maximum differences were within +/- 0.35% and 0.80%, respectively, compared to arbitrarily chosen IBA system as reference. In the penumbrae region, the distance to agreement (DTA) of the region where dose difference exceed +/- 1% was less than 1 mm. Repeated PDD measurement showed small intravariability with SD < 0.5%, although large SD was observed in the buildup region. All four water phantom scanning systems demonstrated adequate accuracy for beam data collection (i.e., within 1% of dose difference or 1 mm of DTA among each other). It is concluded that every system is capable of acquiring accurate beam. Thus the selection of a water scanning system should be based on institutional comfort, personal preference of software and hardware, and financial consideration.
引用
收藏
页码:251 / 258
页数:8
相关论文
共 16 条
[1]   POLARITY EFFECT FOR VARIOUS IONIZATION CHAMBERS WITH MULTIPLE IRRADIATION CONDITIONS IN ELECTRON-BEAMS [J].
AGET, H ;
ROSENWALD, JC .
MEDICAL PHYSICS, 1991, 18 (01) :67-72
[2]  
[Anonymous], 2004, IAEA TECHN REP SER, V430
[3]   COMMISSIONING OF VOLUMETRIC MODULATED ARC THERAPY (VMAT) [J].
Bedford, James L. ;
Warrington, Alan P. .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2009, 73 (02) :537-545
[4]  
Born E, 1997, SSRPM RECOMMENDATION
[5]  
Bruinvis IAD, 2005, 15 NCS NETH COMM RAD
[6]   Accelerator beam data commissioning equipment and procedures: Report of the TG-106 of the Therapy Physics Committee of the AAPM [J].
Das, Indra J. ;
Cheng, Chee-Wai ;
Watts, Ronald J. ;
Ahnesjo, Anders ;
Gibbons, John ;
Li, X. Allen ;
Lowenstein, Jessica ;
Mitra, Raj K. ;
Simon, William E. ;
Zhu, Timothy C. .
MEDICAL PHYSICS, 2008, 35 (09) :4186-4215
[7]   Small fields:: Nonequilibrium radiation dosimetry [J].
Das, Indra J. ;
Ding, George X. ;
Ahnesjo, Anders .
MEDICAL PHYSICS, 2008, 35 (01) :206-215
[8]   Commissioning stereotactic radiosurgery beams using both experimental and theoretical methods [J].
Ding, GX ;
Duggan, DM ;
Coffey, CW .
PHYSICS IN MEDICINE AND BIOLOGY, 2006, 51 (10) :2549-2566
[9]   IMRT commissioning: Multiple institution planning and dosimetry comparisons, a report from AAPM Task Group 119 [J].
Ezzell, Gary A. ;
Burmeister, Jay W. ;
Dogan, Nesrin ;
LoSasso, Thomas J. ;
Mechalakos, James G. ;
Mihailidis, Dimitris ;
Molineu, Andrea ;
Palta, Jatinder R. ;
Ramsey, Chester R. ;
Salter, Bill J. ;
Shi, Jie ;
Xia, Ping ;
Yue, Ning J. ;
Xiao, Ying .
MEDICAL PHYSICS, 2009, 36 (11) :5359-5373
[10]   Commissioning of the Varian TrueBeam linear accelerator: A multi-institutional study [J].
Glide-Hurst, C. ;
Bellon, M. ;
Foster, R. ;
Altunbas, C. ;
Speiser, M. ;
Altman, M. ;
Westerly, D. ;
Wen, N. ;
Zhao, B. ;
Miften, M. ;
Chetty, I. J. ;
Solberg, T. .
MEDICAL PHYSICS, 2013, 40 (03)