Modulation of the foreign body response to implanted sensor models through device-based delivery of the tyrosine kinase inhibitor, masitinib

被引:37
作者
Avula, Mahender Nath [1 ]
Rao, Archana Nagaraja [2 ]
McGill, Lawrence D. [3 ,4 ]
Grainger, David William [1 ,2 ]
Solzbacher, Florian [1 ,5 ]
机构
[1] Univ Utah, Dept Bioengn, Salt Lake City, UT 84112 USA
[2] Univ Utah, Dept Pharmaceut & Pharmaceut Chem, Salt Lake City, UT 84112 USA
[3] Univ Utah, Salt Lake City, UT 84112 USA
[4] Univ Utah, Univ Pathologist Labs, Salt Lake City, UT 84112 USA
[5] Univ Utah, Dept Elect & Comp Engn, Salt Lake City, UT 84112 USA
关键词
Foreign body response; Continuous glucose monitoring sensors; Mast cells; Tyrosine kinase inhibitor; Local drug delivery; Biomedical implants; ACUTE INFLAMMATORY RESPONSES; ENDOTHELIAL GROWTH-FACTOR; GLUCOSE SENSORS; MAST-CELLS; IMATINIB MESYLATE; IN-VITRO; DRUG; BIOCOMPATIBILITY; MICROSPHERES; PERFORMANCE;
D O I
10.1016/j.biomaterials.2013.08.090
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The host foreign body response (FBR) adversely effects the performance of numerous implanted biomaterials especially biosensors, including clinically popular glucose-monitoring sensors. Reactive formation of a fibrous capsule around implanted sensors hinders the transport of essential analytes to the sensor from the surrounding tissue, resulting in loss of glucose response sensitivity and eventual sensor failure. Several strategies have sought to mitigate the foreign body response's effects on CGM sensors through the use of local delivery of pharmaceuticals and biomolecules with limited success. This study describes release of a tyrosine kinase inhibitor - masitinib - from the sensor implant to target tissue resident mast cells as key mediators of the FBR. Model implants are coated with a composite polymer hydrophilic matrix that rapidly dissolves upon tissue implantation to deposit slower-degrading polymer microparticles containing masitinib. Matrix dissolution limits coating interference with sensor function while establishing a local controlled-release delivery depot formulation to alter implant tissue pharmacology and addressing the FBR. Drug efficacy was evaluated in a murine subcutaneous pocket implant model. Drug release extends to more than 30 days in vitro. The resulting FBR in vivo, evaluated by implant capsule thickness and inflammatory cell densities at 14, 21, and 28 days, displays statistically significant reduction in capsule thickness around masitinib-releasing implant sites compared to control implant sites. (C) 2013 Elsevier Ltd. All rights reserved.
引用
收藏
页码:9737 / 9746
页数:10
相关论文
共 49 条
[1]   Foreign body reaction to biomaterials [J].
Anderson, James M. ;
Rodriguez, Analiz ;
Chang, David T. .
SEMINARS IN IMMUNOLOGY, 2008, 20 (02) :86-100
[2]   Biodegradation and biocompatibility of PLA and PLGA microspheres [J].
Anderson, JM ;
Shive, MS .
ADVANCED DRUG DELIVERY REVIEWS, 1997, 28 (01) :5-24
[3]  
[Anonymous], 2009, In vivo glucose sensing
[4]  
Avula M, 2013, DRUG DEVICE COMBINAT
[5]   NEOVASCULARIZATION OF SYNTHETIC MEMBRANES DIRECTED BY MEMBRANE MICROARCHITECTURE [J].
BRAUKER, JH ;
CARRBRENDEL, VE ;
MARTINSON, LA ;
CRUDELE, J ;
JOHNSTON, WD ;
JOHNSON, RC .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 1995, 29 (12) :1517-1524
[6]  
Butler K, 1999, Biomed Sci Instrum, V35, P267
[7]   Cytological evaluation of the tissue-implant reaction associated with S/C and I/P implantation of ALCAP and HA bioceramics in vivo [J].
Butler, K ;
Benghuzzi, H ;
Puckett, A .
PATHOLOGY RESEARCH AND PRACTICE, 2001, 197 (01) :29-39
[8]  
Coster S., 2000, Health Technology Assessment, V4, pi, DOI DOI 10.3310/HTA4120
[9]   In vitro pyrogen test for toxic or immunomodulatory drugs [J].
Daneshian, Mardas ;
Guenther, Armin ;
Wendel, Albrecht ;
Hartung, Thomas ;
von Aulock, Sonja .
JOURNAL OF IMMUNOLOGICAL METHODS, 2006, 313 (1-2) :169-175
[10]   Imatinib mesylate reduces production of extracellular matrix and prevents development of experimental dermal fibrosis [J].
Distler, Joerg H. W. ;
Juengel, Astrid ;
Huber, Lars C. ;
Schulze-Horsel, Ursula ;
Zwerina, Jochen ;
Gay, Renate E. ;
Michel, Beat A. ;
Hauser, Thomas ;
Schett, Georg ;
Gay, Steffen ;
Distler, Oliver .
ARTHRITIS AND RHEUMATISM, 2007, 56 (01) :311-322