Accurate determination and certification of bromine in plastic by isotope dilution inductively coupled plasma mass spectrometry

被引:21
|
作者
Ohata, Masaki [1 ]
Miura, Tsutomu [1 ]
机构
[1] Natl Metrol Inst Japan, Inorgan Analyt Chem Div, Inorgan Stand Sect, Tsukuba, Ibaraki 3058563, Japan
关键词
Isotope dilution inductively coupled plasma; mass spectrometry (ID-ICPMS); Bromine analysis in plastic; Microwave acid digestion procedure; Memory effect; Certified reference material (CRM); RoHS directive; CERTIFIED REFERENCE MATERIALS; HEAVY-METAL ANALYSIS; ICP-MS; ORGANIC-COMPOUNDS; HALOGENS; UNCERTAINTY; RESPECT; SAMPLES; SULFUR; LEAD;
D O I
10.1016/j.aca.2014.06.023
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
The accurate analytical method of bromine (Br) in plastic was developed by an isotope dilution inductively coupled plasma mass spectrometry (ID-ICPMS). The figures of merit of microwave acid digestion procedures using polytetrafluoroethylene (PTFE) or quartz vessels were studied and the latter one was suitable for Br analysis since its material was free from Br contamination. The sample dilution procedures using Milli-Q water or ammonium (NH3) solution were also studied to remove memory effect for ICPMS measurement. Although severe memory effect was observed on Milli-Q water dilution, NH3 solution could remove it successfully. The accuracy of the ID-ICPMS was validated by a certified reference material (CRM) as well as the comparison with the analytical result obtained by an instrumental neutron activation analysis (INAA) as different analytical method. From these results, the ID-ICPMS developed in the present study could be evaluated as accurate analytical method of Br in plastic materials and it could apply to certification of Br in candidate plastic CRM with respect to such regulations related to RoHS (restriction of the use of hazardous substances in electrical and electronics equipment) directive. (C) 2014 Elsevier B. V. All rights reserved.
引用
收藏
页码:23 / 30
页数:8
相关论文
共 50 条
  • [41] A rapid isotope dilution inductively coupled plasma mass spectrometry procedure for uranium bioassay
    William S. Kinman
    Stephen P. Lamont
    Robert E. Steiner
    Journal of Radioanalytical and Nuclear Chemistry, 2009, 282 : 1027 - 1030
  • [42] Determination of mercury in tuna fish tissue using isotope dilution-inductively coupled plasma mass spectrometry
    Lee, SH
    Suh, JK
    MICROCHEMICAL JOURNAL, 2005, 80 (02) : 233 - 236
  • [43] THE DETERMINATION OF LEAD IN BLOOD USING ISOTOPE-DILUTION INDUCTIVELY-COUPLED PLASMA-MASS SPECTROMETRY
    MURPHY, KE
    PAULSEN, PJ
    FRESENIUS JOURNAL OF ANALYTICAL CHEMISTRY, 1995, 352 (1-2): : 203 - 208
  • [44] Determination of boron in steel by isotope-dilution inductively coupled plasma mass spectrometry after matrix separation
    Park, CJ
    BULLETIN OF THE KOREAN CHEMICAL SOCIETY, 2002, 23 (11) : 1541 - 1544
  • [45] The determination of dissolved zirconium and hafnium from seawater using isotope dilution inductively coupled plasma mass spectrometry
    McKelvey, BA
    Orians, KJ
    MARINE CHEMISTRY, 1998, 60 (3-4) : 245 - 255
  • [46] The determination of dissolved zirconium and hafnium from seawater using isotope dilution inductively coupled plasma mass spectrometry
    Depts. of Earth and Ocean S., University of British Columbia, Vanvouver, BC V6T 1Z4, Canada
    不详
    Mar. Chem., 3-4 (245-255):
  • [47] Determination of metal impurities in sulfamic acid by isotope dilution electrothermal vaporization inductively coupled plasma mass spectrometry
    Nonose, N
    Kubota, M
    JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1998, 13 (02) : 151 - 156
  • [48] Speciation of Mercury in Microalgae by Isotope Dilution-inductively Coupled Plasma Mass Spectrometry
    Song, Beibei
    He, Huijun
    Chen, Lufeng
    Yang, Shifeng
    Yin Yongguang
    Li, Yanbin
    ANALYTICAL LETTERS, 2017, 50 (13) : 2161 - 2176
  • [49] Determination of serum calcium levels by 42Ca isotope dilution inductively coupled plasma mass spectrometry
    Han, Bingqing
    Ge, Menglei
    Zhao, Haijian
    Yan, Ying
    Zeng, Jie
    Zhang, Tianjiao
    Zhou, Weiyan
    Zhang, Jiangtao
    Wang, Jing
    Zhang, Chuanbao
    CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2018, 56 (01) : 51 - 58
  • [50] Comparison of external calibration and isotope dilution methods for iodine determination in foods by inductively coupled plasma mass spectrometry
    Photi, Juntima
    Srisakda, Premmin
    Jurkvon, Tiyanun
    Zeder, Christophe
    Judprasong, Kunchit
    FOOD CHEMISTRY-X, 2025, 25