Bone stress for a mini-implant close to the roots of adjacent teeth-3D finite element analysis

被引:59
作者
Motoyoshi, M. [1 ]
Ueno, S. [1 ]
Okazaki, K. [1 ]
Shimizu, N. [1 ]
机构
[1] Nihon Univ, Sch Dent, Dept Orthodont, Div Clin Res,Dent Res Ctr,Chiyoda Ku, Tokyo 1018310, Japan
关键词
temporary anchorage devices; anchorage; finite element method; bone resorption; POSTERIOR REGION; PLACEMENT TORQUE; TITANIUM SCREWS; ANCHORAGE; THICKNESS; STABILITY; FAILURE;
D O I
10.1016/j.ijom.2009.02.011
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
This study aimed to evaluate stress in the bone when an orthodontic mini-implant is close to the roots of adjacent teeth using finite element models (FEMs), and to investigate the causes of the high implant failure rate in the mandible. Four FEMs were used: the implant touches nothing; the implant touches the surface of the periodontal membrane; part of the screw thread is embedded in the periodontal membrane; and the implant touches the root. The effect of cortical bone thickness was evaluated using values of 1, 2 and 3 mm. Maximum stress value and stress distribution on the bone elements was determined. Maximum stress on the bone increased when the mini-implant was close to the root. When the implant touched the root, stress increased to 140 MPa or more, and bone resorption could be predicted. Stress was higher for a cortical bone thickness of 2 mm than for other thicknesses. Cortical bone 2 mm thick had a higher risk for bone resorption. A mandible with an average cortical bone thickness of 2 mm may have a higher risk for implant loosening than a maxilla with the same degree of root proximity, which may be related to the lower success rate in the mandible.
引用
收藏
页码:363 / 368
页数:6
相关论文
共 50 条
[21]   Association between the pen-implant bone structure and stress distribution around the mandibular canal: A three-dimensional finite element analysis [J].
Kinoshita, Hideaki ;
Nakahara, Ken ;
Matsunaga, Satoru ;
Usami, Akinobu ;
Yoshinari, Masao ;
Takano, Naoki ;
Ide, Yoshinobu ;
Abe, Shinichi .
DENTAL MATERIALS JOURNAL, 2013, 32 (04) :637-642
[22]   Effect of Integration Patterns Around Implant Neck on Stress Distribution in Peri-Implant Bone: A Finite Element Analysis [J].
Han, Jingyun ;
Sun, Yuchun ;
Wang, Chao .
JOURNAL OF PROSTHODONTICS-IMPLANT ESTHETIC AND RECONSTRUCTIVE DENTISTRY, 2017, 26 (06) :549-558
[23]   Finite element analysis of bone and implant stresses for customized 3D-printed orthopaedic implants in fracture fixation [J].
Yan, Lina ;
Lim, Joel Louis ;
Lee, Jun Wei ;
Tia, Clement Shi Hao ;
O'Neill, Gavin Kane ;
Chong, Desmond Y. R. .
MEDICAL & BIOLOGICAL ENGINEERING & COMPUTING, 2020, 58 (05) :921-931
[24]   Influence of marginal bone resorption on stress around an implant - a three-dimensional finite element analysis [J].
Kitamura, E ;
Stegaroiu, R ;
Nomura, S ;
Miyakawa, O .
JOURNAL OF ORAL REHABILITATION, 2005, 32 (04) :279-286
[25]   Mechanical and Thermal Stress Behavior of a Conservative Proposed Veneer Preparation Design for Restoring Misaligned Anterior Teeth: A 3D Finite Element Analysis [J].
Dawood, Shilan Nawzad ;
Al-Zahawi, Abdulsalam Rasheed ;
Sabri, Laith Abed .
APPLIED SCIENCES-BASEL, 2020, 10 (17)
[26]   The effect of implant neck microthread design on stress distribution of peri-implant bone with different level: A finite element analysis [J].
Jin, Zhi-Heng ;
Peng, Meng-Dong ;
Li, Qing .
JOURNAL OF DENTAL SCIENCES, 2020, 15 (04) :466-471
[27]   Bone stress and strain modification in diastema closure: 3D analysis using finite element method [J].
Geramy, Allahyar ;
Bouserhal, Joseph ;
Martin, Domingo ;
Baghaeian, Pedram .
INTERNATIONAL ORTHODONTICS, 2015, 13 (03) :274-286
[28]   Stress distribution on teeth restored with veneers using various incisal preparation designs: A 3D finite element analysis study [J].
Kintopp, Cibele Candida de Almeida ;
Diogenes, Alysson Nunes ;
Lopes, Ricardo Tadeu ;
Weber, Katia Raquel ;
Rezende, Carlos Eduardo Edwards ;
Kaizer, Marina da Rosa ;
Gonzaga, Carla Castiglia .
JOURNAL OF PROSTHETIC DENTISTRY, 2024, 131 (06) :1159.e1-1159.e10
[29]   Optimum selection of the dental implant diameter and length in the posterior mandible with poor bone quality - A 3D finite element analysis [J].
Li, Tao ;
Hu, Kaijin ;
Cheng, Libo ;
Ding, Yin ;
Ding, Yuxiang ;
Shao, Jinling ;
Kong, Liang .
APPLIED MATHEMATICAL MODELLING, 2011, 35 (01) :446-456
[30]   Effect of fibre posts, bone losses and fibre content on the biomechanical behaviour of endodontically treated teeth: 3D-finite element analysis [J].
Chieruzzi, Manila ;
Pagano, Stefano ;
Cianetti, Stefano ;
Lombardo, Guido ;
Kenny, Jose M. ;
Torre, Luigi .
MATERIALS SCIENCE AND ENGINEERING C-MATERIALS FOR BIOLOGICAL APPLICATIONS, 2017, 74 :334-346