Efficacy of multi-electrode linear irreversible electroporation

被引:6
作者
Buist, Thomas J. [1 ,2 ]
Groen, Marijn H. A. [2 ]
Wittkampf, Fred H. M. [2 ]
Loh, Peter [2 ]
Doevendans, Pieter A. F. M. [2 ,3 ,4 ]
van Es, Rene [2 ]
Elvan, Arif [1 ]
机构
[1] Isala Hosp, Heart Ctr, Dept Cardiol, Dr Van Heesweg 2, NL-8025 AB Zwolle, Netherlands
[2] Univ Med Ctr Utrecht, Dept Cardiol, Heidelberglaan 100, NL-3584 CX Utrecht, Netherlands
[3] Netherlands Heart Inst, Utrecht, Netherlands
[4] Cent Mil Hosp, Utrecht, Netherlands
来源
EUROPACE | 2021年 / 23卷 / 03期
关键词
Linear multi-electrode; Epicardial; Irreversible; Electroporation; Ablation; ATRIAL-FIBRILLATION; CATHETER ABLATION; FEASIBILITY; CONTACT;
D O I
10.1093/europace/euaa280
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Aims We investigated the efficacy of linear multi-electrode irreversible electroporation (IRE) ablation in a porcine model. Methods and results The study was performed in six pigs (weight 60-75 kg). After median sternotomy and opening of the pericardium, a pericardial cradle was formed and filled with blood. A linear seven polar 7-Fr electrode catheter with 2.5 mm electrodes and 2.5 mm inter-electrode spacing was placed in good contact with epicardial tissue. A single IRE application was delivered using 50 J at one site and 100J at two other sites, in random sequence, using a standard monophasic defibrillator connected to all seven electrodes connected in parallel. The pericardium and thorax were closed and after 3 weeks survival animals were euthanized. A total of 82 histological sections from all 18 electroporation lesions were analysed. A total of seven 50J and fourteen 100J epicardial IRE applications were performed. Mean peak voltages at 50 and 100 J were 1079.2V +/- 81.1 and 1609.5 V +/- 56.8, with a mean peak current of 15.4 A +/- 2.3 and 20.2 A +/- 1.7, respectively. Median depth of the 50 and 100J lesions were 3.2mm [interquartile range (IQR) 3.1-3.6] and 5.5 mm (IQR 4.6-6.6) (P < 0.001), respectively. Median lesion width of the 50 and 100J lesions was 3.9 mm (IQR 3.7-4.8) and 5.4 mm (IQR 5.0-6.3), respectively (P < 0.001). Longitudinal sections showed continuous lesions for 100J applications. Conclusion Epicardial multi-electrode Linear application of IRE pulses is effective in creating continuous deep lesions.
引用
收藏
页码:464 / 468
页数:5
相关论文
共 21 条
[1]   SOME FACTORS AFFECTING BUBBLE FORMATION WITH CATHETER-MEDIATED DEFIBRILLATOR PULSES [J].
BARDY, GH ;
COLTORTI, F ;
IVEY, TD ;
ALFERNESS, C ;
RACKSON, M ;
HANSEN, K ;
STEWART, R ;
GREENE, HL .
CIRCULATION, 1986, 73 (03) :525-538
[2]  
Calkins H, 2017, J ARRYTHM, V33, P369, DOI 10.1016/j.joa.2017.08.001
[3]   2019 HRS/EHRA/APHRS/LAHRS expert consensus statement on catheter ablation of ventricular arrhythmias [J].
Cronin, Edmond M. ;
Bogun, Frank M. ;
Maury, Philippe ;
Peichl, Petr ;
Chen, Minglong ;
Namboodiri, Narayanan ;
Aguinaga, Luis ;
Leite, Luiz Roberto ;
Al-Khatib, Sana M. ;
Anter, Elad ;
Berruezo, Antonio ;
Callans, David J. ;
Chung, Mina K. ;
Cuculich, Phillip ;
d'Avila, Andre ;
Deal, Barbara J. ;
Della Bella, Paolo ;
Deneke, Thomas ;
Dickfeld, Timm-Michael ;
Hadid, Claudio ;
Haqqani, Haris M. ;
Kay, G. Neal ;
Latchamsetty, Rakesh ;
Marchlinski, Francis ;
Miller, John M. ;
Nogami, Akihiko ;
Patel, Akash R. ;
Pathak, Rajeev Kumar ;
Morales, Luis C. Saenz ;
Santangeli, Pasquale ;
Sapp, John L., Jr. ;
Sarkozy, Andrea ;
Soejima, Kyoko ;
Stevenson, William G. ;
Tedrow, Usha B. ;
Tzou, Wendy S. ;
Varma, Niraj ;
Zeppenfeld, Katja .
EUROPACE, 2019, 21 (08) :1143-+
[4]   Minimal coronary artery damage by myocardial electroporation ablation [J].
du Pre, Bastiaan C. ;
van Driel, Vincent J. ;
van Wessel, Harry ;
Loh, Peter ;
Doevendans, Pieter A. ;
Goldschmeding, Roel ;
Wittkampf, Fred H. ;
Vink, Aryan .
EUROPACE, 2013, 15 (01) :144-149
[5]   Effective contact and outcome after pulmonary vein isolation in novel circular multi-electrode atrial fibrillation ablation [J].
Gal, P. ;
Buist, T. J. ;
Smit, J. J. J. ;
Adiyaman, A. ;
Misier, A. R. Ramdat ;
Delnoy, P. P. H. M. ;
Elvan, A. .
NETHERLANDS HEART JOURNAL, 2017, 25 (01) :16-23
[6]   CATHETER TECHNIQUE FOR CLOSED-CHEST ABLATION OF THE ATRIOVENTRICULAR-CONDUCTION SYSTEM - A THERAPEUTIC ALTERNATIVE FOR THE TREATMENT OF REFRACTORY SUPRA-VENTRICULAR TACHYCARDIA [J].
GALLAGHER, JJ ;
SVENSON, RH ;
KASELL, JH ;
GERMAN, LD ;
BARDY, GH ;
BROUGHTON, A ;
CRITELLI, G .
NEW ENGLAND JOURNAL OF MEDICINE, 1982, 306 (04) :194-200
[7]   Variation in left atrial transmural wall thickness at sites commonly targeted for ablation of atrial fibrillation [J].
Hall, Burr ;
Jeevanantham, Vinodh ;
Simon, Rochelle ;
Filippone, John ;
Vorobiof, Gabriel ;
Daubert, James .
JOURNAL OF INTERVENTIONAL CARDIAC ELECTROPHYSIOLOGY, 2006, 17 (02) :127-132
[8]   PROARRHYTHMIC EFFECTS OF VENTRICULAR ELECTRICAL CATHETER ABLATION IN DOGS [J].
HAUER, RNW ;
DEMEDINA, EOR ;
BORST, C .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1987, 10 (06) :1350-1356
[9]   A novel nonthermal energy source for surgical epicardial atrial ablation: Irreversible electroporation [J].
Lavee, Jacob ;
Onik, Gary ;
Mikus, Paul ;
Rubinsky, Boris .
HEART SURGERY FORUM, 2007, 10 (02) :E162-E167
[10]  
Loh P, 2019, 1 IN HUM EXP