Accuracy of Template-Based Dental Implant Placement

被引:0
作者
Eggers, Georg [1 ]
Patellis, Evangelos [1 ]
Muehling, Joachim [1 ]
机构
[1] Univ Heidelberg Hosp, Dept Oral & Craniomaxillofacial Surg, D-69120 Heidelberg, Germany
关键词
accuracy; dental implants; image-guided surgery; oral templates; stereotaxy; IMAGE-GUIDED IMPLANTOLOGY; ORAL IMPLANTS; IN-VITRO; TOMOGRAPHY; SURGERY; REGISTRATION; PRECISION;
D O I
暂无
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Purpose: The accuracy of a system for image-guided dental implant planning and placement based on individually fabricated oral templates was investigated in a phantom study. Materials and Methods: The study was performed on identical phantoms of a partially edentulous mandible. In one phantom mandible (the master model), pilot boreholes for dental implants were placed. Computerized tomography image data were then acquired. Using a system for template-based dental implant placement, identical boreholes were planned on the image data and then reproduced in the 10 remaining phantom mandibles (the "slave" models). The resulting positions were compared to the positions in the master model using a three-dimensional digitizer arm. Results: The mean deviation of borehole apex positions in the slave phantoms, as compared to the master phantom, was 0.50 +/- 0.04 mm and was below 1 mm in 29 of 30 models. The accuracy was highly reproducible, irrespective of the template used and the surgeon involved. Conclusions: The use of templates for image-guided insertion of dental implants is a method with high and reproducible accuracy. INT J ORAL MAXILLOFAC IMPLANTS 2009; 24:447-454
引用
收藏
页码:447 / 454
页数:8
相关论文
共 28 条
[1]   Accuracy of image-guided implantology [J].
Brief, J ;
Edinger, D ;
Hassfeld, S ;
Eggers, G .
CLINICAL ORAL IMPLANTS RESEARCH, 2005, 16 (04) :495-501
[2]   Navigation surgery for dental implants: Assessment of accuracy of the image guided implantology system [J].
Casap, N ;
Wexler, A ;
Persky, N ;
Schneider, A ;
Lustmann, J .
JOURNAL OF ORAL AND MAXILLOFACIAL SURGERY, 2004, 62 (09) :116-119
[3]  
Di Giacomo GAP, 2005, J PERIODONTOL, V76, P503
[4]   Image-to-patient registration techniques in head surgery [J].
Eggers, G. ;
Muehling, J. ;
Marmulla, R. .
INTERNATIONAL JOURNAL OF ORAL AND MAXILLOFACIAL SURGERY, 2006, 35 (12) :1081-1095
[5]  
Eufinger H, 1997, Clin Oral Investig, V1, P169
[6]   Precision of transfer of preoperative planning for oral implants based on cone-beam CT-scan images through a robotic drilling machine - An in vitro study [J].
Fortin, T ;
Champleboux, G ;
Bianchi, S ;
Buatois, H ;
Coudert, FL .
CLINICAL ORAL IMPLANTS RESEARCH, 2002, 13 (06) :651-656
[7]   Low-dose-Computertomographie des Kieferknochens in der präimplantologischen DiagnostikGrenzen der Dosisreduzierung und Genauigkeit von LängenmessungenLow-dose CT of the mandible and maxilla for dental implant treatment diagnosis: dose reduction limits and distance measurement precision [J].
S. Haßfeld ;
S. Streib ;
H. Sahl ;
U. Stratmann ;
D. Fehrentz ;
J. Zöller .
Mund-, Kiefer- und Gesichtschirurgie, 1998, 2 (4) :188-193
[8]  
Heurich Th, 2002, Biomed Tech (Berl), V47 Suppl 1 Pt 1, P25, DOI 10.1515/bmte.2002.47.s1a.25
[9]  
Hoffmann J, 2005, INT J ORAL MAX IMPL, V20, P382
[10]   Precision for computer-guided implant placement: Using 3D planning software and fixed intraoral reference points [J].
Holst, Stefan ;
Blatz, Markus B. ;
Eitner, Stephan .
JOURNAL OF ORAL AND MAXILLOFACIAL SURGERY, 2007, 65 (03) :393-399