Can we improve the dose distribution for single or multi-lumen breast balloons used for Accelerated Partial Breast Irradiation?

被引:5
作者
Skowronek, Janusz [1 ,2 ]
Bieleda, Grzegorz [3 ]
Laski, Piotr [4 ]
Kycler, Witold [4 ]
机构
[1] Greater Poland Canc Ctr, Brachytherapy Dept, PL-61866 Poznan, Poland
[2] Poznan Univ Med Sci, Electroradiol Dept, Poznan, Poland
[3] Greater Poland Canc Ctr, Dept Med Phys, Poznan, Poland
[4] Greater Poland Canc Ctr, Surg Dept 2, Poznan, Poland
关键词
APBI; brachytherapy; breast cancer; Contura; IPSA; BRACHYTHERAPY CATHETER; INTERSTITIAL BRACHYTHERAPY; DOSIMETRIC FINDINGS; CONSERVING SURGERY; MAMMOSITE; TOXICITY; TRIAL; APBI;
D O I
10.5114/jcb.2013.37776
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Purpose: The aim of the study was to verify dose distribution parameters for multi-lumen, and artificially created single-lumen balloon applicator used for the same patient with two optimization algorithms: inverse planning simulated annealing (IPSA) and dose point optimization with distance option. Material and methods: Group of 24 patients with multi-lumen balloon applied were investigated. Each patient received 10 fractions of 3.4 Gy (2 fractions daily). For every patient, four treatment plans were prepared. Firstly, for five-lumen balloon optimized with IPSA algorithm and optimization parameters adjusted for each case. Secondly, for the same applicator optimized with dose point optimization and distant option. Two other plans were prepared for single-lumen applicator, created by removing four peripheral lumens, optimized with both algorithms. Results: The highest D-95 parameter was obtained for plans optimized with IPSA algorithm, mean value 99.3 percent of prescribed dose, and it was significantly higher than plans optimized with dose point algorithm (mean = 83.50%, p < 0.0001), IPSA single-lumen balloon plan (mean = 83.50%, p = 0.0037) and optimized to dose point single-lumen balloon (mean = 85.51%, p < 0.0001). There were no statistically significant differences concerning maximum doses distributed to skin surface for neither application nor optimization method. Volumes receiving 200% of prescribed dose in PTV were higher for multi-lumen balloon dose point optimized plans (mean = 8.78%), than for other plans (IPSA multi-lumen balloon plan: mean = 7.37%, p < 0.0001, single-lumen IPSA: mean = 7.20%, p < 0.0001, single-lumen dose point: mean = 7.19%, p < 0.0001). Conclusions: Basing on performed survey, better dose distribution parameters are obtained for patients with multi-lumen balloon applied and optimized using IPSA algorithm with individualized optimization parameters.
引用
收藏
页码:134 / 138
页数:5
相关论文
共 25 条
[1]   Contura Multi-Lumen Balloon Breast Brachytherapy Catheter: Comparative Dosimetric Findings of a Phase 4 Trial [J].
Arthur, Douglas W. ;
Vicini, Frank A. ;
Todor, Dorin A. ;
Julian, Thomas B. ;
Cuttino, Laurie W. ;
Mukhopadhyay, Nitai D. .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2013, 86 (02) :264-269
[2]   IMPROVEMENTS IN CRITICAL DOSIMETRIC ENDPOINTS USING THE CONTURA MULTILUMEN BALLOON BREAST BRACHYTHERAPY CATHETER TO DELIVER ACCELERATED PARTIAL BREAST IRRADIATION: PRELIMINARY DOSIMETRIC FINDINGS OF A PHASE IV TRIAL [J].
Arthur, Douglas W. ;
Vicini, Frank A. ;
Todor, Dorin A. ;
Julian, Thomas B. ;
Lyden, Maureen R. .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2011, 79 (01) :26-33
[3]   A dosimetric comparison of the Contura multilumen balloon breast brachytherapy catheter vs. the single-lumen MammoSite balloon device in patients treated with accelerated partial breast irradiation at a single institution [J].
Brown, Sheree ;
McLaughlin, Mark ;
Pope, Doyle Keith ;
Haile, Kenneth ;
Hughes, Lone ;
Israel, Philip Z. ;
Lyden, Maureen .
BRACHYTHERAPY, 2011, 10 (01) :68-73
[4]   Patient selection for accelerated partial-breast irradiation (APBI) after breast-conserving surgery: Recommendations of the Groupe Europeen de Curietherapie-European Society for Therapeutic Radiology and Oncology (GEC-ESTRO) breast cancer working group based on clinical evidence (2009) [J].
Csaba Polgar ;
Van Limbergen, Erik ;
Poetter, Richard ;
Kovacs, Gyoergy ;
Polo, Alfredo ;
Lyczek, Jaroslaw ;
Hildebrandt, Guido ;
Niehoff, Peter ;
Luis Guinot, Jose ;
Guedea, Ferran ;
Johansson, Bengt ;
Ott, Oliver J. ;
Major, Tibor ;
Strnad, Vratislav .
RADIOTHERAPY AND ONCOLOGY, 2010, 94 (03) :264-273
[5]   A COMPARISON OF SKIN AND CHEST WALL DOSE DELIVERED WITH MULTICATHETER, CONTURA MULTILUMEN BALLOON, AND MAMMOSITE BREAST BRACHYTHERAPY [J].
Cuttino, Laurie W. ;
Todor, Dorin ;
Rosu, Mihaela ;
Arthur, Douglas W. .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2011, 79 (01) :34-38
[6]   Optimization and comparison of balloon-based partial breast brachytherapy using a single source, a standard plan line source, and both forward and inverse planned multilumen techniques [J].
Eyre, Katie ;
Whitney, Diane ;
Mukesh, Mukesh ;
Wilson, Charles ;
Coles, Charlotte .
BRACHYTHERAPY, 2013, 12 (02) :107-113
[7]   LONG-TERM OUTCOME FOLLOWING BREAST-CONSERVING SURGERY AND RADIATION-THERAPY [J].
GAGE, I ;
RECHT, A ;
GELMAN, R ;
NIXON, AJ ;
SILVER, B ;
BORNSTEIN, BA ;
HARRIS, JR .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1995, 33 (02) :245-251
[8]   Three-year clinical outcome using the Contura multilumen balloon breast brachytherapy catheter to deliver accelerated partial breast irradiation (APBI): Improving radiation standards for the optimal application of APBI [J].
Israel, Philip Z. ;
Robbins, Angela ;
Shroff, Paulomi ;
Brown, Sheree ;
McLaughlin, Mark ;
Pope, Keith .
BRACHYTHERAPY, 2012, 11 (04) :316-321
[9]  
Israel PZ, 2009, AM SURGEON, V75, P1042
[10]  
Kini Vijay, 2005, Brachytherapy, V4, P181, DOI 10.1016/j.brachy.2005.07.003