Computerized categorization of TLD glow curve anomalies using multi-class classification support vector machines

被引:4
作者
Amit, Gal [1 ]
Datz, Hanan [1 ]
机构
[1] Soreq Nucl Res Ctr, Radiat Safety Div, Yavne, Israel
关键词
Glow curve; Machine learning; Multi-class classification; Thermoluminescent dosimetry; DECONVOLUTION;
D O I
10.1016/j.radmeas.2019.04.015
中图分类号
TL [原子能技术]; O571 [原子核物理学];
学科分类号
0827 ; 082701 ;
摘要
A novel method for automatic categorization of thermoluminescent dosimeter (TLD) glow curve (GC) anomalies is presented. This automatic categorization will improve the metrological process of dose estimation by enhancing both its repeatability and its accuracy. Moreover, it will help external dosimetry laboratories to forecast some of the malfunctions of their TLD readers. A degenerated automatic approach was previously used in order to differentiate between a regular GC and an anomalous one, without being able to distinguish between different types of anomalies. That approach is now substantially extended to implicitly enable the categorization of GCs into five different kinds of anomalies. The machine learning algorithm applied for this purpose is support vector machines (SVM). The SVM algorithm categorizes TLD GCs into either a 'good' GC or into five types of TLD GC anomalies. When applied on an uncategorized GC, SVM associates it with a classification probability for each of the six categories. Results show an accuracy rate between 87.5% and 89% for the correct categorization of GCs to either of the six classes, depending on the presence of 'spikes' class in the data.
引用
收藏
页码:1 / 6
页数:6
相关论文
共 17 条
  • [1] IMPROVEMENT OF DOSE ESTIMATION PROCESS USING ARTIFICIAL NEURAL NETWORKS
    Amit, Gal
    Datz, Hanan
    [J]. RADIATION PROTECTION DOSIMETRY, 2019, 184 (01) : 36 - 43
  • [2] Automatic detection of anomalous thermoluminescent dosimeter glow curves using machine learning
    Amit, Gal
    Datz, Hanan
    [J]. RADIATION MEASUREMENTS, 2018, 117 : 80 - 85
  • [3] [Anonymous], RAD PROTECT DOSIM
  • [4] Support vector machines for speaker and language recognition
    Campbell, WM
    Campbell, JP
    Reynolds, DA
    Singer, E
    Torres-Carrasquillo, PA
    [J]. COMPUTER SPEECH AND LANGUAGE, 2006, 20 (2-3) : 210 - 229
  • [5] Support vector machine classification and validation of cancer tissue samples using microarray expression data
    Furey, TS
    Cristianini, N
    Duffy, N
    Bednarski, DW
    Schummer, M
    Haussler, D
    [J]. BIOINFORMATICS, 2000, 16 (10) : 906 - 914
  • [6] Hierarchical classification and feature reduction for fast face detection with support vector machines
    Heisele, B
    Serre, T
    Prentice, S
    Poggio, T
    [J]. PATTERN RECOGNITION, 2003, 36 (09) : 2007 - 2017
  • [7] The thermoluminescence dose-response and other characteristics of the high-temperature TL in LiF:Mg,Ti (TLD-100)
    Horowitz, Y. S.
    Oster, L.
    Datz, H.
    [J]. RADIATION PROTECTION DOSIMETRY, 2007, 124 (02) : 191 - 205
  • [8] Highlights and pitfalls of 20 years of application of computerised glow curve analysis to thermoluminescence research and dosimetry
    Horowitz, Y. S.
    Moscovitch, M.
    [J]. RADIATION PROTECTION DOSIMETRY, 2013, 153 (01) : 1 - 22
  • [9] ANALYSIS OF THERMOLUMINESCENCE GLOW CURVES USING DERIVATIVES OF DIFFERENT ORDERS
    Karmakar, Mahua
    Bhattacharyya, S.
    Sarkar, A.
    Mazumdar, P. S.
    Singh, S. D.
    [J]. RADIATION PROTECTION DOSIMETRY, 2017, 175 (04) : 493 - 502
  • [10] An example of abnormal glow curves identification in personnel thermoluminescent dosimetry
    Piniella, VO
    Stadtmann, H
    Lankmayr, E
    [J]. JOURNAL OF BIOCHEMICAL AND BIOPHYSICAL METHODS, 2002, 53 (1-3): : 117 - 122