SPECTRUM WEIGHTED RESPONSES OF SEVERAL DETECTORS IN MIXED FIELDS OF FAST AND THERMAL NEUTRONS

被引:3
作者
Kim, Sang In [1 ]
Chang, Insu [1 ,2 ]
Kim, Bong Hwan [1 ]
Kim, Yang Lyul [1 ]
Lee, Jung Il [1 ]
机构
[1] Korea Atom Energy Res Inst, Hlth Phys Team, Taejon 305353, South Korea
[2] Hanyang Univ, Dept Nucl Engn, Seoul 133791, South Korea
关键词
Spectrum Weighted Response; Response of Neutron Survey meter; IAEA Technical Reports Series 403; Reference neutron Field; DT Neutron Generator; Bonner Sphere Spectrometer; Fast Neutron Field; MAXIMUM-ENTROPY DECONVOLUTION; SPECTROMETER; ENERGY; CODE;
D O I
10.5516/NET.08.2013.029
中图分类号
TL [原子能技术]; O571 [原子核物理学];
学科分类号
0827 ; 082701 ;
摘要
The spectrum weighted responses of various detectors were calculated to provide guidance on the proper selection and use of survey instruments on the basis of their energy response characteristics on the neutron fields. To yield the spectrum weighted response, the detector response functions of 17 neutron-measuring devices were numerically folded with each of the produced calibration neutron spectra through the in-house developed software 'K-SWR'. The detectors' response functions were taken from the IAEA Technical Reports Series No. 403 (TRS-403). The reference neutron fields of 21 kinds with 2 spectra groups with different proportions of thermal and fast neutrons have been produced using neutrons from the Am-241-Be sources held in a graphite pile, a bare Am-241-Be source, and a DT neutron generator. Fluence-average energy (E-ave) varied from 3.8 MeV to 16.9 MeV, and the ambient-dose-equivalent rate [H*(10)/h] varied from 0.99 to 16.5 mSv/h.
引用
收藏
页码:273 / 280
页数:8
相关论文
共 17 条
[1]  
[Anonymous], 2005, MCNP-X TM User's Manual
[2]   The neutron fluence and H*(10) response of the new LB 6411 rem counter [J].
Burgkhardt, B ;
Fieg, G ;
Klett, A ;
Plewnia, A ;
Siebert, BRL .
RADIATION PROTECTION DOSIMETRY, 1997, 70 (1-4) :361-364
[3]   A NEW BONNER-SPHERE SET FOR HIGH-ENERGY NEUTRON MEASUREMENTS - MONTE-CARLO SIMULATION [J].
HSU, HH ;
ALVAR, KR ;
VASILIK, DG .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1994, 41 (04) :938-940
[4]  
*INT ORG STAND, 2000, ISO12789
[5]  
International Atomic Energy Agency, 2001, IAEA TECHN REP SER, V403
[6]  
International Atomic Energy Agency, 1990, IAEA TECHN REP SER, V318
[7]  
International Commission on Radiation Units and Measurements, 1998, 57 ICRU, Vos29, DOI 10.1093/jicru/os29.2.Report57
[8]  
International Organization for Standardization, 1996, ISO10647
[9]  
International Organization for Standardization, 1989, ISO8529
[10]   Response of several neutron measuring devices under the fractional change of thermal and fast neutrons [J].
Kim, B. H. ;
Kim, S. I. ;
Kim, J. S. ;
Kim, J. L. ;
Lee, J. I. .
RADIATION MEASUREMENTS, 2010, 45 (10) :1541-1543