Experimental Demonstration of X-Ray Fluorescence CT Using a Spatially Distributed Multi-Beam X-Ray Source

被引:3
作者
Chen, Zhiqiang [1 ,2 ]
Zhang, Siyuan [1 ,2 ]
Li, Liang [1 ,2 ]
机构
[1] Tsinghua Univ, Dept Engn Phys, Beijing, Peoples R China
[2] Tsinghua Univ, Key Lab Particle & Radiat Imaging, Minist Educ, Beijing, Peoples R China
基金
国家重点研发计划;
关键词
x-ray fluorescence; computed tomography; carbon nanotubes; spatially distributed multi-beam x-ray source; x-ray spectrometer; COMPUTED-TOMOGRAPHY XFCT; NANOPARTICLE-LOADED OBJECTS; SYSTEM; MICROPROBE;
D O I
10.3389/fphy.2020.587625
中图分类号
O4 [物理学];
学科分类号
0702 ;
摘要
X-ray fluorescence computed tomography (XFCT) is a high-sensitivity imaging modality for high-atomic elements such as gadolinium (Gd) or gold (Au). In order to improve the contrast of x-ray fluorescence (XRF) signals in the raw projection data, common XFCT systems use single-pixel x-ray spectrometers to record XRF photons stimulated by the pencil-beam x-ray source and achieve line-by-line scan of the whole object by translating the x-ray source or the object. However, this kind of design results in waiting time of the translation device. One improvement would be to replace the traditional x-ray tube with a spatially distributed multi-beam x-ray source and scan the whole object by switching the exposure of each focus. In this study, we present a design of the XFCT system using a spatially distributed multi-beam source, and a XFCT imaging experiment was performed to investigate its feasibility. Each cold cathode of the x-ray source used in this study was made of carbon nanotubes, and each focus was independently controlled by the electronic control system. The incident beam was collimated by a pinhole array to produce a pencil-beam source. The object scanned by the system was a polymethyl methacrylate (PMMA) cylinder (8 cm in diameter) with Gd (20 mg/ml) and I (100 mg/ml) insertions. Results show that the distribution of iodine (I) and Gd in a PMMA phantom was successfully reconstructed, but the imaging performance was limited by the collimation and exposure mode of the current distributed x-ray source. The practicality of the current distributed x-ray source used for XFCT scan and further optimization of the proposed system are discussed according to the experimental results.
引用
收藏
页数:9
相关论文
共 31 条
[1]   Order of Magnitude Sensitivity Increase in X-ray Fluorescence Computed Tomography (XFCT) Imaging With an Optimized Spectro-Spatial Detector Configuration: Theory and Simulation [J].
Ahmad, Moiz ;
Bazalova, Magdalena ;
Xiang, Liangzhong ;
Xing, Lei .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 2014, 33 (05) :1119-1128
[2]   L-shell x-ray fluorescence computed tomography (XFCT) imaging of Cisplatin [J].
Bazalova, Magdalena ;
Ahmad, Moiz ;
Pratx, Guillem ;
Xing, Lei .
PHYSICS IN MEDICINE AND BIOLOGY, 2014, 59 (01) :219-232
[3]   Investigation of X-ray Fluorescence Computed Tomography (XFCT) and K-Edge Imaging [J].
Bazalova, Magdalena ;
Kuang, Yu ;
Pratx, Guillem ;
Xing, Lei .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 2012, 31 (08) :1620-1627
[4]   Synchrotron hard x-ray microprobe:: Fluorescence imaging of single cells [J].
Bohic, S ;
Simionovici, A ;
Snigirev, A ;
Ortega, R ;
Devès, G ;
Heymann, D ;
Schroer, CG .
APPLIED PHYSICS LETTERS, 2001, 78 (22) :3544-3546
[5]   FLUORESCENCE TOMOGRAPHY USING SYNCHROTRON RADIATION AT THE NSLS [J].
BOISSEAU, P ;
GRODZINS, L .
HYPERFINE INTERACTIONS, 1987, 33 (1-4) :283-292
[6]   A carbon nanotube field emission cathode with high current density and long-term stability [J].
Calderon-Colon, Xiomara ;
Geng, Huaizhi ;
Gao, Bo ;
An, Lei ;
Cao, Guohua ;
Zhou, Otto .
NANOTECHNOLOGY, 2009, 20 (32)
[7]   A dynamic micro-CT scanner based on a carbon nanotube field emission x-ray source [J].
Cao, G. ;
Lee, Y. Z. ;
Peng, R. ;
Liu, Z. ;
Rajaram, R. ;
Calderon-Colon, X. ;
An, L. ;
Wang, P. ;
Phan, T. ;
Sultana, S. ;
Lalush, D. S. ;
Lu, J. P. ;
Zhou, O. .
PHYSICS IN MEDICINE AND BIOLOGY, 2009, 54 (08) :2323-2340
[8]   A stationary computed tomography system with cylindrically distributed sources and detectors [J].
Chen, Yi ;
Xi, Yan ;
Zhao, Jun .
JOURNAL OF X-RAY SCIENCE AND TECHNOLOGY, 2014, 22 (06) :707-725
[9]   X-ray fluorescence computed tomography (XFCT) imaging of gold nanoparticle-loaded objects using 110 kVp x-rays [J].
Cheong, Seong-Kyun ;
Jones, Bernard L. ;
Siddiqi, Arsalan K. ;
Liu, Fang ;
Manohar, Nivedh ;
Cho, Sang Hyun .
PHYSICS IN MEDICINE AND BIOLOGY, 2010, 55 (03) :647-662
[10]   X-Ray Fluorescence Computed Tomography With Polycapillary Focusing [J].
Cong, Wenxiang ;
Xi, Yan ;
Wang, Ge .
IEEE ACCESS, 2014, 2 :1138-1142