DIAGNOSTIC REFERENCE LEVELS FOR PAEDIATRIC HEAD COMPUTED TOMOGRAPHY IN MOROCCO: A NATIONWIDE SURVEY

被引:15
作者
Benmessaoud, M. [1 ]
Dadouch, A. [1 ]
Talbi, M. [2 ]
Tahiri, M. [3 ]
El-ouardi, Y. [1 ]
机构
[1] Sidi Mohamed Ben Abdellah Univ, Fac Sci Dhar EL Mahraz, BP 30000, Fes, Morocco
[2] Moulay Ismail Univ, Fac Sci, BP 11201, Meknes, Morocco
[3] Hassan First Univ, Higher Inst Hlth Sci, BP 26000, Settat, Morocco
关键词
RADIATION-EXPOSURE; DOSE REDUCTION; CT; COUNTRIES; CHILDREN;
D O I
10.1093/raddos/ncaa170
中图分类号
X [环境科学、安全科学];
学科分类号
08 ; 0830 ;
摘要
The purpose of this study was to establish the diagnostic reference levels (DRLs) for paediatric head computed tomography (CT) in Morocco and to assess the effective doses received during head CT examinations. The data of 1007 patients were collected retrospectively from Moroccan university children's hospitals. The sample was classified per age group:<1, 1-5, 5-10 and 10-15 years. The proposed DRLs were defined as 75th percentile of the distributions, which were in terms of CT dose index of 26.98, 28.88, 34.00 and 38.20 mGy and dose length product of 461.64, 540.06, 627.20 and 705.98 mGy.cm, respectively. The effective doses estimated were 3.6, 2.9, 2 and 1.79 mSv, respectively. The DRLs reported in Morocco were compared with those of other countries, which were based on the same age grouping method, including Thailand, Switzerland, Japan and the international DRLs. Our initiative via the determination of the first Moroccan diagnostic reference levels for paediatric head CT must be a starting point to spread this investigation towards other examinations and imaging modalities.
引用
收藏
页码:400 / 408
页数:9
相关论文
共 36 条
[1]  
[Anonymous], ANN ICRP, DOI DOI 10.1177/0146645320936035
[2]  
[Anonymous], 2004, NRPBPE12004
[3]  
[Anonymous], 2007, RAD DOSE ADULT PEDIA
[4]   Radiation exposure from CT in early childhood: a French large-scale multicentre study [J].
Bernier, M-O ;
Rehel, J-L ;
Brisse, H. J. ;
Wu-Zhou, X. ;
Caer-Lorho, S. ;
Jacob, S. ;
Chateil, J. F. ;
Aubert, B. ;
Laurier, D. .
BRITISH JOURNAL OF RADIOLOGY, 2012, 85 (1009) :53-60
[5]   Diagnostic reference levels of paediatric computed tomography examinations performed at a dedicated Australian paediatric hospital [J].
Bibbo, Giovanni ;
Brown, Scott ;
Linke, Rebecca .
JOURNAL OF MEDICAL IMAGING AND RADIATION ONCOLOGY, 2016, 60 (04) :475-484
[6]  
Brenner David J., 2010, Reviews on Environmental Health, V25, P63
[7]   Estimated risks of radiation-induced fatal cancer from pediatric CT [J].
Brenner, DJ ;
Elliston, CD ;
Hall, EJ ;
Berdon, WE .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2001, 176 (02) :289-296
[8]   CT exposure from pediatric MDCT: results from the 2007-2008 SFIPP/ISRN survey [J].
Brisse, H. J. ;
Aubert, B. .
JOURNAL DE RADIOLOGIE, 2009, 90 (02) :207-215
[9]   Effective dose in paediatric computed tomography [J].
Chapple, CL ;
Willis, S ;
Frame, J .
PHYSICS IN MEDICINE AND BIOLOGY, 2002, 47 (01) :107-115
[10]  
European Commission, 2018, RAD PROT, V2315-2826, P33