Positive Bias in Mycophenolic Acid Concentrations Determined by the CEDIA Assay Compared to HPLC-UV Method: Is CEDIA Assay Suitable for Therapeutic Drug Monitoring of Mycophenolic Acid?

被引:6
作者
Dasgupta, Amitava
Johnson, Myrtle
机构
[1] Univ Texas Houston, Sch Med, Dept Pathol, Houston, TX 77030 USA
[2] Texas Med Ctr, Mem Hermann Hosp, Lab Serv, Houston, TX USA
关键词
mycophenolic acid; CEDIA; HPLC-UV; positive bias; PERFORMANCE LIQUID-CHROMATOGRAPHY; MULTIPLIED IMMUNOASSAY TECHNIQUE; GLUCURONIDE METABOLITES; TRANSPLANT RECIPIENTS; HUMAN PLASMA; MOFETIL;
D O I
10.1002/jcla.21565
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
BackgroundBoth immunoassays and chromatographic methods are available for therapeutic drug monitoring of mycophenolic acid (MPA), an immunosuppressant. We studied the suitability of cloned enzyme donor immunoassay (CEDIA) assay for routine monitoring of MPA by comparing values obtained by the CEDIA assay with corresponding values obtained by using a high-performance liquid chromatography combined with ultraviolet detection (HPLC-UV) method. J. Clin. Lab. Anal. 27:77-80, 2013. (c) 2012 Wiley Periodicals, Inc. MethodsWe compared MPA concentrations obtained by a reference HPLC-UV method and CEDIA assay on Hitachi 917 analyzer (Roche Diagnostics, Indianapolis, IN) using 60 patient specimens (18 liver transplant recipient and 42 kidney transplant recipients). ResultsWhen MPA concentrations in all 60 transplant recipients obtained by the HPLC-UV (x-axis) method were compared with corresponding values obtained by the CEDIA method (y-axis), the following regression equation was obtained: y = 1.1558x + 0.2876 (r = 0.97). Interestingly, much lower bias was observed in 42 renal transplant recipients as revealed by the following regression equation; y = 1.1181x + 0.2745 (r = 0.98). However, more significant positive bias was observed in 18 liver transplant recipients as following regression equation as observed: y = 1.3337x + 0.1493 (r = 0.94). ConclusionsWe conclude that MPA concentrations determined by the CEDIA assay showed significant positive bias compared to HPLC-UV method. Therefore, caution must be exercised in interpreting therapeutic drug monitoring result of MPA if CEDIA assay is used.
引用
收藏
页码:77 / 80
页数:4
相关论文
共 14 条
  • [1] An Investigation Into the Bias Between Liquid Chromatography-Tandem Mass Spectrometry and an Enzyme Multiplied Immunoassay Technique for the Measurement of Mycophenolic Acid
    Brown, Nigel W.
    Franklin, Michael E.
    Einarsdottir, Eyrun N.
    Gonde, Christopher E.
    Pires, Maria
    Taylor, Paul J.
    Tredger, J. Michael
    [J]. THERAPEUTIC DRUG MONITORING, 2010, 32 (04) : 420 - 426
  • [2] Clinical pharmacokinetics of mycophenolate mofetil
    Bullingham, RES
    Nicholls, AJ
    Kanmm, BR
    [J]. CLINICAL PHARMACOKINETICS, 1998, 34 (06) : 429 - 455
  • [3] Simple high performance liquid chromatographic assay for mycophenolic acid in renal transplant patients
    Elbarbry, Fawzy A.
    Shoker, Ahmed
    [J]. JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 2007, 43 (02) : 788 - 792
  • [4] Comparison of high-performance liquid chromatography and enzyme-multiplied immunoassay technique to monitor mycophenolic acid in paediatric renal recipients
    Irtan, Sabine
    Azougagh, Said
    Monchaud, Caroline
    Popon, Michel
    Baudouin, Veronique
    Jacqz-Aigrain, Evelyne
    [J]. PEDIATRIC NEPHROLOGY, 2008, 23 (10) : 1859 - 1865
  • [5] The effect of renal insufficiency on mycophenolic acid protein binding
    Kaplan, B
    Meier-Kriesche, HU
    Friedman, G
    Mulgaonkar, S
    Gruber, S
    Korecka, M
    Brayman, KL
    Shaw, LM
    [J]. JOURNAL OF CLINICAL PHARMACOLOGY, 1999, 39 (07) : 715 - 720
  • [6] Korecka M, 1999, CLIN CHEM, V45, P1047
  • [7] High-performance liquid chromatography method for the determination of mycophenolic acid and its acyl and phenol glucuronide metabolites in human plasma
    Patel, CG
    Akhlaghi, F
    [J]. THERAPEUTIC DRUG MONITORING, 2006, 28 (01) : 116 - 122
  • [8] Shipkova M, 2001, CLIN CHEM, V47, P1485
  • [9] Investigation of the Crossreactivity of Mycophenolic Acid Glucuronide Metabolites and of Mycophenolate Mofetil in the Cedia MPA Assay
    Shipkova, Maria
    Schuetz, Ekkehard
    Besenthal, Ingo
    Fraunberger, Peter
    Wieland, Eberhard
    [J]. THERAPEUTIC DRUG MONITORING, 2010, 32 (01) : 79 - 85
  • [10] Monitoring mycophenolate in liver transplant recipients: Toward a therapeutic range
    Tredger, JM
    Brown, NW
    Adams, J
    Gonde, CE
    Dhawan, A
    Rela, M