Assessment of vertebral artery stents using 16-slice multi-detector row CT angiography in vivo evaluation: Comparison of a medium-smooth kernel and a sharp kernel

被引:2
作者
Yoo, Won Jong [1 ]
Lim, Yeon Soo [1 ]
Ahn, Kook Jin [2 ]
Choi, Byung Gil [3 ]
Kim, Ji Young [1 ]
Kim, Sung Hoon [3 ]
机构
[1] Catholic Univ Korea, Coll Med, Holy Family Hosp, Dept Radiol, Puchon 420717, Gyeonggi, South Korea
[2] Catholic Univ Korea, Coll Med, St Marry Hosp, Dept Radiol, Seoul 150713, South Korea
[3] Catholic Univ Korea, Coll Med, Kangnam St Marry Hosp, Dept Radiol, Seoul 137701, South Korea
关键词
Vertebral artery; Stent; Multi-detector row CT; CT angiography; MAGNETIC-RESONANCE ANGIOGRAPHY; MULTISLICE COMPUTED-TOMOGRAPHY; INITIAL-EXPERIENCE; VITRO EVALUATION; CORONARY; PATENCY; RESTENOSIS; ISCHEMIA; PHANTOM;
D O I
10.1016/j.ejrad.2008.01.039
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Objectives: To assess the lumen visibility of extracranial vertebral artery stents examined with 16-slice multi-detector row computed tomography (MDCT) angiography in vivo using a medium-smooth kernel (B30s) and a sharp kernel (B60s), and to compare these with digital subtraction angiography (DSA) after stent placement. Methods: Twenty stents from 20 patients (14 men, 6 women; mean age, 62.7 +/- 10.1 years) who underwent CT angiography (CTA) with 16-slice MDCT were retrospectively analyzed. In CT angiograms using a B30s and a B60s, the lumen diameters and CT attenuations of the stented vessels were measured three times by three observers, and artificial luminal narrowing (ALN) was calculated. To assess measurement reliability on CT angiograms, the intraclass correlation coefficient (ICC) was used. DSA served as the reference standard for the in-stent luminal measurements on CT angiography. The median interval between CT angiography and DSA was 1 day (range 1-10). Results: For interobserver reliability, intraclass correlation coefficients for the lumen diameters on CT angiograms with a B30s and a B60s were 0.90 and 0.96, respectively. The lumen diameters on CT angiograms using a B30s were consistently smaller than that on CT angiograms using a B60s (p < 0.01). The mean ALN was 37 +/- 7% on CT angiograms using a B30s and 25 +/- 9% on CT angiograms using a B60s. The mean CT attenuation in in-stent lumen was 347 +/- 55 HU on CT angiograms using a B30s and 295 +/- 46 HU on CT angiograms using a B60s. The ALN and CT attenuation within the stented vessels between CT angiograms using a B30s and a B60s was significant (p < 0.01). Conclusions: 16-slice MDCT using a sharp kernel allows good visualization of the stented vessels and is useful in the assessment of vertebral artery stent patency after stent placement. (C) 2008 Elsevier Ireland Ltd. All rights reserved.
引用
收藏
页码:362 / 368
页数:7
相关论文
共 20 条
[1]   Endovascular treatment of the vertebral artery origin in patients with symptoms of vertebrobasilar ischemia [J].
Dabus, Guilherme ;
Gerstle, Ronald J. ;
Derdeyn, Colin P. ;
Cross, DeWitte T., III ;
Moran, Christopher J. .
NEURORADIOLOGY, 2006, 48 (12) :917-923
[2]   Coronary artery stent patency assessed with in-stent contrast enhancement measured at multi-detector row CT angiography: Initial experience [J].
Hong, C ;
Chrysant, GS ;
Woodard, PK ;
Bae, KT .
RADIOLOGY, 2004, 233 (01) :286-291
[3]  
*KOR NAT STAT OFF, 2005, ANN REP CAUS DEATH S
[4]   Multislice spiral computed tomography for the detection of coronary stent restenosis and patency [J].
Krüger, S ;
Mahnken, AH ;
Sinha, AM ;
Borghans, A ;
Dedden, K ;
Hoffmann, R ;
Hanrath, P .
INTERNATIONAL JOURNAL OF CARDIOLOGY, 2003, 89 (2-3) :167-172
[5]   Symptomatic ostial vertebral artery stenosis treated with tubular coronary stents: Clinical results and restenosis analysis [J].
Lin, YH ;
Juang, JM ;
Jeng, JS ;
Yip, PK ;
Kao, HL .
JOURNAL OF ENDOVASCULAR THERAPY, 2004, 11 (06) :719-726
[6]  
Mahnken AH, 2005, INVEST RADIOL, V40, P8
[7]   Coronary artery stents in multislice computed tomography -: In vitro artifact evaluation [J].
Mahnken, AH ;
Buecker, A ;
Wildberger, JE ;
Ruebben, A ;
Stanzel, S ;
Vogt, F ;
Günther, RW ;
Blindt, R .
INVESTIGATIVE RADIOLOGY, 2004, 39 (01) :27-33
[8]  
Maintz D, 2003, INVEST RADIOL, V38, P790, DOI 10.1097/01.rli.0000086495.96457.54
[9]   64-slice multidetector coronary CT angiography:: in vitro evaluation of 68 different stents [J].
Maintz, D ;
Seifarth, H ;
Raupach, R ;
Flohr, T ;
Rink, M ;
Sommer, T ;
Özgün, M ;
Heindel, W ;
Fischbach, R .
EUROPEAN RADIOLOGY, 2006, 16 (04) :818-826
[10]   Imaging of coronary artery stents using multislice computed tomography: in vitro evaluation [J].
Maintz, D ;
Juergens, KU ;
Wichter, T ;
Grude, M ;
Heindel, W ;
Fischbach, R .
EUROPEAN RADIOLOGY, 2003, 13 (04) :830-835