Reducing radiation dose by application of optimized low-energy x-ray filters to K-edge imaging with a photon counting detector

被引:6
作者
Choi, Yu-Na [1 ]
Lee, Seungwan [2 ]
Kim, Hee-Joung [1 ]
机构
[1] Yonsei Univ, Coll Hlth Sci & Res, Inst Hlth Sci, Dept Radiol Sci, Wonju 220710, Gangwon Do, South Korea
[2] Konyang Univ, Coll Med Sci, Dept Radiol Sci, Daejeon 302812, South Korea
基金
新加坡国家研究基金会;
关键词
photon counting detector; K-edge imaging; radiation dose; COMPUTED-TOMOGRAPHY; DIGITAL MAMMOGRAPHY; CONTRAST AGENTS; PULSE PILEUP; SPECTRAL CT; CDTE; PERFORMANCE; QUALITY;
D O I
10.1088/0031-9155/61/2/N35
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
K-edge imaging with photon counting x-ray detectors (PCXDs) can improve image quality compared with conventional energy integrating detectors. However, low-energy x-ray photons below the K-edge absorption energy of a target material do not contribute to image formation in the K-edge imaging and are likely to be completely absorbed by an object. In this study, we applied x-ray filters to the K-edge imaging with a PCXD based on cadmium zinc telluride for reducing radiation dose induced by low-energy x-ray photons. We used aluminum (Al) filters with different thicknesses as the low-energy x-ray filters and implemented the iodine K-edge imaging with an energy bin of 34-48 keV at the tube voltages of 50, 70 and 90 kVp. The effects of the low-energy x-ray filters on the K-edge imaging were investigated with respect to signal-difference-to-noise ratio (SDNR), entrance surface air kerma (ESAK) and figure of merit (FOM). The highest value of SDNR was observed in the K-edge imaging with a 2 mm Al filter, and the SDNR decreased as a function of the filter thicknesses. Compared to the K-edge imaging with a 2 mm Al filter, the ESAK was reduced by 66%, 48% and 39% in the K-edge imaging with a 12 mm Al filter for 50 kVp, 70 kVp and 90 kVp, respectively. The FOM values, which took into account the ESAK and SDNR, were maximized for 8, 6 to 8 and 4 mm Al filters at 50 kVp, 70 kVp and 90 kVp, respectively. We concluded that the use of an optimal low-energy filter thickness, which was determined by maximizing the FOM, could significantly reduce radiation dose while maintaining image quality in the K-edge imaging with the PCXD.
引用
收藏
页码:N35 / N49
页数:15
相关论文
共 43 条
[1]   A 20 kpixels CdTe photon-counting imager using XPAD chip [J].
Basolo, S. ;
Berar, J. F. ;
Boudet, N. ;
Breugnon, P. ;
Chantepie, B. ;
Clemens, J. C. ;
Delpierre, P. ;
Dinkespiler, B. ;
Hustache, S. ;
Medjoubi, K. ;
Menouni, M. ;
Morel, C. ;
Pangaud, P. ;
Vigeolas, E. .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2008, 589 (02) :268-274
[2]   Spectral CT imaging of vulnerable plaque with two independent biomarkers [J].
Baturin, Pavlo ;
Alivov, Yahya ;
Molloi, Sabee .
PHYSICS IN MEDICINE AND BIOLOGY, 2012, 57 (13) :4117-4138
[3]  
Bushberg J.T., 2011, The Essential Physics of Medical Imaging, Vthird
[4]   The effects of photon flux on energy spectra and imaging characteristics in a photon-counting x-ray detector [J].
Cho, H-M ;
Kim, H-J ;
Choi, Y-N ;
Lee, S-W ;
Ryu, H-J ;
Lee, Y-J .
PHYSICS IN MEDICINE AND BIOLOGY, 2013, 58 (14) :4865-4879
[5]  
Choi YN, 2011, IEEE NUCL SCI CONF R, P2855, DOI 10.1109/NSSMIC.2011.6152504
[6]   Contrast agents in magnetic resonance imaging [J].
Dawson, P .
EUROPEAN JOURNAL OF RADIOLOGY, 1996, 23 (03) :201-204
[7]   Image-based spectral distortion correction for photon-counting x-ray detectors [J].
Ding, Huanjun ;
Molloi, Sabee .
MEDICAL PHYSICS, 2012, 39 (04) :1864-1876
[8]   Application of contrast-to-noise ratio in optimizing beam quality for digital chest radiography: comparison of experimental measurements and theoretical simulations [J].
Doyle, P. ;
Martin, C. J. ;
Gentle, D. .
PHYSICS IN MEDICINE AND BIOLOGY, 2006, 51 (11) :2953-2970
[9]   CdTe and CdZnTe materials for room-temperature X-ray and gamma ray detectors [J].
Eisen, Y ;
Shor, A .
JOURNAL OF CRYSTAL GROWTH, 1998, 184 :1302-1312
[10]   Performance of computed tomography for contrast agent concentration measurements with monochromatic x-ray beams:: comparison of K-edge versus temporal subtraction [J].
Elleaume, H ;
Charvet, AM ;
Corde, S ;
Estéve, F ;
Le Bas, JF .
PHYSICS IN MEDICINE AND BIOLOGY, 2002, 47 (18) :3369-3385