An Analytical Model for Inductively Coupled Implantable Biomedical Devices With Ferrite Rods

被引:53
作者
Theilmann, Paul T. [1 ]
Asbeck, Peter M. [1 ]
机构
[1] Univ Calif San Diego, Dept Elect & Comp Engn, La Jolla, CA 92093 USA
关键词
Biomedical power supplies; biomedical telemetry; ferrites; implantable biomedical devices; mutual coupling; DEMAGNETIZING FACTORS; SYSTEM; DESIGN; COILS; CORE;
D O I
10.1109/TBCAS.2008.2004776
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Using approximations applicable to near field coupled implants simplified expressions for the complex mutual inductance of coaxial aligned coils with and without a cylindrical ferrite rod are derived. Experimental results for ferrite rods of various sizes and permeabilities are presented to verify the accuracy of this expression. An equivalent circuit model for the inductive link between an implant and power coil is then presented and used to investigate how ferrite size, permeability and loss affect the power available to the implant device. Enhancements in coupling provided by high frequency, low permeability nickel zinc rods are compared with low frequency high permeability manganese zinc rods.
引用
收藏
页码:43 / 52
页数:10
相关论文
共 50 条
  • [21] Navigating the future of healthcare with innovations and challenges in implantable battery technology for biomedical devices
    Krishnamoorthy, Umapathi
    Lakshmipathy, Priya
    Ramya, Manohar
    Fayek, Hady H.
    DISCOVER APPLIED SCIENCES, 2024, 6 (11)
  • [22] Review on next generation wireless power transmission technology for implantable biomedical devices
    Saranya, N.
    Kesavamurthy, T.
    INTERNATIONAL JOURNAL OF BIOMEDICAL ENGINEERING AND TECHNOLOGY, 2021, 35 (03) : 207 - 222
  • [23] Kinetic and thermal energy harvesters for implantable medical devices and biomedical autonomous sensors
    Cadei, Andrea
    Dionisi, Alessandro
    Sardini, Emilio
    Serpelloni, Mauro
    MEASUREMENT SCIENCE AND TECHNOLOGY, 2014, 25 (01)
  • [24] A High Data Rate Implantable MIMO Antenna for Deep Implanted Biomedical Devices
    Alazemi, Abdullah J.
    Iqbal, Amjad
    IEEE TRANSACTIONS ON ANTENNAS AND PROPAGATION, 2022, 70 (02) : 998 - 1007
  • [25] CMOS Front-End Architecture for In-Vivo Biomedical Implantable Devices
    Colomer-Farrarons, J.
    Miribel-Catala, P.
    Rodriguez, I.
    Samitier, J.
    IECON: 2009 35TH ANNUAL CONFERENCE OF IEEE INDUSTRIAL ELECTRONICS, VOLS 1-6, 2009, : 4188 - +
  • [26] Implantable High-Gain Dental Antennas for Minimally Invasive Biomedical Devices
    Yang, Chin-Lung
    Tsai, Chi-Lin
    Chen, Sheng-Hao
    IEEE TRANSACTIONS ON ANTENNAS AND PROPAGATION, 2013, 61 (05) : 2380 - 2387
  • [27] Review of analytical methods to extract magnetic parameters of an inductively coupled circuit
    Prasanth, Venugopal
    Bauer, Pavol
    Ferreira, Jan Abraham
    Polinder, Henk
    2015 IEEE PELS WORKSHOP ON EMERGING TECHNOLOGIES - WIRELESS POWER (WOW), 2015,
  • [28] Cylindrical ferrite coupled slotline junction for Faraday nonreciprocal devices
    Kusiek, A.
    Marynowski, W.
    Mazur, J.
    2012 19TH INTERNATIONAL CONFERENCE ON MICROWAVE RADAR AND WIRELESS COMMUNICATIONS (MIKON), VOLS 1 AND 2, 2012, : 699 - 702
  • [29] High Voltage Integrated Chip Power Recovering Topology for Implantable Wireless Biomedical Devices
    Nair, Vijith Vijayakumaran
    Youn, Jinhwan
    Choi, Jun Rim
    BIODEVICES: PROCEEDINGS OF THE 2015 INTERNATIONAL CONFERENCE ON BIOMEDICAL ELECTRONICS AND DEVICES, 2015, : 13 - 22
  • [30] Compact Dual-Band Zig Zag Shaped Implantable Antenna for Biomedical Devices
    Nehra, Rajiv Kumar
    Raghava, N. S.
    INDIAN JOURNAL OF PURE & APPLIED PHYSICS, 2022, 60 (10) : 841 - 848