Ballistocardiogram as Proximal Timing Reference for Pulse Transit Time Measurement: Potential for Cuffless Blood Pressure Monitoring

被引:104
|
作者
Kim, Chang-Sei [1 ]
Carek, Andrew M. [2 ]
Mukkamala, Ramakrishna [3 ]
Inan, Omer T. [2 ]
Hahn, Jin-Oh [1 ]
机构
[1] Univ Maryland, Dept Mech Engn, College Pk, MD 20742 USA
[2] Georgia Inst Technol, Sch Elect & Comp Engn, Atlanta, GA 30332 USA
[3] Michigan State Univ, Dept Elect & Comp Engn, E Lansing, MI 48824 USA
基金
美国国家卫生研究院;
关键词
Ballistocardiogram (BCG); blood pressure; electrocardiogram (ECG); pulse arrival time (PAT); pulse transit time (PTT); ARRIVAL-TIME; PREEJECTION PERIOD; HEART-RATE; CONTRACTILITY; CALIBRATION; MARKER; ECG;
D O I
10.1109/TBME.2015.2440291
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Goal: We tested the hypothesis that the ballistocardiogram (BCG) waveform could yield a viable proximal timing reference for measuring pulse transit time (PTT). Methods: From 15 healthy volunteers, we measured PTT as the time interval between BCG and a noninvasively measured finger blood pressure (BP) waveform. To evaluate the efficacy of the BCG-based PTT in estimating BP, we likewise measured pulse arrival time (PAT) using the electrocardiogram (ECG) as proximal timing reference and compared their correlations to BP. Results: BCG-based PTT was correlated with BP reasonably well: the mean correlation coefficient (r) was 0.62 for diastolic (DP), 0.65 for mean (MP), and 0.66 for systolic (SP) pressures when the intersecting tangent method was used as distal timing reference. Comparing four distal timing references (intersecting tangent, maximum second derivative, diastolic minimum, and systolic maximum), PTT exhibited the best correlation with BP when the systolic maximum method was used (mean r value was 0.66 for DP, 0.67 for MP, and 0.70 for SP). PTT was more strongly correlated with DP than PAT regardless of the distal timing reference: mean r value was 0.62 versus 0.51 (p = 0.07) for intersecting tangent, 0.54 versus 0.49 (p = 0.17) for maximum second derivative, 0.58 versus 0.52 (p = 0.37) for diastolic minimum, and 0.66 versus 0.60 (p = 0.10) for systolic maximum methods. The difference between PTT and PAT in estimating DP was significant (p = 0.01) when the r values associated with all the distal timing references were compared altogether. However, PAT appeared to outperform PTT in estimating SP (p = 0.31 when the r values associated with all the distal timing references were compared altogether). Conclusion: We conclude that BCG is an adequate proximal timing reference in deriving PTT, and that BCG-based PTT may be superior to ECG-based PAT in estimating DP. Significance: PTT with BCG as proximal timing reference has potential to enable convenient and ubiquitous cuffless BP monitoring.
引用
收藏
页码:2657 / 2664
页数:8
相关论文
共 50 条
  • [31] Restrictive factors of the measurement accuracy to estimate Blood Pressure with Pulse wave Transit Time
    Chen, X. M.
    Li, Y. B.
    Zhang, Y.
    Deng, N.
    BIOINFORMATICS AND BIOMEDICAL ENGINEERING: NEW ADVANCES, 2016, : 435 - 441
  • [32] A novel calibration method for noninvasive blood pressure measurement using pulse transit time
    Yan, Y. S.
    Zhang, Y. T.
    2007 4TH IEEE/EMBS INTERNATIONAL SUMMER SCHOOL AND SYMPOSIUM ON MEDICAL DEVICES AND BIOSENSORS, 2007, : 1 - +
  • [33] Novel blood pressure and pulse pressure estimation based on pulse transit time and stroke volume approximation
    Joonnyong Lee
    JangJay Sohn
    Jonghyun Park
    SeungMan Yang
    Saram Lee
    Hee Chan Kim
    BioMedical Engineering OnLine, 17
  • [34] An implantable optical blood pressure sensor based on pulse transit time
    Fiala, Jens
    Bingger, Philipp
    Ruh, Dominic
    Foerster, Katharina
    Heilmann, Claudia
    Beyersdorf, Friedhelm
    Zappe, Hans
    Seifert, Andreas
    BIOMEDICAL MICRODEVICES, 2013, 15 (01) : 73 - 81
  • [35] Estimation of Cuffless Systolic and Diastolic Blood Pressure Using Pulse Transient Time
    Mehrabbeik, Mahtab
    Rashidi, Saeid
    2019 5TH IRANIAN CONFERENCE ON SIGNAL PROCESSING AND INTELLIGENT SYSTEMS (ICSPIS 2019), 2019,
  • [36] Wearable Continuous Blood Pressure Monitoring Devices Based on Pulse Wave Transit Time and Pulse Arrival Time: A Review
    Zhou, Zi-Bo
    Cui, Tian-Rui
    Li, Ding
    Jian, Jin-Ming
    Li, Zhen
    Ji, Shou-Rui
    Li, Xin
    Xu, Jian-Dong
    Liu, Hou-Fang
    Yang, Yi
    Ren, Tian-Ling
    MATERIALS, 2023, 16 (06)
  • [37] The Wavelet Transform of Pulse Wave and Electrocardiogram Improves Accuracy of Blood Pressure Estimation in Cuffless Blood Pressure Measurement
    Yamanaka, Syunsuke
    Morikawa, Koji
    Yamamura, Osamu
    Morita, Hiroshi
    Huh, Ji Young
    CIRCULATION, 2016, 134
  • [38] Continuous blood pressure measurement using the pulse transit time: Comparison to intra-arterial measurement
    Patzak, Andreas
    Mendoza, Yuri
    Gesche, Heiko
    Konermann, Martin
    BLOOD PRESSURE, 2015, 24 (04) : 217 - 221
  • [39] Continuous monitoring of cerebrovascular reactivity through pulse transit time and intracranial pressure
    Liu, Xiuyun
    Gadhoumi, Kais
    Xiao, Ran
    Tran, Nate
    Smielewski, Peter
    Czosnyka, Marek
    Hetts, Steven W.
    Ko, Nerissa
    Hu, Xiao
    PHYSIOLOGICAL MEASUREMENT, 2019, 40 (01)
  • [40] Pulse arrival time is not an adequate surrogate for pulse transit time as a marker of blood pressure
    Zhang, Guanqun
    Gao, Mingwu
    Xu, Da
    Olivier, N. Bari
    Mukkamala, Ramakrishna
    JOURNAL OF APPLIED PHYSIOLOGY, 2011, 111 (06) : 1681 - 1686