Accuracy evaluation of computer-guided implant surgery associated with prototyped surgical guides

被引:29
作者
Cunha, Rodrigo Moreira [1 ]
Souza, Francisley Avila [2 ]
Hadad, Henrique [2 ]
Poli, Pier Paolo [3 ]
Maiorana, Carlo [3 ]
Carvalho, Paulo Sergio Perri [4 ]
机构
[1] Sao Leopoldo Mand Sch Dent & Res Ctr, Implant Dent Postgrad Program, Campinas, Brazil
[2] Sao Paulo State Univ UNESP, Sch Dent, Dept Surg & Integrated Clin, R Jose Bonifacio,1193 Vila Mendonca, BR-16015050 Aracatuba, SP, Brazil
[3] Univ Milan, Fdn IRCCS Ca Granda Maggiore Policlin Hosp, Implant Ctr Edentulism & Jawbone Atrophies Maxill, Milan, Italy
[4] Sao Leopoldo Mand Sch Dent & Res Ctr, Implant Dent Postgrad Program, Campinas, Brazil
关键词
D O I
10.1016/j.prosdent.2019.07.010
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Statement of problem. Correct implant placement by means of a pretreatment planning software program is still subject to deviations between the planned and achieved implant positions. Inaccuracy at this level may have drastic consequences, including neurovascular trauma. Further data are therefore needed to evaluate the accuracy of such computer-guided implant planning software programs. Purpose. The purpose of this clinical study was to evaluate the accuracy of computer-guided implant surgery associated with prototyped surgical guides. Material and methods. Cone beam computed tomography (CBCT) scans were made of the participants with a tomographic guide to merge anatomic and prosthetic data. This allowed virtual planning with a prosthetically guided approach respecting the anatomy of the participant. A prototyped surgical guide was then fabricated from the virtual plan, determining the intrasurgical position of the implants. Flapless guided implant surgery was carried out according to the manufacturer's instructions. A second CBCT was made 30 days after the surgery, to enable overlapping of the data from before and after the implant placement. The angular, coronal, central, and apical deviations of the placed implants were measured and compared with those virtually planned. The data were submitted to descriptive statistic and intraclass correlation coefficient (ICC), analysis of variance, and the Student t test (alpha=.05). Results. A total of 61 implants were analyzed. The mean angular deviation was 2.04 degrees. The mean coronal, central, and apical linear deviations were 0.68 mm, 0.72 mm, and 0.82 mm, respectively. No statistically significant difference was found between the virtual and the real position of the implants inserted. A tendency toward a greater absolute error was observed in the mandible than in the maxilla in terms of angular (P=.047), central (P=.043), and coronal error (P=.031). Conclusions. Flapless computer-guided implant surgery with virtual planning had some angular and linear deviations; nevertheless, this technique should be acceptably safe and accurate. (J Prosthet Dent 2021;125:266-72)
引用
收藏
页码:266 / 272
页数:7
相关论文
共 36 条
[1]   Implant Placement Accuracy When Using Stereolithographic Template as a Surgical Guide: Preliminary Results [J].
Al-Harbi, Saad A. ;
Sun, Albert Y. T. .
IMPLANT DENTISTRY, 2009, 18 (01) :46-51
[2]   Image distortion and spatial resolution of a commercially available cone-beam computed tomography machine [J].
Ballrick, John W. ;
Palomo, J. Martin ;
Ruch, Edward ;
Amberman, B. Douglas ;
Hans, Mark G. .
AMERICAN JOURNAL OF ORTHODONTICS AND DENTOFACIAL ORTHOPEDICS, 2008, 134 (04) :573-582
[3]  
Balshi SF, 2006, INT J PERIODONT REST, V26, P239
[4]   Computer-aided flapless implant surgery and immediate loading. A technical note [J].
Baruffaldi A. ;
Poli P.P. ;
Baruffaldi A. ;
Giberti L. ;
Pigozzo M. ;
Maiorana C. .
Oral and Maxillofacial Surgery, 2016, 20 (3) :313-319
[5]   Accuracy of computer-aided template-guided oral implant placement: a prospective clinical study [J].
Beretta, Mario ;
Poli, Pier Paolo ;
Maiorana, Carlo .
JOURNAL OF PERIODONTAL AND IMPLANT SCIENCE, 2014, 44 (04) :184-193
[6]  
Bergkvist G, 2005, INT J ORAL MAX IMPL, V20, P399
[7]   Accuracy of implant treatment planning utilizing template-guided reformatted computed tomography [J].
Besimo, CE ;
Lambrecht, JT ;
Guindy, JS .
DENTOMAXILLOFACIAL RADIOLOGY, 2000, 29 (01) :46-51
[8]   Modular Preoperative Planning Software for Computer-Aided Oral Implantology and the Application of a Novel Stereolithographic Template: A Pilot Study [J].
Chen, Xiaojun ;
Yuan, Jianbing ;
Wang, Chengtao ;
Huang, Yuanliang ;
Kang, Lu .
CLINICAL IMPLANT DENTISTRY AND RELATED RESEARCH, 2010, 12 (03) :181-193
[9]   IMPACT OF OPERATOR EXPERIENCE ON THE ACCURACY OF IMPLANT PLACEMENT WITH STEREOLITHOGRAPHIC SURGICAL TEMPLATES: AN IN VITRO STUDY [J].
Cushen, Sarra E. ;
Turkyilmaz, Ilser .
JOURNAL OF PROSTHETIC DENTISTRY, 2013, 109 (04) :248-254
[10]   A Prospective Study on the Accuracy of Mucosally Supported Stereolithographic Surgical Guides in Fully Edentulous Maxillae [J].
D'haese, Jan ;
Van De Velde, Tommie ;
Elaut, Lucien ;
De Bruyn, Hugo .
CLINICAL IMPLANT DENTISTRY AND RELATED RESEARCH, 2012, 14 (02) :293-303