Safety and Effectiveness of the Celect Inferior Vena Cava Filter: Preliminary Results

被引:39
|
作者
Sangwaiya, Minal Jagtiani [1 ]
Marentis, Theodore C.
Walker, T. Gregory [1 ]
Stecker, Michael [2 ]
Wicky, Stephan T. [1 ]
Kalva, Sanjeeva P. [1 ]
机构
[1] Harvard Univ, Massachusetts Gen Hosp, Sch Med, Dept Radiol, Boston, MA 02114 USA
[2] Harvard Univ, Brigham & Womens Hosp, Sch Med, Dept Radiol, Boston, MA 02114 USA
关键词
CLINICAL-EXPERIENCE; PULMONARY-EMBOLISM; FOLLOW-UP; RETRIEVAL; PENETRATION; PROTECTION; TEMPORARY; OUTCOMES;
D O I
10.1016/j.jvir.2009.05.033
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
PURPOSE: To evaluate the safety and effectiveness of the Celect inferior vena cava (IVC) filter during implantation, retrieval, and short-term follow-up. MATERIALS AND METHODS: The clinical data of 73 patients (46 men; age range, 22-89 years) who had a Celect IVC filter implanted between August 2007 and June 2008 were reviewed. Twenty-one (28.8%) presented with pulmonary embolism (PE), 15 (20.54%) with deep vein thrombosis (DVT), 12 (16.4%) with both, and the rest (34.26%) with other symptoms. Indications for filter placement were contraindication to anticoagulation (n = 38; 52%), prophylaxis/added protection (n = 22; 30%), failure of anticoagulation (n = 11; 15%), and complications of anticoagulation (n = 2; 3%). Filters were placed in the infrarenal (n = 71) or suprarenal (n = 2) IVC. Follow-up data were reviewed for filter-related complications and recurrent PE. RESULTS: All filters were successfully deployed. Immediately after fluoroscopy-guided filter deployment in 61 patients, four filters (6.5%) showed significant tilt. During follow-up (mean, 68 days +/- 73), three patients developed symptoms of PE after filter placement; however, computed tomographic (CT) pulmonary angiography demonstrated new PE in only two. Imaging follow-up with radiography (n = 32), CT (n = 11), and/or angiography (n = 4) in 47 patients (at a mean of 62 days +/- 75) showed no filter migration. Follow-up abdominal CT (at a mean of 69 days +/- 58) was available in 18 patients and demonstrated filter-related problems in seven (39%). These included penetration of filter legs in four and fracture/migration of filter components in one. Fourteen filters were successfully retrieved after a median period of 84 days. CONCLUSIONS: The Celect IVC filter can be safely placed but is related to a high incidence of caval filter leg penetration. Symptomatic PE after filter placement confirmed by CT occurred in 2.8% of patients.
引用
收藏
页码:1188 / 1192
页数:5
相关论文
共 50 条
  • [1] Symptomatic caval penetration by a Celect inferior vena cava filter
    Conor O. Bogue
    Philip R. John
    Bairbre L. Connolly
    David J. Rea
    Joao G. Amaral
    Pediatric Radiology, 2009, 39 : 1110 - 1113
  • [2] Symptomatic caval penetration by a Celect inferior vena cava filter
    Bogue, Conor O.
    John, Philip R.
    Connolly, Bairbre L.
    Rea, David J.
    Amaral, Joao G.
    PEDIATRIC RADIOLOGY, 2009, 39 (10) : 1110 - 1113
  • [3] Safety and Effectiveness of the Denali Inferior Vena Cava Filter: Intermediate Follow-Up Results
    Reis, Stephen P.
    Kovoor, Jerry
    Sutphin, Patrick D.
    Toomay, Seth
    Trimmer, Clayton
    Pillai, Anil
    Reddick, Mark
    Kalva, Sanjeeva P.
    VASCULAR AND ENDOVASCULAR SURGERY, 2016, 50 (06) : 385 - 390
  • [4] Preliminary Experience With Option Inferior Vena Cava Filter
    Iqbal, Shams I.
    Elmi, Azadeh
    Hedgire, Sandeep
    Yeddula, Kalpana
    Ganguli, Suvranu
    Walker, T. Gregory
    Salazar, Gloria M.
    Wicky, Stephan
    Kalva, Sanjeeva P.
    VASCULAR AND ENDOVASCULAR SURGERY, 2013, 47 (01) : 24 - 29
  • [5] The RETRIEVE Trial: Safety and Effectiveness of the Retrievable Crux Vena Cava Filter
    Smouse, H. Bob
    Mendes, Robert
    Bosiers, Marc
    Van Ha, Thuong G.
    Crabtree, Tami
    JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 2013, 24 (05) : 609 - 621
  • [6] Retrospective Review of 120 Celect Inferior Vena Cava Filter Retrievals: Experience at a Single Institution
    Zhou, Dayong
    Spain, James
    Moon, Eunice
    Mclennan, Gordon
    Sands, Mark J.
    Wang, Weiping
    JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 2012, 23 (12) : 1557 - 1563
  • [7] Inferior vena cava filter use and patient safety: legacy or science?
    Geerts, William
    Selby, Rita
    HEMATOLOGY-AMERICAN SOCIETY OF HEMATOLOGY EDUCATION PROGRAM, 2017, : 686 - 692
  • [8] Initial experience in 115 patients with the retrievable Cook Celect vena cava filter
    Doody, O.
    Given, M. F.
    Kavnoudias, H.
    Street, M.
    Thomson, K. R.
    Lyon, S. M.
    JOURNAL OF MEDICAL IMAGING AND RADIATION ONCOLOGY, 2009, 53 (01) : 64 - 68
  • [9] Predicting the Safety and Effectiveness of Inferior Vena Cava Filters Study: Design of a unique safety and effectiveness study of inferior vena cava filters in clinical practice
    Gillespie, David L.
    Spies, James B.
    Siami, F. Sandra
    Rectenwald, John E.
    White, Rodney A.
    Johnson, Matthew S.
    JOURNAL OF VASCULAR SURGERY-VENOUS AND LYMPHATIC DISORDERS, 2020, 8 (02) : 187 - +
  • [10] Safety of inferior vena cava filter retrieval in anticoagulated patients
    Hoppe, Hanno
    Kaufman, John A.
    Barton, Robert E.
    Petersen, Bryan D.
    Lakin, Paul C.
    DeLoughery, Thomas G.
    Irani, Zubin
    Yavuz, Kivilcim
    Keller, Frederick S.
    CHEST, 2007, 132 (01) : 31 - 36