Design of a real-time laser stimulus system based on MEMS-based confocal laser scanning microscope

被引:0
|
作者
Jin, Xingjie [1 ,2 ]
Xu, Xin [1 ]
Gao, Feng [1 ]
Kong, Wen [1 ]
Shi, Guohua [1 ]
机构
[1] Chinese Acad Sci, Jiangsu Key Lab Med Opt, Suzhou Inst Biomed Engn & Technol, Suzhou 215163, Peoples R China
[2] Shanghai Univ, Sch Commun & Informat Engn, Shanghai 200444, Peoples R China
来源
FIFTH SYMPOSIUM ON NOVEL OPTOELECTRONIC DETECTION TECHNOLOGY AND APPLICATION | 2019年 / 11023卷
关键词
Light stimulus; confocal microscopy; MEMS-mirror; FPGA;
D O I
10.1117/12.2521544
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
We design a real-time laser stimulus system for laser confocal scanning microscope. By introducing the FPGA and AOM to achieve high speed modulation of a scanning laser, we can adjust the laser lighting area freely. For reducing the size of the optical path, we use MEMS-mirror instead of traditional fast and slow axis mirrors. The size of MEMS-mirror is 1.5 mm diameter and the scanning frequencies are set 16 kHz and 12 Hz at the fast and slow axis, respectively. Our system is capable of delivering stabilized large stimulus pattern (up to 500 x 500 pixels) to the biological tissues.
引用
收藏
页数:4
相关论文
共 50 条
  • [1] Large field of view MEMS-based confocal laser scanning microscope for fluorescence imaging
    Bechtel, Christin
    Knobbe, Jens
    Grueger, Heinrich
    Lakner, Hubert
    OPTIK, 2014, 125 (02): : 876 - 882
  • [2] MEMS-based confocal laser scanning fluorescence microscopy for tumor demarcation in oncological surgery
    Reinig, Peter
    Dallmann, Hans-Georg
    Schwarzenberg, Markus
    Ziebarth, Johannes
    Knobbe, Jens
    Junek, Jiri
    Herbst, Rudolf
    Rathert, Julian
    Gerlach, Ruediger
    Blache, Ulrich
    Tretbar, Sandy
    Fricke, Stephan
    ADVANCED BIOMEDICAL AND CLINICAL DIAGNOSTIC AND SURGICAL GUIDANCE SYSTEMS XX, 2022, 11949
  • [3] 3-D Confocal Laser Scanning Microscopy Based on a Full-MEMS Scanning System
    Liu, Lin
    Xie, Huikai
    IEEE PHOTONICS TECHNOLOGY LETTERS, 2013, 25 (15) : 1478 - 1480
  • [4] MEMS based laser scanning microscope for endoscopic use
    Schelinski, U.
    Knobbe, J.
    Dallmann, H.
    Grueger, H.
    Foerster, M.
    Scholles, M.
    Schwarzenberg, M.
    Rieske, R.
    MOEMS AND MINIATURIZED SYSTEMS X, 2011, 7930
  • [5] REAL-TIME CONFOCAL LASER SCAN MICROSCOPE FOR EXAMINATION AND DIAGNOSIS OF THE EYE IN-VIVO
    MASSIG, JH
    PREISSLER, M
    WEGENER, AR
    GAIDA, G
    APPLIED OPTICS, 1994, 33 (04): : 690 - 694
  • [6] Real-Time Resolution Enhancement of Confocal Laser Scanning Microscopy via Deep Learning
    Cui, Zhiying
    Xing, Yi
    Chen, Yunbo
    Zheng, Xiu
    Liu, Wenjie
    Kuang, Cuifang
    Chen, Youhua
    PHOTONICS, 2024, 11 (10)
  • [7] Coaxial Scanning Real-Time Light Stimulus System
    Jin Xingjie
    Gao Feng
    Xu Xin
    Kong Wen
    Yan Zhuangzhi
    Shi Guohua
    ACTA OPTICA SINICA, 2019, 39 (03)
  • [8] Coaxial Scanning Real-Time Light Stimulus System
    Jin X.
    Gao F.
    Xu X.
    Kong W.
    Yan Z.
    Shi G.
    Guangxue Xuebao/Acta Optica Sinica, 2019, 39 (03):
  • [9] Application of real-time confocal laser scanning microscopy to observe living cells in tissue specimens
    Saino, T
    Satoh, Y
    JOURNAL OF ELECTRON MICROSCOPY, 2004, 53 (01): : 49 - 56
  • [10] Design of Galvanometer Laser Scanning System Based on FPGA
    Lv, Qingying
    Sun, Yu
    MACHINE DESIGN AND MANUFACTURING ENGINEERING III, 2014, : 246 - 249