Culture of dialysate in suspected CAPD associated peritonitis using the BacT/Alert system

被引:11
作者
Khare, S [1 ]
Yurack, J [1 ]
Toye, B [1 ]
机构
[1] UNIV OTTAWA,OTTAWA GEN HOSP,DEPT PATHOL & LAB MED,DIV MICROBIOL,OTTAWA,ON K1H 8L6,CANADA
关键词
D O I
10.1016/S0732-8893(96)00124-1
中图分类号
R51 [传染病];
学科分类号
100401 ;
摘要
The BacT/Alert blood culture system was evaluated as a method of culturing dialysates by comparing inoculation of culture bottles directly (DBTA) or after centrifugation of 50 mL of dialysate (CBTA) with conventional culture. Of the 122 dialysates cultured, 84 were positive by one of the 3 methods. After eliminating contaminants, DBTA and CBTA detected 84% (59 of 70) and 93% (65 of 70) of the positive cultures, respectively, compared to 77% (54 to 70) for conventional culture. CBTA and DBTA detected 87% (82 of 94) and 73% (68 of 94) of the significant organisms isolated, respectively, compared to 61% (57 of 94) by conventional culture. However, 60% of the contaminants occurred with the CBTA method. When a dialysate was positive by all 3 methods, both BacT/Alert methods detected growth earlier by a mean of almost 19 hours. The BacT/Alert system is a useful alternative method for culturing dialysates with the advantages of an earlier detection of positive cultures and minimal handling for the processing of negative cultures. (C) 1996 Elsevier Science Inc.
引用
收藏
页码:101 / 106
页数:6
相关论文
共 21 条
[1]   TOTAL VOLUME CULTURE TECHNIQUE FOR THE ISOLATION OF MICROORGANISMS FROM CONTINUOUS AMBULATORY PERITONEAL-DIALYSIS PATIENTS WITH PERITONITIS [J].
DAWSON, MS ;
HARFORD, AM ;
GARNER, BK ;
SICA, DA ;
LANDWEHR, DM ;
DALTON, HP .
JOURNAL OF CLINICAL MICROBIOLOGY, 1985, 22 (03) :391-394
[2]   CLINICAL AND MICROBIOLOGICAL EVALUATION OF 4 CULTURE METHODS FOR THE DIAGNOSIS OF PERITONITIS IN PATIENTS ON CONTINUOUS AMBULATORY PERITONEAL-DIALYSIS [J].
DOYLE, PW ;
CRICHTON, EP ;
MATHIAS, RG ;
WERB, R .
JOURNAL OF CLINICAL MICROBIOLOGY, 1989, 27 (06) :1206-1209
[3]   EVALUATION OF THE LYSIS-CENTRIFUGATION SYSTEM FOR CULTURING DIALYSATES FROM CONTINUOUS AMBULATORY PERITONEAL-DIALYSIS PATIENTS WITH PERITONITIS [J].
FORBES, BA ;
FRYMOYER, PA ;
KOPECKY, RT ;
WOJTASZEK, JM ;
PETTIT, DJ .
AMERICAN JOURNAL OF KIDNEY DISEASES, 1988, 11 (02) :176-179
[4]  
GOPAUL D, 1984, CAN J MED TECHNOL, V46, P186
[5]  
HOWE PA, 1991, MED LAB SCI, V48, P114
[6]  
KEANE WF, 1993, PERITON DIALYSIS INT, V13, P14
[7]   LABORATORY DIAGNOSIS OF PERITONITIS IN PATIENTS ON CONTINUOUS AMBULATORY PERITONEAL-DIALYSIS [J].
LUDLAM, HA ;
PRICE, TNC ;
BERRY, AJ ;
PHILLIPS, I .
JOURNAL OF CLINICAL MICROBIOLOGY, 1988, 26 (09) :1757-1762
[8]  
Lye W C, 1994, Adv Perit Dial, V10, P166
[9]  
LYON R, 1995, AM J CLIN PATHOL, V103, P660
[10]   ADDI-CHEK FILTRATION, BACTEC, AND 10-ML CULTURE METHODS FOR RECOVERY OF MICROORGANISMS FROM DIALYSIS EFFLUENT DURING EPISODES OF PERITONITIS [J].
MALES, BM ;
WALSHE, JJ ;
GARRINGER, L ;
KOSCINSKI, D ;
AMSTERDAM, D .
JOURNAL OF CLINICAL MICROBIOLOGY, 1986, 23 (02) :350-353