CHARACTERISATION OF RADIATION FIELD FOR IRRADIATION OF BIOLOGICAL SAMPLES AT NUCLEAR REACTOR-COMPARISON OF TWIN DETECTOR AND RECOMBINATION METHODS

被引:16
作者
Golnik, N. [1 ]
Gryzinski, M. A. [2 ]
Kowalska, M. [3 ]
Meronka, K. [3 ]
Tulik, P. [1 ]
机构
[1] Warsaw Univ Technol, Inst Metrol & Biomed Engn, PL-02525 Warsaw, Poland
[2] Natl Ctr Nucl Res, PL-05400 Otwock, Poland
[3] Cent Lab Radiol Protect, PL-03194 Warsaw, Poland
关键词
IONIZATION-CHAMBER; QUALITY; INDEX;
D O I
10.1093/rpd/nct341
中图分类号
X [环境科学、安全科学];
学科分类号
08 ; 0830 ;
摘要
Central Laboratory for Radiological Protection is involved in achieving scientific project on biological dosimetry. The project includes irradiation of blood samples in radiation fields of nuclear reactor. A simple facility for irradiation of biological samples has been prepared at horizontal channel of the nuclear reactor MARIA in NCBJ in Poland. The radiation field, composed mainly of gamma radiation and thermal neutrons, has been characterised in terms of tissue kerma using twin-detector technique and recombination chambers.
引用
收藏
页码:196 / 200
页数:5
相关论文
共 10 条
[1]  
GOLNIK N, 1994, RADIAT PROT DOSIM, V52, P35
[2]   Recombination index of radiation quality of medical high energy neutron beams [J].
Golnik, N ;
Cherevatenko, EP ;
Serov, AY ;
Shvidkij, SV ;
Sychev, BS ;
Zielczynski, M .
RADIATION PROTECTION DOSIMETRY, 1997, 70 (1-4) :215-218
[3]   Photon kerma determination using a high pressure graphite ionisation chamber in mixed fields of neutrons of energy 44 and 66 MeV [J].
Golnik, N ;
Brede, HJ ;
Schrewe, UJ ;
Zielczynski, M .
RADIATION PROTECTION DOSIMETRY, 2000, 88 (02) :135-142
[4]  
Golnik N., 1996, IAE20A
[5]  
Golnik N., 1985, P 5 S NEUTR DOS SEP, P717
[6]  
International Atomic Energy Agency, 1986, 260 INT AT EN AG
[7]  
International Atomic Energy Agency, 2001, 405 INT AT AN AG
[8]  
International Organization for Standardization, 2004, 192382004 ISO
[9]  
Pytel K., COMMUNICATION
[10]  
ZIELCZYNSKI M, 1994, RADIAT PROT DOSIM, V52, P419