Optimal Design of Wireless Power Transmission Links for Millimeter-Sized Biomedical Implants

被引:182
作者
Ahn, Dukju [1 ]
Ghovanloo, Maysam [2 ]
机构
[1] Univ Calif San Diego, La Jolla, CA 92093 USA
[2] Georgia Inst Technol, Sch Elect & Comp Engn, GT Bion Lab, Atlanta, GA 30308 USA
基金
美国国家科学基金会;
关键词
Inductive link; mm-sized distributed implants; optimal operating frequency; wireless power transmission; PRINTED SPIRAL COILS; OPTIMIZATION; FREQUENCY; SYSTEMS; TRANSMITTER; TISSUE;
D O I
10.1109/TBCAS.2014.2370794
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
This paper presents a design methodology for RF power transmission to millimeter-sized implantable biomedical devices. The optimal operating frequency and coil geometries are found such that power transfer efficiency (PTE) and tissue-loss-constrained allowed power are maximized. We define receiver power reception susceptibility (Rx-PRS) and transmitter figure of merit (Tx-FoM) such that their multiplication yields the PTE. Rx-PRS and Tx-FoM define the roles of the Rx and Tx in the PTE, respectively. First, the optimal Rx coil geometry and operating frequency range are identified such that the Rx-PRS is maximized for given implant constraints. Since the Rx is very small and has lesser design freedom than the Tx, the overall operating frequency is restricted mainly by the Rx. Rx-PRS identifies such operating frequency constraint imposed by the Rx. Secondly, the Tx coil geometry is selected such that the Tx-FoM is maximized under the frequency constraint at which the Rx-PRS was saturated. This aligns the target frequency range of Tx optimization with the frequency range at which Rx performance is high, resulting in the maximum PTE. Finally, we have found that even in the frequency range at which the PTE is relatively flat, the tissue loss per unit delivered power can be significantly different for each frequency. The Rx-PRS can predict the frequency range at which the tissue loss per unit delivered power is minimized while PTE is maintained high. In this way, frequency adjustment for the PTE and tissue-loss-constrained allowed power is realized by characterizing the Rx-PRS. The design procedure was verified through full-wave electromagnetic field simulations and measurements using de-embedding method. A prototype implant, 1 mm in diameter, achieved PTE of 0.56% (22.5 dB) and power delivered to load (PDL) was 224 mu W at 200 MHz with 12 mm Tx-to-Rx separation in the tissue environment.
引用
收藏
页码:125 / 137
页数:13
相关论文
共 36 条
[1]   Wireless Power Transmission With Self-Regulated Output Voltage for Biomedical Implant [J].
Ahn, Dukju ;
Hong, Songcheol .
IEEE TRANSACTIONS ON INDUSTRIAL ELECTRONICS, 2014, 61 (05) :2225-2235
[2]   Effect of Coupling Between Multiple Transmitters or Multiple Receivers on Wireless Power Transfer [J].
Ahn, Dukju ;
Hong, Songcheol .
IEEE TRANSACTIONS ON INDUSTRIAL ELECTRONICS, 2013, 60 (07) :2602-2613
[3]  
[Anonymous], 2011, 2011 DESIGN AUTOMATI, DOI DOI 10.1109/DATE.2011.5763123
[4]  
[Anonymous], C9532002R2008 IEEE
[5]   Feedback Analysis and Design of RF Power Links for Low-Power Bionic Systems [J].
Baker, Michael W. ;
Sarpeshkar, Rahul .
IEEE TRANSACTIONS ON BIOMEDICAL CIRCUITS AND SYSTEMS, 2007, 1 (01) :28-38
[6]   Wireless Implants [J].
Bashirullah, Rizwan .
IEEE MICROWAVE MAGAZINE, 2010, 11 (07) :S14-S23
[7]   A Fully-Integrated, Miniaturized (0.125 mm2) 10.5 μW Wireless Neural Sensor [J].
Biederman, William ;
Yeager, Daniel J. ;
Narevsky, Nathan ;
Koralek, Aaron C. ;
Carmena, Jose M. ;
Alon, Elad ;
Rabaey, Jan M. .
IEEE JOURNAL OF SOLID-STATE CIRCUITS, 2013, 48 (04) :960-970
[8]   A 3-STEP METHOD FOR THE DE-EMBEDDING OF HIGH-FREQUENCY S-PARAMETER MEASUREMENTS [J].
CHO, HJ ;
BURK, DE .
IEEE TRANSACTIONS ON ELECTRON DEVICES, 1991, 38 (06) :1371-1375
[9]   A Transponder-Based, Nonradiative Wireless Power Transfer [J].
Ettorre, Mauro ;
Grbic, Anthony .
IEEE ANTENNAS AND WIRELESS PROPAGATION LETTERS, 2012, 11 :1150-1153
[10]  
Finkenzeller K., 2010, RFID HDB, P66