Development of a transfer standard for the measurement of low Rn-222 activity concentration in air

被引:4
作者
Linzmaier, Diana [1 ]
Roettger, Annette [1 ]
机构
[1] Phys Tech Bundesanstalt, D-38116 Braunschweig, Germany
关键词
Radon; alpha-Spectrometry; Calibration; Reference atmospheres; Metrology; RADON; CHAMBER;
D O I
10.1016/j.apradiso.2013.11.076
中图分类号
O61 [无机化学];
学科分类号
070301 ; 081704 ;
摘要
A large volume transfer standard has been developed to calibrate commercial radon measurement devices in a homogeneous Rn-222 reference atmosphere. The transfer standard serves for the realization, maintenance and dissemination of the unit Bq/m(3) below 1 kBq/m(3). The transfer standard consists of a multi-wire impulse ionization chamber, electronic measurement equipment and the corresponding software. The multi-wire impulse ionization chamber is composed of two parallel printed boards. Between the boards, 478 vertical electrode wires are soldered in to include a volume of 10 L. In the dependency of the energy, the alpha-particles of Rn-222 and the decay products are detected in the active volume of the multi-wire impulse ionization chamber. By means of the measurement software an alpha-energy spectrum is produced. Via different algorithms, the activity concentration of the Rn-222 reference atmosphere is determined. (c) 2013 Elsevier Ltd. All rights reserved.
引用
收藏
页码:306 / 309
页数:4
相关论文
共 11 条