Clinical study of the laser sheath for lead extraction: The total experience in the United States

被引:230
作者
Byrd, CL
Wilkoff, BL
Love, CJ
Sellers, TD
Reiser, C
机构
[1] Spectranetics, Colorado Springs, CO 80907 USA
[2] Broward Gen Med Ctr, Ft Lauderdale, FL USA
[3] Cleveland Clin Fdn, Cleveland, OH 44195 USA
[4] Ohio State Univ, Med Ctr, Columbus, OH 43210 USA
[5] Mem Hosp, Colorado Springs, CO USA
来源
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY | 2002年 / 25卷 / 05期
关键词
pacing lead; defibrillator lead; lead removal; ultraviolet;
D O I
10.1046/j.1460-9592.2002.t01-1-00804.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
The laser sheath uses optical fibers, delivering pulsed ultraviolet excimer laser light, to vaporize fibrotic tissue binding intravenous cardiac leads to the vein or heart wall during lead extraction from the implant vein. The total investigational experience with laser sheaths is reported. During the period from October 1995 to December 1999, 2,561 pacing and defibrillator leads were treated in 1,684 patients at 89 sites in the United States with three sizes of laser sheath. Endpoints were complete removal of the lead, partial removal (leaving the tip behind), or failure (abandoning the lead, onset of complications, change to transfemoral or transatrial approach). Minimal follow-up at 30 days was recorded. Of the leads, 90% were completely removed, 3% were partially removed, and the balance were failures. Major perioperative complications (tamponade, hemothorax, pulmonary embolism, lead migration, and death) were observed in 1.9% of patients with in hospital death in 13 (0.8%). Minor complications were seen in an additional 1.4% of patients. Multivariate analysis showed that implant duration was the only preoperative independent predictor of failure; female sex was the only multivariate predictor of complications. Success and complications were not dependent on laser sheath size. At follow-up, various extraction related complications were observed in 2% of patients. The learning curve showed a trend toward fewer complications with experience. Lead extraction with the laser sheath can be safely practiced with high success rates. Success is independent of laser sheath size. Major complications can be expected in < 2% of patients, and occur more often during an investigator's early experience.
引用
收藏
页码:804 / 808
页数:5
相关论文
共 18 条
  • [1] AGGARWAL RK, 1995, BRIT HEART J, V73, P571
  • [2] Three-year experience with a stylet for lead extraction: A multicenter study
    Alt, E
    Neuzner, J
    Binner, L
    Gohl, K
    Res, JCJ
    Knabe, UH
    Zehender, M
    Reinhardt, J
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1996, 19 (01): : 18 - 25
  • [3] Learning curve characteristics of pacing lead extraction with a laser sheath
    Bracke, FA
    Meijer, A
    Van Gelder, B
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1998, 21 (11): : 2309 - 2313
  • [4] Byrd Charles L., 1992, Cardiology Clinics, V10, P735
  • [5] BYRD CL, 1991, J THORAC CARDIOV SUR, V101, P989
  • [6] INTRAVASCULAR LEAD EXTRACTION USING LOCKING STYLETS AND SHEATHS
    BYRD, CL
    SCHWARTZ, SJ
    HEDIN, NB
    GOODE, LB
    FEARNOT, NE
    SMITH, HJ
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1990, 13 (12): : 1871 - 1875
  • [7] Initial experience with larger laser sheaths for the removal of transvenous pacemaker and implantable defibrillator leads
    Epstein, LM
    Byrd, CL
    Wilkoff, BL
    Love, CJ
    Sellers, TD
    Hayes, DL
    Reiser, C
    [J]. CIRCULATION, 1999, 100 (05) : 516 - 525
  • [8] Excimer Laser assisted extraction of permanent pacemaker and ICD leads: present experiences of a European multi-centre study
    Kennergren, C
    [J]. EUROPEAN JOURNAL OF CARDIO-THORACIC SURGERY, 1999, 15 (06) : 856 - 860
  • [9] Love CJ, 2000, PACE, V23, P544
  • [10] MEIBOM J, 1985, PACE, V8, pA54