Effect of design on the initial stability of press-fit cups in the presence of acetabular rim defects: experimental evaluation of the effect of adding circumferential fins

被引:12
作者
Huber, Wolfgang O. [1 ]
Noble, Philip C. [2 ]
机构
[1] Herz Jesu Hosp Vienna, Dept Orthoped, A-1030 Vienna, Austria
[2] Baylor Coll Med, Inst Orthoped Res & Educ, Houston, TX 77030 USA
关键词
Acetabularimplants; Initial stability; Primary and revision arthroplasty; Fin system; TOTAL HIP-ARTHROPLASTY; SCREW-FIXATION; FOLLOW-UP; COMPONENTS; REVISION; REPLACEMENT; OSTEOLYSIS; 10-YEAR; CONTACT; SOCKETS;
D O I
10.1007/s00264-013-2187-8
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
This experimental study was undertaken to examine the fixation characteristics of a six-finned acetabular cup in both primary and revision arthroplasty in comparison with two other commonly used cup designs without fins. All three cup designs (AnanovaA (R) [Intraplant], PlasmacupA (R) NSC [Aesculap]; Exceed ABT (TM) [Biomet]) were implanted into validated models of normal and revision acetabula. The defect models were designed to simulate a dorso-cranial rim defect of 90A degrees width and 10 mm in depth (moderate rim defect) and a dorso-cranial rim defect of 130A degrees width and 15 mm in depth (severe rim defect). The fixation strength of the three cup designs was tested by cyclically edge-loading the implanted cups using a mechanical testing machine. The six-finned Ananova implant exhibited greater resistance to foam-cup interface motion than both the Plasmacup and Exceed ABT implants. The largest average differences were observed in the resistance to ultimate spin-out, with Ananova outperforming Exceed ABT and Plasmacup by 26 % and 17 % in the moderate and by 36 % and 38 % in the severe defect models, respectively. The six-finned Ananova cup was significantly more resistant to edge loading both in the normal acetabulum and in acetabula with moderate to severe dorso-cranial rim defects than cup designs without fins, indicating that it may cover a wider range of clinical indications than conventional press-fit cups and provide clinicians with the confidence that, in primary and simple revision arthroplasty, adequate fixation strength can be obtained.
引用
收藏
页码:725 / 731
页数:7
相关论文
共 30 条
[1]  
Adler E, 1992, J Arthroplasty, V7, P295, DOI 10.1016/0883-5403(92)90052-R
[2]   Initial stability of a cementless acetabular cup design: Experimental investigation on the effect of adding fins to the rim of the cup [J].
Baleani, M ;
Fognani, R ;
Toni, A .
ARTIFICIAL ORGANS, 2001, 25 (08) :664-669
[3]   Loss in mechanical contact of cementless acetabular prostheses due to post-operative weight bearing: A biomechanical model [J].
Bellini, Chiara Maria ;
Galbusera, Fabio ;
Ceroni, Roberto Giacometti ;
Raimondi, Manuela Teresa .
MEDICAL ENGINEERING & PHYSICS, 2007, 29 (02) :175-181
[4]  
BERMAN AT, 1994, ORTHOPEDICS, V17, P963
[5]  
Cipriano Cara A, 2008, Am J Orthop (Belle Mead NJ), V37, pE18
[6]   THE INITIAL STABILITY OF UNCEMENTED ACETABULAR COMPONENTS [J].
CURTIS, MJ ;
JINNAH, RH ;
WILSON, VD ;
HUNGERFORD, DS .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1992, 74 (03) :372-376
[7]   Acetabular revision in total hip replacement with a press-fit cup [J].
Elke, R ;
Berli, B ;
Wagner, A ;
Morscher, EW .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2003, 85B (08) :1114-1119
[8]   Radiographic evaluation of screw position in revision total hip arthroplasty [J].
Galat, DD ;
Petrucci, JA ;
Wasielewski, RC .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2004, (419) :124-129
[9]   Does An Enhanced Surface Finish Improve Acetabular Fixation in Revision Total Hip Arthroplasty? [J].
Jamieson, Miranda L. ;
Russell, Robert D. ;
Incavo, Stephen J. ;
Noble, Philip C. .
JOURNAL OF ARTHROPLASTY, 2011, 26 (04) :644-648
[10]  
JASTY M, 1994, CLIN ORTHOP RELAT R, P111