Fast estimation of scatter components using the ordered subset expectation maximization algorithm for scatter compensation

被引:5
作者
Liu, Z [1 ]
Obi, T [1 ]
Yamaguchi, M [1 ]
Ohyama, N [1 ]
机构
[1] Tokyo Inst Technol, Imaging Sci & Engn Lab, Midori Ku, Yokohama, Kanagawa 2268503, Japan
关键词
FESC method; scatter estimation; scatter compensation; OS-EM algorithm;
D O I
10.1007/s10043-999-0415-7
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
In this work we propose a method for scatter compensation in single photon emission computed tomography imaging, by which we can estimate the scatter components in projections in high speed with good accuracy. The method is that we first estimate the scatter components in projections based on scatter response kernels by one time of ordered subsets expectation maximization iterative reconstruction, and then subtract the estimated scatter components from the projections and complete reconstruction by filtered back-projection method. The principle is that the image corresponding to the scatter components in projections consists largely of low-frequency components of an activity distribution; these low-frequency components will converge faster than the high ones in iterative reconstruction. Therefore, we can estimate the low-frequency component image before the image converges with the high-frequency ones, and obtain the scatter components by re-projecting the low-frequency component image with scatter response kernels. The effects of the proposed method were compared with the dual- and triple-energy window methods using experimental measurements. The results show that good accuracy in estimated scatter components, good uniformity of scatter compensation at the center and the side of an object, and good noise property can be acquired by this method.
引用
收藏
页码:415 / 423
页数:9
相关论文
共 25 条
[1]  
AXELSSON B, 1984, J NUCL MED, V25, P490
[2]  
BECK JW, 1982, IEEE T NUCL SCI, V29, P506, DOI 10.1109/TNS.1982.4335896
[4]   ATTENUATION CORRECTION AND INCOMPLETE PROJECTION IN SINGLE PHOTON-EMISSION COMPUTED-TOMOGRAPHY [J].
CHANG, LT .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1979, 26 (02) :2780-2789
[5]   MAXIMUM LIKELIHOOD FROM INCOMPLETE DATA VIA EM ALGORITHM [J].
DEMPSTER, AP ;
LAIRD, NM ;
RUBIN, DB .
JOURNAL OF THE ROYAL STATISTICAL SOCIETY SERIES B-METHODOLOGICAL, 1977, 39 (01) :1-38
[6]   SCATTER DETECTION IN SPECT IMAGING - DEPENDENCE ON SOURCE DEPTH, ENERGY, AND ENERGY WINDOW [J].
FLOYD, CE ;
JASZCZAK, RJ ;
COLEMAN, RE .
PHYSICS IN MEDICINE AND BIOLOGY, 1988, 33 (09) :1075-1081
[7]   ENERGY AND SPATIAL-DISTRIBUTION OF MULTIPLE ORDER COMPTON SCATTER IN SPECT - A MONTE-CARLO INVESTIGATION [J].
FLOYD, CE ;
JASZCZAK, RJ ;
HARRIS, CC ;
COLEMAN, RE .
PHYSICS IN MEDICINE AND BIOLOGY, 1984, 29 (10) :1217-1230
[8]   REVISED SCATTER FRACTION RESULTS FOR SPECT [J].
FLOYD, CE ;
JASZCZAK, RJ ;
HARRIS, CC ;
COLEMAN, RE .
PHYSICS IN MEDICINE AND BIOLOGY, 1987, 32 (12) :1663-1666
[9]  
GILARDI MC, 1988, J NUCL MED, V29, P1971
[10]   ACCELERATED IMAGE-RECONSTRUCTION USING ORDERED SUBSETS OF PROJECTION DATA [J].
HUDSON, HM ;
LARKIN, RS .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 1994, 13 (04) :601-609