An Analytical Model for Inductively Coupled Implantable Biomedical Devices With Ferrite Rods

被引:53
作者
Theilmann, Paul T. [1 ]
Asbeck, Peter M. [1 ]
机构
[1] Univ Calif San Diego, Dept Elect & Comp Engn, La Jolla, CA 92093 USA
关键词
Biomedical power supplies; biomedical telemetry; ferrites; implantable biomedical devices; mutual coupling; DEMAGNETIZING FACTORS; SYSTEM; DESIGN; COILS; CORE;
D O I
10.1109/TBCAS.2008.2004776
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Using approximations applicable to near field coupled implants simplified expressions for the complex mutual inductance of coaxial aligned coils with and without a cylindrical ferrite rod are derived. Experimental results for ferrite rods of various sizes and permeabilities are presented to verify the accuracy of this expression. An equivalent circuit model for the inductive link between an implant and power coil is then presented and used to investigate how ferrite size, permeability and loss affect the power available to the implant device. Enhancements in coupling provided by high frequency, low permeability nickel zinc rods are compared with low frequency high permeability manganese zinc rods.
引用
收藏
页码:43 / 52
页数:10
相关论文
共 50 条
  • [41] STUDY OF NONRECIPROCAL DEVICES USING THREE-STRIP FERRITE COUPLED LINE
    Marynowski, W.
    Mazur, J.
    PROGRESS IN ELECTROMAGNETICS RESEARCH-PIER, 2011, 118 : 487 - 504
  • [42] Ferrite Concentrating and Shielding Structure Design of Wireless Power Transmitting Coil for Inductively Coupled Capsule Robot
    Zhuang, Haoyu
    Wang, Wei
    Yan, Guozheng
    IEEE TRANSACTIONS ON BIOMEDICAL CIRCUITS AND SYSTEMS, 2023, 17 (01) : 45 - 53
  • [43] Flexible Metasurface-Coupled Efficient Wireless Power Transfer System for Implantable Devices
    Shah, Izaz Ali
    Zada, Muhammad
    Shah, Syed Ahson Ali
    Basir, Abdul
    Yoo, Hyoungsuk
    IEEE TRANSACTIONS ON MICROWAVE THEORY AND TECHNIQUES, 2024, 72 (04) : 2534 - 2547
  • [44] A Low-Power 13.56 MHz RF Front-End Circuit for Implantable Biomedical Devices
    Lee, Shuenn-Yuh
    Hong, Jia-Hua
    Hsieh, Cheng-Han
    Liang, Ming-Chun
    Kung, Jing-Yang
    IEEE TRANSACTIONS ON BIOMEDICAL CIRCUITS AND SYSTEMS, 2013, 7 (03) : 256 - 265
  • [45] Analytical performance of the Conical torch in axially viewed inductively coupled plasma optical emission spectroscopy
    Guo, Xiaoman
    Alavi, Sina
    Mostaghimi, Javad
    JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2019, 34 (10) : 2126 - 2135
  • [46] Analysis of reflected load model for inductively coupled power transfer systems
    Yang Minsheng
    Wang Yaonan
    Zhang Xizheng
    Li Jianqi
    2010 ASIA-PACIFIC POWER AND ENERGY ENGINEERING CONFERENCE (APPEEC), 2010,
  • [47] Evaluation of various commercially available nebulization devices for inductively coupled plasma atomic emission spectrometry
    Hoenig, M
    Baeten, H
    Vanhentenrijk, S
    Ploegaerts, G
    Bertholet, T
    ANALUSIS, 1997, 25 (01) : 13 - 19
  • [48] Ferrite-Enhanced U-Shaped Internal Antenna for Large-Area Inductively Coupled Plasma System
    Kim, Kyong Nam
    Lim, Jong Hyeuk
    Lim, Woong Sun
    Yeom, Geun Young
    IEEE TRANSACTIONS ON PLASMA SCIENCE, 2010, 38 (02) : 133 - 136
  • [49] Evaluation of Specific Absorption Rate in Three-Layered Tissue Model at 13.56 MHz and 40.68 MHz for Inductively Powered Biomedical Implants
    Mohanarangam, Krithikaa
    Palagani, Yellappa
    Choi, Jun Rim
    APPLIED SCIENCES-BASEL, 2019, 9 (06):
  • [50] A 40.68-MHz Active Rectifier With Hybrid Adaptive On/Off Delay-Compensation Scheme for Biomedical Implantable Devices
    Cheng, Lin
    Ge, Xinyuan
    Hu, Langyu
    Yao, Yuan
    Ki, Wing-Hung
    Tsui, Chi-Ying
    IEEE TRANSACTIONS ON CIRCUITS AND SYSTEMS I-REGULAR PAPERS, 2020, 67 (02) : 516 - 525