Filling accuracy and imprecision of commercial evacuated sodium citrate coagulation tubes

被引:7
|
作者
Lippi, Giuseppe [1 ]
Salvagno, Gian Luca [1 ]
Biljak, Vanja Radisic [2 ]
Kralj, Ana-Katarina [2 ]
Kuktic, Ivona [2 ]
Gelati, Matteo [1 ]
Simundic, Ana-Maria [2 ]
机构
[1] Univ Verona, Sect Clin Biochem, Verona, Italy
[2] Univ Hosp Sveti Duh, Dept Med Lab Diagnost, Zagreb, Croatia
来源
SCANDINAVIAN JOURNAL OF CLINICAL & LABORATORY INVESTIGATION | 2019年 / 79卷 / 04期
关键词
Quality; errors; coagulation; blood collection; blood tubes; PREANALYTICAL PHASE; COLLECTION; QUALITY; ROUTINE; HEMOSTASIS;
D O I
10.1080/00365513.2019.1609696
中图分类号
R-3 [医学研究方法]; R3 [基础医学];
学科分类号
1001 ;
摘要
Current recommendations advocate that blood tubes for coagulation testing should be filled not less than 90% of their nominal filling volume, since under- or over-filling >10% may generate unreliable results of some hemostasis assays. This study was hence aimed to explore filling accuracy and precision of commercial blood tubes. Between-lot variations of 3 different lots (20 tubes per lot) of 3.2% citrate blood tubes manufactured by Becton Dickinson, Greiner and Kima were studied. One additional lot from each manufacturer was assessed in triplicate (three series of 20 tubes), to assess within-lot variation. All tubes were first weighed empty and then filled with distilled water by a syringe, under ideal filling conditions. Filled tubes were weighed again, in duplicate. For each 20 tubes series, mean bias (deviation from the ideal tube filling volume) and imprecision (coefficient of variation; CV%) were calculated. All biases were within +/- 10%. Within-lot and between-lot variation in filling volume was acceptable, and comprised between 0.4 and 2.4%. Greiner tubes were the most accurate (bias, -1.0 to 2.4%), followed by Kima (bias, -7.8 to -5.9%) and Becton Dickinson (bias, -9.6 to 3.3%) tubes. The highest between-lot difference was noted for Becton Dickinson tubes (up to 12.9%), followed by Greiner and Kima tubes (up to 3.4 and 1.8%, respectively). Although coagulation tubes filling accuracy was within +/- 10% for all three tested manufacturers, the overall bias was found to be variable among manufacturers and lots. Major effort shall be made by blood tube manufacturers for improving standardization of their products.
引用
收藏
页码:276 / 279
页数:4
相关论文
共 10 条
  • [1] THROMBIN GENERATION IN DIFFERENT COMMERCIAL SODIUM CITRATE BLOOD TUBES
    Salvagno, Gian Luca
    Demonte, Davide
    Gelati, Matteo
    Poli, Giovanni
    Favaloro, Emmanuel J.
    Lippi, Giuseppe
    JOURNAL OF MEDICAL BIOCHEMISTRY, 2020, 39 (01) : 19 - 24
  • [2] Sodium citrate vacuum tubes validation: preventing preanalytical variability in routine coagulation testing
    Lima-Oliveira, Gabriel
    Lippi, Giuseppe
    Salvagno, Gian Luca
    Montagnana, Martina
    Picheth, Geraldo
    Guidi, Gian Cesare
    BLOOD COAGULATION & FIBRINOLYSIS, 2013, 24 (03) : 252 - 255
  • [3] Routine coagulation testing in Vacutainer® Citrate Plus tubes filled at minimum or optimal volume
    Balboni, Fiamma
    Barbui, Stefania
    Gallo, Marco
    Berardi, Margherita
    Vezzosi, Marco
    Lippi, Giuseppe
    DIAGNOSIS, 2020, 7 (01) : 55 - 60
  • [4] Guidance on the critical shortage of sodium citrate coagulation tubes for hemostasis testing
    Gosselin, Robert C.
    Bowyer, Annette
    Favaloro, Emmanuel J.
    Johnsen, Jill M.
    Lippi, Giuseppe
    Marlar, Richard A.
    Neeves, Keith
    Rollins-Raval, Marian A.
    JOURNAL OF THROMBOSIS AND HAEMOSTASIS, 2021, 19 (11) : 2857 - 2861
  • [5] Impact of incorrect filling of citrate blood sampling tubes on thromboelastometry
    Haas, Thorsten
    Spielmann, Nelly
    Cushing, Melissa
    SCANDINAVIAN JOURNAL OF CLINICAL & LABORATORY INVESTIGATION, 2015, 75 (08): : 717 - 719
  • [6] Accuracy of Citrate Anticoagulant Amount, Volume, and Concentration in Evacuated Blood Collection Tubes Evaluated with UV Molecular Absorption Spectrometry on a Purified Water Model
    Gros, Natasa
    Klobucar, Tadej
    Gaber, Klara
    MOLECULES, 2023, 28 (02):
  • [7] The determination of euglobulin clot lysis time reference intervals in Beckton Dickinson and Kima 3.2% sodium citrate coagulation tubes
    Kuktic, Ivona
    Pasalic, Ante
    Cancarevic, Ognjen
    Galic, Edvard
    Saracevic, Andrea
    Biljak, Vanja Radisic
    BLOOD COAGULATION & FIBRINOLYSIS, 2025, 36 (01) : 14 - 17
  • [8] A comparison of two sodium citrate concentrations in two evacuated blood collection systems for prothrombin time and ISI determination
    van den Besselaar, AMHP
    Chantarangkul, V
    Tripodi, A
    THROMBOSIS AND HAEMOSTASIS, 2000, 84 (04) : 664 - 667
  • [9] Incomplete filling of spray-dried K2EDTA evacuated blood tubes: impact on measuring routine hematological parameters on Sysmex XN-10
    Lippi, Giuseppe
    Dima, Francesco
    Tosi, Marco
    Carpene, Giovanni
    Celegon, Giovanni
    Favaloro, Emmanuel J.
    Salvagno, Gian Luca
    DIAGNOSIS, 2023, 10 (04) : 440 - 445
  • [10] Sodium citrate blood contamination by K2-ethylenediaminetetraacetic acid (EDTA): impact on routine coagulation testing
    Lima-Oliveira, G.
    Salvagno, G. L.
    Danese, E.
    Favaloro, E. J.
    Guidi, G. C.
    Lippi, G.
    INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY, 2015, 37 (03) : 403 - 409