Implementing and Validating Transcutaneous Bilirubinometry for Neonates

被引:17
作者
Holland, Lorne [1 ]
Blick, Kenneth [2 ]
机构
[1] Univ Colorado, Dept Pathol, Aurora, CO 80045 USA
[2] Univ Oklahoma, Hlth Sci Ctr, Oklahoma City, OK USA
关键词
Chemistry; Clinical chemistry; Transcutaneous bilirubin; Bilirubinometry; Neonates; TOTAL SERUM BILIRUBIN; BILICHECK; JAUNDICE; HYPERBILIRUBINEMIA; NEWBORNS; INFANTS; TERM; UNIT;
D O I
10.1309/AJCPN9BMFW8COTWP
中图分类号
R36 [病理学];
学科分类号
100104 ;
摘要
Noninvasive, transcutaneous bilirubin (TcB) measurement is an attractive option for neonates, but opinions about its usefulness vary among studies. We collected paired measurements of TcB and serum bilirubin (SB) in 343 term neonates using the BiliCheck meter (SpectRx, Norcross, GA) and 3 different SB methods. Correlations between SB and TcB were similar for all laboratory methods and TcB measurement sites. However, TcB bias varied depending on the comparison SB method and TcB measurement site. TcB bias also varied with race when measurements were done on the forehead but not when they were done on the sternum. Several factors must be considered before implementing TcB measurement: (1) Each laboratory instrument has its own unique relationship to TcB. (2) The chosen measurement site affects the relationship. (3) Race can affect TcB bias when the measurement is taken on the forehead. Properly used, TcB measurement, especially when taken from? the sternum, can be a useful screening method for neonatal jaundice.
引用
收藏
页码:555 / 561
页数:7
相关论文
共 33 条
[1]  
Beck M, 2003, ARCH DIS CHILD-FETAL, V88, P350
[2]   Predictive ability of a predischarge hour-specific serum bilirubin for subsequent significant hyperbilirubinemia in healthy term and near-term newborns [J].
Bhutani, VK ;
Johnson, L ;
Sivieri, EM .
PEDIATRICS, 1999, 103 (01) :6-14
[3]   Prediction of severe hyperbilirubinaemia using the Bilicheck transcutaneous bilirubinometer [J].
Boo, Nem-Yun ;
Ishak, Shareena .
JOURNAL OF PAEDIATRICS AND CHILD HEALTH, 2007, 43 (04) :306-311
[4]   Comparison of transcutaneous and plasma bilirubin measurement [J].
da Cunha Leite, Maria das Gragas ;
Granato, Valeria de Araujo ;
Facchini, Fernando Peraizini ;
Martins Marba, SergioTadeu .
JORNAL DE PEDIATRIA, 2007, 83 (03) :283-286
[5]   Attempt to improve transcutaneous bilirubinometry: a double-blind study of Medick BiliMed versus Respironics BiliCheck [J].
De Luca, D. ;
Zecca, E. ;
Corsello, M. ;
Tiberi, E. ;
Semeraro, C. ;
Romagnoli, C. .
ARCHIVES OF DISEASE IN CHILDHOOD-FETAL AND NEONATAL EDITION, 2008, 93 (02) :F135-F139
[6]   Skin bilirubin nomogram for the first 96 h of life in a European normal health newborn population, obtained with multiwavelength transcutaneous bilirubinometry [J].
De Luca, Daniele ;
Romagnoli, Costantino ;
Tiberi, Eloisa ;
Zuppa, Antonio Alberto ;
Zecca, Enrico .
ACTA PAEDIATRICA, 2008, 97 (02) :146-150
[7]   Using Bilicheck™ for preterm neonates in a sub-intensive unit:: Diagnostic usefulness and suitability [J].
De Luca, Daniele ;
Zecca, Enrico ;
de Turris, Pierluigi ;
Barbato, Giada ;
Marras, Marco ;
Romagnoli, Costantino .
EARLY HUMAN DEVELOPMENT, 2007, 83 (05) :313-317
[8]   A new transcutaneous bilirubinometer, BiliCheck, used in the neonatal intensive care unit and the maternity ward [J].
Ebbesen, F ;
Rasmussen, LM ;
Wimberley, PD .
ACTA PAEDIATRICA, 2002, 91 (02) :203-211
[9]   Assessment of a transcutaneous device in the evaluation of neonatal hyperbilirubinemia in a primarily Hispanic population [J].
Engle, WD ;
Jackson, GL ;
Sendelbach, D ;
Manning, D ;
Frawley, WH .
PEDIATRICS, 2002, 110 (01) :61-67
[10]  
Ho E. Y. W., 2006, Hong Kong Medical Journal, V12, P99