The transformation of common office supplies into a low-cost optical biosensing platform

被引:23
作者
Han, Yong Duk [1 ]
Chun, Hyeong Jin [1 ]
Yoon, Hyun C. [1 ]
机构
[1] Ajou Univ, Dept Mol Sci & Technol, Suwon 443749, South Korea
关键词
Optical biosensor; Laser diode; Solar cell; Glucose monitoring; Osteoarthritis immunoassay; HYDROGEN-PEROXIDE; RESONANCE; OSTEOARTHRITIS;
D O I
10.1016/j.bios.2014.03.037
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
By reassembling common office supplies, an optical biosensing system was developed. A laser pointer and the solar cell from a calculator were utilized in the developed optical biosensing system as the light source and signal transducer, respectively. For intuitive signal evaluation, a multimeter was used. The following two types of conventional enzymatic colorimetric assays were employed with the optical biosensing system: (i) the Trinder's reaction-based enzymatic assay; and (ii) the competitive enzyme-linked immunosorbent assay. These colorimetric assays were performed in reaction channels made from transparent polymer and glass. By matching the maximum absorption spectra of the colored end products from the assays with the emission spectra of the laser diodes, the biochemical reaction rate was manifested as a change in the intensity of the laser beam. This change was then converted by the solar cell into voltage and displayed on the connected multimeter. To verify the detection performance of the system, glucose and an osteoarthritis biomarker (urinary collagen type II C-telopeptide fragments [uCTX-II]) were quantified. With glucose, the voltages registered were linearly correlated with the glucose concentration, from 0 to 10 mM. Using a competitive immunoassay for uCTX-II the system exhibited a calibration curve with a dynamic detection range between 1.3 and 10 ng/mL uCTX-II. Given the advantages of the proposed biosensing system, including its high sensitivity, facile fabrication, and the high obtainability and cost-effectiveness of the components used to make it, we expect that this study will provide a basis for the production of a low-cost optical biosensor. (C) 2014 Elsevier B.V. All rights reserved.
引用
收藏
页码:259 / 268
页数:10
相关论文
共 27 条
[1]   New trends in instrumental design for surface plasmon resonance-based biosensors [J].
Abbas, Abdennour ;
Linman, Matthew J. ;
Cheng, Quan .
BIOSENSORS & BIOELECTRONICS, 2011, 26 (05) :1815-1824
[2]  
Birmingham JD, 2006, BIOMARK INSIGHTS, V1, P61
[3]   Optical biosensors [J].
Borisov, Sergey M. ;
Wolfbeis, Otto S. .
CHEMICAL REVIEWS, 2008, 108 (02) :423-461
[4]   ELECTRODE SYSTEMS FOR CONTINUOUS MONITORING IN CARDIOVASCULAR SURGERY [J].
CLARK, LC ;
LYONS, C .
ANNALS OF THE NEW YORK ACADEMY OF SCIENCES, 1962, 102 (01) :29-&
[5]   Review: Collagen markers in early arthritic diseases [J].
Elsaid, KA ;
Chichester, CO .
CLINICA CHIMICA ACTA, 2006, 365 (1-2) :68-77
[6]   Advances in collagen cross-link analysis [J].
Eyre, David R. ;
Weis, Mary Ann ;
Wu, Jiann-Jiu .
METHODS, 2008, 45 (01) :65-74
[7]   Sensitive optical biosensors for unlabeled targets: A review [J].
Fan, Xudong ;
White, Ian M. ;
Shopova, Siyka I. ;
Zhu, Hongying ;
Suter, Jonathan D. ;
Sun, Yuze .
ANALYTICA CHIMICA ACTA, 2008, 620 (1-2) :8-26
[8]   Bone marrow abnormalities on magnetic resonance imaging are associated with type II collagen degradation in knee osteoarthritis - A three-month longitudinal study [J].
Garnero, P ;
Peterfy, C ;
Zaim, S ;
Schoenharting, M .
ARTHRITIS AND RHEUMATISM, 2005, 52 (09) :2822-2829
[9]   Point of Care Diagnostics: Status and Future [J].
Gubala, Vladimir ;
Harris, Leanne F. ;
Ricco, Antonio J. ;
Tan, Ming X. ;
Williams, David E. .
ANALYTICAL CHEMISTRY, 2012, 84 (02) :487-515
[10]   Multienzyme-modified biosensing surface for the electrochemical analysis of aspartate transaminase and alanine transaminase in human plasma [J].
Han, Yong Duk ;
Song, Seung Yeon ;
Lee, Jun Hwang ;
Lee, Dae Sik ;
Yoon, Hyun C. .
ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2011, 400 (03) :797-805