Determination of Aluminium by Electrothermal Atomization Atomic Absorption Spectrometry in Serum to Characterize Hemodialysis Toxicity

被引:43
|
作者
Qadir, Muhammad Abdul [1 ]
Ahmed, Mahmood [1 ]
Shahzad, Shabnam [1 ]
机构
[1] Univ Punjab, Inst Chem, Lahore, Pakistan
关键词
Hemodialysis; Electrothermal atomization atomic absorption spectrometry; Serum; ET-AAS; Aluminum; BONE; FOOD;
D O I
10.1080/00032719.2014.930872
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A simple and cost-effective method is described for the determination of aluminum by electrothermal atomic absorption spectrometry (ET-AAS) in serum of hemodialysis patients and healthy subjects. The only preparative step required is the dissolution of the serum sample in 0.2% magnesium nitrate matrix modifier and separate diluents 0.01M EDTA and 0.1% Triton X-100. The calibration curve was linear from 20 to 100 mu g/L with correlation coefficients of 0.9993 and 0.9998 for EDTA and Triton X-100, respectively. The sensitivity of the method for aluminum at the 309.3nm line was 74 pg. The instrumental and method limits of detection were 2.2 mu g/L and 4.4 mu g/L, respectively. The aluminum concentrations of forty serum samples from hemodialysis patients and healthy subjects were determined and the mean values were 170.9 +/- 6.8 mu g/L and 47.3 +/- 9.3 mu g/L, respectively, whereas the permissible limit for aluminum in blood serum is 10 mu g/L. The high level of Al in serum was related to oral phosphate binding agents and dialysis treatment.
引用
收藏
页码:147 / 153
页数:7
相关论文
共 50 条
  • [41] Determination of lead in aluminum and magnesium antacids using electrothermal atomic absorption spectrometry
    Portugal, Lindomar A.
    Matos, Geraldo D.
    Lima, Daniel C.
    Brito, Geysa B.
    Fernandes, Andrea P.
    Ferreira, Sergio L. C.
    MICROCHEMICAL JOURNAL, 2011, 98 (01) : 29 - 31
  • [42] DETERMINATION OF BERYLLIUM IN VARIOUS TYPES OF NATURAL WATERS BY ELECTROTHERMAL ATOMIC ABSORPTION SPECTROMETRY
    Hagarova, I.
    Machackova, L.
    Zemberyova, M.
    Navojova, V.
    GLOBAL NEST JOURNAL, 2012, 14 (01): : 40 - 47
  • [43] Heated Slurry Sampling for the Determination of Cadmium in Food by Electrothermal Atomic Absorption Spectrometry
    Peng, Yue'e
    Guo, Wei
    Zhang, Ping
    Jin, Lanlan
    Hu, Shenghong
    ANALYTICAL LETTERS, 2015, 48 (18) : 2894 - 2907
  • [44] Determination of Cd, Pb and As in sediments of the Sava River by electrothermal atomic absorption spectrometry
    Murko, Simona
    Milacic, Radmila
    Veber, Marjan
    Scancar, Janez
    JOURNAL OF THE SERBIAN CHEMICAL SOCIETY, 2010, 75 (01) : 113 - 128
  • [45] Selective preconcentration and determination of tributyltin in fresh water by electrothermal atomic absorption spectrometry
    Bermejo-Barrera, P
    Anllo-Sendín, RM
    Cantelar-Barbazán, MJ
    Bermejo-Barrera, A
    ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2002, 372 (7-8) : 837 - 839
  • [46] Palladium release in electrothermal atomic absorption spectrometry
    Slaveykova, VI
    Manev, SG
    Lazarov, DL
    SPECTROSCOPY LETTERS, 1997, 30 (02) : 297 - 307
  • [47] Adaptive Calibration in Electrothermal Atomic Absorption Spectrometry
    Sadagov, Yu. M.
    Sadagov, A. Yu.
    JOURNAL OF ANALYTICAL CHEMISTRY, 2023, 78 (08) : 988 - 994
  • [48] Determination of cadmium by electrothermal atomic absorption spectrometry using electrochemical separation in a microcell
    Knápek, J
    Komárek, J
    Krásensky, P
    SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 2005, 60 (03) : 393 - 398
  • [49] Slurry sampling for determination of lead in marine plankton by electrothermal atomic absorption spectrometry
    Arslan, Zikri
    Tyson, Julian F.
    MICROCHEMICAL JOURNAL, 2007, 86 (02) : 227 - 234
  • [50] Coprecipitation with disulfide for the determination of traces of cobalt in water by electrothermal atomic absorption spectrometry
    Hiraide, M
    Chen, ZS
    Kawaguchi, H
    TALANTA, 1996, 43 (07) : 1131 - 1136