Intermediate Range Wireless Power Transfer With Segmented Coil Transmitters for Implantable Heart Pumps

被引:84
作者
Tang, Sai Chun [1 ]
Lun, Tian Le Tim [2 ]
Guo, Ziyan [2 ]
Kwok, Ka-Wai [2 ]
McDannold, Nathan J. [1 ]
机构
[1] Harvard Med Sch, Brigham & Womens Hosp, Dept Radiol, 75 Francis St, Boston, MA 02115 USA
[2] Univ Hong Kong, Dept Engn Mech, Pokfulam, Hong Kong, Peoples R China
关键词
Artificial heart; implantable medical device; left ventricular assist device (LVAD); mid range wireless power transfer; segmented coil transmitter; ENERGY TRANSMISSION-SYSTEM; DENSITY; PERFORMANCE; DC;
D O I
10.1109/TPEL.2016.2584558
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
In wireless power transfer systems, substantial differences in transmission range and alignment sensitivity can occur when the transmitting coil dimensions are varied. A coil with larger inner and outer diameter was found to provide a wider transmission range and lower alignment sensitivity. Accordingly, we developed a larger coil (24x30 cm(2)) designed to be embedded in the back of a vest to power dc pumps for artificial hearts or left ventricular assist devices. To significantly reduce the required transmitting coil voltage, the coil was divided into eight segments with resonant capacitors. The coil was operated at 6.78 MHz and evaluated with a 5.3-cm diameter receiving coil. A circuit model for the energy coupling coils was developed to predict the output power and efficiency. Having a coil separation of 7.7 cm, we measured an output power of 48.2 W and a corresponding energy efficiency higher than 80%. The energy coupling coils were also evaluated with a dc pump in parallel to an additional load representing the power consumed by auxiliary circuits of an implanted heart pump. Experimental results showed that the proposed coil segmentation technique for mid-range wireless energy transfer can significantly reduce the transmitter voltage to a safe level (similar to 10 V-rms).
引用
收藏
页码:3844 / 3857
页数:14
相关论文
共 36 条
  • [1] Ahlbom A, 1998, HEALTH PHYS, V74, P494
  • [2] Examination of circuit parameters for stable high efficiency TETS for artificial hearts
    Arai, S
    Miura, H
    Sato, F
    Matsuki, H
    Sato, T
    [J]. IEEE TRANSACTIONS ON MAGNETICS, 2005, 41 (10) : 4170 - 4172
  • [3] Bartlett R., 2009, ENCY INT SPORTS STUD, P164
  • [4] Analysis, Design, and Control of a Transcutaneous Power Regulator for Artificial Hearts
    Chen, Qianhong
    Wong, Siu Chung
    Tse, Chi K.
    Ruan, Xinbo
    [J]. IEEE TRANSACTIONS ON BIOMEDICAL CIRCUITS AND SYSTEMS, 2009, 3 (01) : 23 - 31
  • [5] A Novel Low Temperature Transcutaneous Energy Transfer System Suitable for High Power Implantable Medical Devices: Performance and Validation in Sheep
    Dissanayake, Thushari D.
    Budgett, David M.
    Hu, Patrick
    Bennet, Laura
    Pyner, Susan
    Booth, Lindsea
    Amirapu, Satya
    Wu, Yanzhen
    Malpas, Simon C.
    [J]. ARTIFICIAL ORGANS, 2010, 34 (05) : E160 - E167
  • [6] DOLAN AM, 1966, T AM SOC ART INT ORG, V12, P275
  • [7] Modeling and Simulation of a Transcutaneous Energy Transmission System Used in Artificial Organ Implants
    Fang, Wan
    Liu, Wei
    Qian, Jie
    Tang, Houjun
    Ye, Pengsheng
    [J]. ARTIFICIAL ORGANS, 2009, 33 (12) : 1069 - 1074
  • [8] PERFORMANCE OF A TUNED FERRITE CORE TRANSCUTANEOUS TRANSFORMER
    FRAIM, FW
    HUFFMAN, FN
    [J]. IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1971, BM18 (05) : 352 - &
  • [9] FULLER JW, 1968, IEEE T BIO-MED ENG, VBM15, P63
  • [10] Gabriel C, 1996, ALOETR19960037 AIR F