eLine100: A Front End ASIC for LCLS Detectors in Low Noise Applications

被引:0
作者
Dragone, A. [1 ]
Caragiulo, P. [1 ]
Freytag, D. [1 ]
Hart, P. [1 ]
Herbst, R. [1 ]
Herrmann, S. [1 ]
Kenney, C. [1 ]
Segal, J. [1 ]
Haller, G. [1 ]
机构
[1] SLAC Natl Accelerator Lab, Res Engn Div, Menlo Pk, CA 94025 USA
关键词
ASIC; detectors; integrating; LCLS; X-rays;
D O I
10.1109/TNS.2014.2301092
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
eLine100 is a fast-frame 96-channel readout ASIC for SLAC Linac Coherent Light Source (LCLS) detectors. The circuit has been designed to integrate the charge from high-capacitance 2D sensors with rolling shutter and 1D sensors. It has a noise floor of 55 e(-) + 8 e(-)/pF r.m.s. at room temperature and it is suitable for applications requiring resolutions on the order of 100 e(-) r.m.s. and signals up to 120 photons/pixel/pulse at 8 keV (260 ke(-)). 2D sensors with a rolling shutter like the X-ray Charge Pump Sensor (XCPS), for which the ASIC has been optimized, present many pixels which are bussed on the same readout line. This characteristic, together with the fixed LCLS beam period, impose limitations on the time available for the read out of each pixel. Given the periodic structure of the LCLS beam, the ASIC developed for this application is a time-variant system, providing a two-stage low-noise charge integration, filtering, correlated double sampling and a processing speed of up to 250 k pixel/s on each channel. To cope with the required input range, a charge pump scheme has been implemented using an asynchronous zero-balance measurement method. It provides on-chip 1-bit coarse analog-to-digital conversion of the integrated charge. The residual charge is sampled using correlated double sampling into an analog memory, multiplexed and measured with the required resolution by an external ADC. In this paper, the ASIC architecture and performances are presented.
引用
收藏
页码:1001 / 1006
页数:6
相关论文
共 11 条
[1]  
Caragiulo P, 2012, IEEE NUCL SCI CONF R, P892
[2]  
Carini GA, 2012, IEEE NUCL SCI CONF R, P515
[3]  
Carini G. A., 2003, NUCL INSTRUM METH A, VA512, P179
[4]  
Dragone A, 2012, IEEE NUCL SCI CONF R, P898
[5]  
Dragone A., 2008, NUCL SCI S NSS MIC, P2970
[6]   Active-edge planar radiation sensors [J].
Kenney, C. J. ;
Segal, J. D. ;
Westbrook, E. ;
Parker, Sherwood ;
Hasi, J. ;
Da Via, C. ;
Watts, S. ;
Morse, J. .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2006, 565 (01) :272-277
[7]  
Nakamura J, 2006, IMAGE SENSORS AND SIGNAL PROCESSING FOR DIGITAL STILL CAMERAS, P315
[8]   Theoretical comparison between two different filtering techniques suitable for the VLSI spectroscopic amplifier ROTOR [J].
Porro, M ;
Fiorini, C ;
Strüder, L .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2003, 512 (1-2) :179-190
[9]   LOW-NOISE TECHNIQUES IN DETECTORS [J].
RADEKA, V .
ANNUAL REVIEW OF NUCLEAR AND PARTICLE SCIENCE, 1988, 38 :217-277
[10]   TRAPEZOIDAL FILTERING OF SIGNALS FROM LARGE GERMANIUM DETECTORS AT HIGH RATES [J].
RADEKA, V .
NUCLEAR INSTRUMENTS & METHODS, 1972, 99 (03) :525-+