Clinical evaluation of the eBsensor hand-held blood glucose monitoring system

被引:3
作者
Chen, Chien-Chih
Tai, Der-Yan
Ho, Ping
Lai, Yiu-Kay
Lin, Jui-Jane
Mao, Simon J. T.
Lin, Chia-Chi
Hsieh, Yu-Chen
Chiang, Hsiang-Cheng
Dai, Ken-Shwo
机构
[1] Visgeneer Inc, Hsinchu 300, Taiwan
[2] Wei Gong Mem Hosp, Miaoli, Taiwan
[3] Natl Tsing Hua Univ, Coll Life Sci, Inst Biotechnol, Hsinchu, Taiwan
[4] Natl Chiao Tung Univ, Dept Biol Sci & Technol, Hsinchu, Taiwan
关键词
diabetes; glucose monitoring system; Clarke error grid analysis; glucose meter; strip; self-monitoring; accuracy; precision; electrochemical;
D O I
10.1016/j.cca.2006.09.023
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Background: Home glucose monitoring system is increasingly recognized as an important tool for glycemic control. We evaluated the clinical performance of the eBsensor glucose monitoring system. Methods: Fingertip capillary blood glucose concentrations from 282 subjects were measured using eBsensor glucose monitoring system and compared against predicate devices and the Yellow Springs Instruments (YSI) 2300 blood glucose analyzer. Accuracy and precision of the eBsensor glucose monitoring system were assessed using several methods. The comparative study between the eBsensor and 2 currently marketed monitoring systems was performed. Results: The 282 eBsensor readings covered a wide range from 2.6 to 24.4 mmol/l. Deming regression and Pearson correlation analyses showed a linear relationship between the eBsensor readings and the YSI reference method (eBsensor=0.9496 YSI+0.4127 mmol/l; r=0.98). Error Grid analysis demonstrated that 100% of the eBsensor readings in clinically acceptable zones A and B. The CVs for the 6 lots of strips were within the satisfactory interval (< 6%). The comparative study showed that the eBsensor readings correlated well with the OneTouch Ultra values (r=0.97) and the Glucocard II values (r=0.97). Conclusions: eBsensor is a reliable glucose monitoring system which provides high accurate and precise glucose readings over a wide range of glucose concentrations. (c) 2006 Elsevier B.V. All rights reserved.
引用
收藏
页码:170 / 173
页数:4
相关论文
共 24 条
  • [1] Alto William A, 2002, J Am Board Fam Pract, V15, P1
  • [2] *AM DIAB ASS, 2004, DIABETES CARE S1, V27, pS15
  • [3] [Anonymous], 1996, DIABETES CARE, V19, pS62
  • [4] [Anonymous], 2003, GLOB STRAT DIET PHYS
  • [5] Validation of home blood glucose meters with respect to clinical and analytical approaches
    Brunner, GA
    Ellmerer, M
    Sendlhofer, G
    Wutte, A
    Trajanoski, Z
    Schaupp, L
    Quehenberger, F
    Wach, P
    Krejs, GJ
    Pieber, TR
    [J]. DIABETES CARE, 1998, 21 (04) : 585 - 590
  • [6] EVALUATING CLINICAL ACCURACY OF SYSTEMS FOR SELF-MONITORING OF BLOOD-GLUCOSE
    CLARKE, WL
    COX, D
    GONDERFREDERICK, LA
    CARTER, W
    POHL, SL
    [J]. DIABETES CARE, 1987, 10 (05) : 622 - 628
  • [7] A comparison of blood glucose meters in Australia
    Cohen, M
    Boyle, E
    Delaney, C
    Shaw, J
    [J]. DIABETES RESEARCH AND CLINICAL PRACTICE, 2006, 71 (02) : 113 - 118
  • [8] ACCURACY OF PERCEIVING BLOOD-GLUCOSE IN IDDM
    COX, DJ
    CLARKE, WL
    GONDERFREDERICK, L
    POHL, S
    HOOVER, C
    SNYDER, A
    ZIMBELMAN, L
    CARTER, WR
    BOBBITT, S
    PENNEBAKER, J
    [J]. DIABETES CARE, 1985, 8 (06) : 529 - 536
  • [9] Accuracy of the EasyTouch blood glucose self-monitoring system: a study of 516 cases
    Dai, KS
    Tai, DY
    Ho, P
    Chen, CC
    Peng, WC
    Chen, ST
    Hsu, CC
    Liu, YP
    Hsieh, HC
    Yang, CC
    Tsai, MC
    Mao, SJT
    [J]. CLINICA CHIMICA ACTA, 2004, 349 (1-2) : 135 - 141
  • [10] The impact of blood glucose self-monitoring on metabolic control and quality of life in type 2 diabetic patients - An urgent need for better educational strategies
    Franciosi, M
    Pellegrini, F
    De Berardis, G
    Belfiglio, M
    Cavaliere, D
    Di Nardo, B
    Greenfield, S
    Kaplan, SH
    Sacco, M
    Tognoni, G
    Valentini, M
    Nicolucci, A
    [J]. DIABETES CARE, 2001, 24 (11) : 1870 - 1877