Performance of a HeartLogic™ Based Care Path in the Management of a Real-World Chronic Heart Failure Population

被引:13
作者
Feijen, Michelle [1 ]
Egorova, Anastasia D. [1 ]
Treskes, Roderick W. [1 ]
Mertens, Bart J. A. [2 ]
Jukema, J. Wouter [1 ]
Schalij, Martin J. [1 ]
Beeres, Saskia L. M. A. [1 ]
机构
[1] Leiden Univ Med Ctr, Dept Cardiol, Leiden, Netherlands
[2] Leiden Univ Med Ctr, Dept Biomed DataSciences, Leiden, Netherlands
关键词
HeartLogic (TM); chronic heart failure; multisensory remote monitoring; heart failure admissions; CIED; HOSPITALIZATIONS; MORTALITY; EVENTS;
D O I
10.3389/fcvm.2022.883873
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Aim: Early detection of impending fluid retention and timely adjustment of (medical) therapy can prevent heart failure related hospitalizations. The multisensory cardiac implantable electronic device (CIED) based algorithm HeartLogic (TM) aims to alert in case of impending fluid retention. The aim of the current analysis is to evaluate the performance of the HeartLogic (TM) guided heart failure care path in a real-world heart failure population and to investigate whether the height of the index and the duration of the alert state are indicative of the degree of fluid retention. MethodsConsecutive adult heart failure patients with a CIED and an activated HeartLogic (TM) algorithm were eligible for inclusion. Patients were followed up according to the hospital's heart failure care path. The device technician reviewed alerts for a technical CIED checkup. Afterwards, the heart failure nurse contacted the patient to identify impending fluid retention. An alert was either true positive or false positive. Without an alert a patient was true negative or false negative. ResultsAmong 107 patients, [82 male, 70 (IQR 60-77) years, left ventricular ejection fraction 37 +/- 11%] 130 HeartLogic (TM) alerts were available for analysis. Median follow up was 14 months [IQR 8-23]. The sensitivity to detect impending fluid retention was 79%, the specificity 88%. The positive predictive was value 71% and the negative predictive value 91%. The unexplained alert rate was 0.23 alerts/patient year and the false negative rate 0.17 alerts/patient year. True positive alerts [42 days (IQR 28-63)] lasted longer than false positive alerts [28 days (IQR 21-44)], p = 0.02. The maximal HeartLogic (TM) index was higher in true positive alerts [26 (IQR 21-34)] compared to false positive alerts [19 (IQR 17-24)], p < 0.01. Patients with higher HeartLogic (TM) indexes required more intense treatment (index height in outpatient setting 25 [IQR 20-32], day clinic treatment 28 [IQR 24-36] and hospitalized patients 45 [IQR 35-58], respectively), p < 0.01. ConclusionThe CIED-based HeartLogic (TM) algorithm facilitates early detection of impending fluid retention and thereby enables clinical action to prevent this at early stage. The current analysis illustrates that higher and persistent alerts are indicative for true positive alerts and higher index values are indicative for more severe fluid retention.
引用
收藏
页数:9
相关论文
共 22 条
[1]  
Adamson Philip B, 2009, Curr Heart Fail Rep, V6, P287
[2]   Incident heart failure hospitalization and subsequent mortality in chronic heart failure: A propensity-matched study [J].
Ahmed, Ali ;
Allman, Richard M. ;
Fonarow, Gregg C. ;
Love, Thomas E. ;
Zannad, Faiez ;
Dell'Italia, Louis J. ;
White, Michel ;
Gheorghiade, Mihai .
JOURNAL OF CARDIAC FAILURE, 2008, 14 (03) :211-218
[3]   CorVue algorithm efficacy to predict heart failure in real life: Unnecessary and potentially misleading information? [J].
Anna Palfy, Julia ;
Benezet-Mazuecos, Juan ;
Martinez Milla, Juan ;
Antonio Iglesias, Jose ;
de la Vieja, Juan Jose ;
Sanchez-Borque, Pepa ;
Miracle, Angel ;
Manuel Rubio, Jose .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2018, 41 (08) :948-952
[4]   A Multisensor Algorithm Predicts Heart Failure Events in Patients With Implanted Devices Results From the MultiSENSE Study [J].
Boehmer, John P. ;
Hariharan, Ramesh ;
Devecchi, Fausto G. ;
Smith, Andrew L. ;
Molon, Giulio ;
Capucci, Alessandro ;
An, Qi ;
Averina, Viktoria ;
Stolen, Craig M. ;
Thakur, Pramodsingh H. ;
Thompson, Julie A. ;
Wariar, Ramesh ;
Zhang, Yi ;
Singh, Jagmeet P. .
JACC-HEART FAILURE, 2017, 5 (03) :216-225
[5]   Multiparametric Implantable Cardioverter-Defibrillator Algorithm for Heart Failure Risk Stratification and Management An Analysis in Clinical Practice [J].
Calo, Leonardo ;
Bianchi, Valter ;
Ferraioli, Donatella ;
Santini, Luca ;
Dello Russo, Antonio ;
Carriere, Cosimo ;
Santobuono, Vincenzo Ezio ;
Andreoli, Chiara ;
La Greca, Carmelo ;
Arena, Giuseppe ;
Talarico, Antonello ;
Pisano, Ennio ;
Santoro, Amato ;
Giammaria, Massimo ;
Ziacchi, Matteo ;
Viscusi, Miguel ;
De Ruvo, Ermenegildo ;
Campari, Monica ;
Valsecchi, Sergio ;
D'Onofrio, Antonio .
CIRCULATION-HEART FAILURE, 2021, 14 (10) :1072-1079
[6]   ICD-measured heart sounds and their correlation with echocardiographic indexes of systolic and diastolic function [J].
Calo, Leonardo ;
Capucci, Alessandro ;
Santini, Luca ;
Pecora, Domenico ;
Favale, Stefano ;
Petracci, Barbara ;
Molon, Giulio ;
Bianchi, Valter ;
Cipolletta, Laura ;
De Ruvo, Ermenegildo ;
Ammirati, Fabrizio ;
La Greca, Carmelo ;
Campari, Monica ;
Valsecchi, Sergio ;
D'Onofrio, Antonio .
JOURNAL OF INTERVENTIONAL CARDIAC ELECTROPHYSIOLOGY, 2020, 58 (01) :95-101
[7]   Preliminary experience with the multisensor HeartLogic algorithm for heart failure monitoring: a retrospective case series report [J].
Capucci, Alessandro ;
Santini, Luca ;
Favale, Stefano ;
Pecora, Domenico ;
Petracci, Barbara ;
Calo, Leonardo ;
Molon, Giulio ;
Cipolletta, Laura ;
Bianchi, Valter ;
Schirripa, Valentina ;
Santobuono, Vincenzo E. ;
La Greca, Carmelo ;
Campari, Monica ;
Valsecchi, Sergio ;
Ammirati, Fabrizio ;
D'Onofrio, Antonio .
ESC HEART FAILURE, 2019, 6 (02) :308-318
[8]   Sensitivity and positive predictive value of implantable intrathoracic impedance monitoring as a predictor of heart failure hospitalizations: the SENSE-HF trial [J].
Conraads, Viviane M. ;
Tavazzi, Luigi ;
Santini, Massimo ;
Oliva, Fabrizio ;
Gerritse, Bart ;
Yu, Cheuk-Man ;
Cowie, Martin R. .
EUROPEAN HEART JOURNAL, 2011, 32 (18) :2266-2273
[9]   Early Detection of Fluid Retention in Patients with Advanced Heart Failure: A Review of a Novel Multisensory Algorithm, HeartLogicTM [J].
Feijen, Michelle ;
Egorova, Anastasia D. ;
Beeres, Saskia L. M. A. ;
Treskes, Roderick W. .
SENSORS, 2021, 21 (04) :1-19
[10]   Multiparameter diagnostic sensor measurements during clinically stable periods and worsening heart failure in ambulatory patients [J].
Gardner, Roy S. ;
Thakur, Pramodsingh ;
Hammill, Eric F. ;
Nair, Devi G. ;
Eldadah, Zayd ;
Stancak, Branislav ;
Ferrick, Kevin ;
Sriratanasathavorn, Charn ;
Duray, Gabor Zoltan ;
Wariar, Ramesh ;
Zhang, Yi ;
An, Qi ;
Averina, Viktoria ;
Boehmer, John P. .
ESC HEART FAILURE, 2021, 8 (02) :1571-1581