Magnetic resonance only workflow and validation of dose calculations for radiotherapy of prostate cancer

被引:25
作者
Christiansen, Rasmus L. [1 ]
Jensen, Henrik R. [1 ]
Brink, Carsten [1 ,2 ]
机构
[1] Odense Univ Hosp, Dept Oncol, Lab Radiat Phys, Odense, Denmark
[2] Univ Southern Denmark, Inst Clin Sci, Odense, Denmark
关键词
RADIATION-THERAPY; MRI SEQUENCES; SYSTEM; CT;
D O I
10.1080/0284186X.2017.1290275
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Background: Current state of the art radiotherapy planning of prostate cancer utilises magnetic resonance (MR) for soft tissue delineation and computed tomography (CT) to provide an electron density map for dose calculation. This dual scan workflow is prone to setup and registration error. This study evaluates the feasibility of an MR-only workflow and the validity of dose calculation from an MR derived pseudo CT.Material and methods: Thirty prostate cancer patients were CT and MR scanned. Clinical treatment plans were generated on CT using a single 18MV arc volumetric modulated arc therapy (VMAT) with a prescription of 78Gy/39 fractions. Dose was recalculated on pseudo CT and assuming uniform water density. Pseudo CT and uniform density based dose calculations were compared to CT dose calculations by gamma analysis. One patient was treated with a plan based solely on MR and pseudo CT including daily image guided radiotherapy (IGRT) performed by manual match of implanted gold markers.Results: A pseudo CT was generated for 29 of the 30 patients. Median gamma pass rates for 1%/1mm passing criteria for dose calculated on pseudo CT when compared to CT were 100% for most evaluated structures. Dose calculated on uniform density also yielded high median pass rates, but with a higher occurrence of pass rates below 95%. Cases of pass rate below 95% on pseudo CT proved to originate from the presence of rectal air on CT, not represented by the pseudo CT. Treatment based on MR alone was successfully delivered to one patient, including manually performed daily IGRT.Conclusions: Median gamma pass rates were high for pseudo CT and proved superior to uniform density. Local differences in dose calculations were concluded not to have clinical relevance. Feasibility of the MR-only workflow was demonstrated through successful delivery of a treatment course planned based on MR alone.
引用
收藏
页码:787 / 791
页数:5
相关论文
共 14 条
[1]   MRI-based treatment planning for radiotherapy: Dosimetric verification for prostate IMRT [J].
Chen, LL ;
Price, RA ;
Wang, L ;
Li, JS ;
Qin, LH ;
McNeeley, S ;
Ma, CMC ;
Freedman, GM ;
Pollack, A .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2004, 60 (02) :636-647
[2]   The Value of Magnetic Resonance Imaging for Radiotherapy Planning [J].
Dirix, Piet ;
Haustermans, Karin ;
Vandecaveye, Vincent .
SEMINARS IN RADIATION ONCOLOGY, 2014, 24 (03) :151-159
[3]   Automatic Substitute Computed Tomography Generation and Contouring for Magnetic Resonance Imaging (MRI)-Alone External Beam Radiation Therapy From Standard MRI Sequences [J].
Dowling, Jason A. ;
Sun, Jidi ;
Pichler, Peter ;
Rivest-Henault, David ;
Ghose, Soumya ;
Richardson, Haylea ;
Wratten, Chris ;
Martin, Jarad ;
Arm, Jameen ;
Best, Leah ;
Chandra, Shekhar S. ;
Fripp, Jurgen ;
Menk, Frederick W. ;
Greer, Peter B. .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2015, 93 (05) :1144-1153
[4]   An Atlas-Based Electron Density Mapping Method for Magnetic Resonance Imaging (MRI)-Alone Treatment Planning and Adaptive MRI-Based Prostate Radiation Therapy [J].
Dowling, Jason A. ;
Lambert, Jonathan ;
Parker, Joel ;
Salvado, Olivier ;
Fripp, Jurgen ;
Capp, Anne ;
Wratten, Chris ;
Denham, James W. ;
Greer, Peter B. .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2012, 83 (01) :E5-E11
[5]   CT substitute derived from MRI sequences with ultrashort echo time [J].
Johansson, Adam ;
Karlsson, Mikael ;
Nyholm, Tufve .
MEDICAL PHYSICS, 2011, 38 (05) :2708-2714
[6]   New developments in Mill for target volume delineation in radiotherapy [J].
Khoo, V. S. ;
Joon, D. L. .
BRITISH JOURNAL OF RADIOLOGY, 2006, 79 :S2-S15
[7]   elastix: A Toolbox for Intensity-Based Medical Image Registration [J].
Klein, Stefan ;
Staring, Marius ;
Murphy, Keelin ;
Viergever, Max A. ;
Pluim, Josien P. W. .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 2010, 29 (01) :196-205
[8]  
Kohler M, 2015, MR ONLY SIMULATION R, P1
[9]   A dual model HU conversion from MRI intensity values within and outside of bone segment for MRI-based radiotherapy treatment planning of prostate cancer [J].
Korhonen, Juha ;
Kapanen, Mika ;
Keyrilainen, Jani ;
Seppala, Tiina ;
Tenhunen, Mikko .
MEDICAL PHYSICS, 2014, 41 (01)
[10]   The Magnetic Resonance Imaging-Linac System [J].
Lagendijk, Jan J. W. ;
Raaymakers, Bas W. ;
van Vulpen, Marco .
SEMINARS IN RADIATION ONCOLOGY, 2014, 24 (03) :207-209