Fixation of patella fractures with braided polyester suture: a biomechanical study

被引:103
作者
Patel, VR [1 ]
Parks, BG [1 ]
Wang, YB [1 ]
Ebert, FR [1 ]
Jinnah, RH [1 ]
机构
[1] St Georges Healthcare, Dept Orthopaed, London, England
来源
INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED | 2000年 / 31卷 / 01期
关键词
D O I
10.1016/S0020-1383(99)00190-4
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
We tested the quality of fixation of displaced transverse patella fractures using braided polyester suture to investigate the suitability of this material as an alternative to stainless steel wire for fixation of these fractures. Osteotomies were created to simulate fractures of the patella in ten cadaveric knee specimens and were sequentially fixed using two techniques: the modified tension-band technique and the longitudinal anterior band (Lotke) technique. Each technique was implemented using either 1.25-mm stainless steel wire or 7-metric braided polyester suture (No. 5 Ethibond). The quality of fixation for each technique was tested by measuring the fracture gap during three simulated extensions of the knee against gravity on a materials testing machine. All techniques behaved comparably under the loading conditions used. In the four groups, there was no fixation failure (fracture gap >3 mm) nor any significant difference between the mean maximum fracture gaps. The quality of fixation for braided polyester suture was comparable to that of stainless steel wire for such fractures, providing sufficient stability to withstand loads likely to be encountered during postoperative rehabilitation. Our results support the use of braided polyester suture as an alternative to stainless steel wire for fixation of displaced patella fractures. (C) 2000 Elsevier Science Ltd. All rights reserved.
引用
收藏
页码:1 / 6
页数:6
相关论文
共 24 条
  • [1] Benjamin J, 1987, J Orthop Trauma, V1, P219, DOI 10.1097/00005131-198701030-00004
  • [2] BIGLIANI LU, 1996, ROCKWOOD GREENS FRAC, V4, P1055
  • [3] COMMINUTED DISPLACED FRACTURES OF THE PATELLA
    BOSTMAN, O
    KIVILUOTO, O
    NIRHAMO, J
    [J]. INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 1981, 13 (03): : 196 - 202
  • [4] FATIGUE TESTING OF CERCLAGE STAINLESS-STEEL WIRE FIXATION
    BOSTROM, MPG
    ASNIS, SE
    ERNBERG, JJ
    WRIGHT, TM
    GIDDINGS, VL
    BERBERIAN, WS
    MISSRI, AAE
    [J]. JOURNAL OF ORTHOPAEDIC TRAUMA, 1994, 8 (05) : 422 - 428
  • [5] EVALUATION OF METHODS OF INTERNAL-FIXATION OF TRANSVERSE PATELLA FRACTURES - A BIOMECHANICAL STUDY
    BURVANT, JG
    THOMAS, KA
    ALEXANDER, R
    HARRIS, MB
    [J]. JOURNAL OF ORTHOPAEDIC TRAUMA, 1994, 8 (02) : 147 - 153
  • [6] Buzzi R, 1989, Arch Putti Chir Organi Mov, V37, P283
  • [7] MECHANICAL-PROPERTIES OF SUTURE MATERIALS INVITRO AND AFTER INVIVO IMPLANTATION IN HORSES
    CAMPBELL, EJ
    BAILEY, JV
    [J]. VETERINARY SURGERY, 1992, 21 (05) : 355 - 361
  • [8] Capperauld I, 1976, Polim Med, V6, P167
  • [9] FRACTURES OF THE PATELLA
    CARPENTER, JE
    KASMAN, R
    MATTHEWS, LS
    [J]. JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1993, 75A (10) : 1550 - 1561
  • [10] The use of non-absorbable suture in the fixation of patellar fractures: a preliminary report
    Chatakondu, SC
    Abhaykumar, S
    Elliott, DS
    [J]. INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 1998, 29 (01): : 23 - 27