Determination of the detective quantum efficiency of a digital X-ray detector:: Comparison of three evaluations using a common image data set

被引:51
|
作者
Neitzel, U
Günther-Kohfahl, S
Borasi, G
Samei, E
机构
[1] Philips Med Syst, D-22335 Hamburg, Germany
[2] Azienda Osped Santa Maria Nuova, Serv Fis Sanitaria, I-42100 Emilia, RE, Italy
[3] Duke Univ, Dept Radiol, Durham, NC 27710 USA
[4] Duke Univ, Dept Phys, Durham, NC 27710 USA
[5] Duke Univ, Dept Biomed Engn, Durham, NC 27710 USA
关键词
digital x-ray detector; detective quantum efficiency; modulation transfer function; noise power spectrum; standardization;
D O I
10.1118/1.1766421
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The detective quantum efficiency (DQE) of an x-ray digital imaging detector was determined independently by the three participants of this study, using the same data set consisting of edge and flat field images. The aim was to assess the possible variation in DQE originating from established, but slightly different, data processing methods used by different groups. For the case evaluated in this study differences in DQE of up to +/- 15% compared to the mean were found. The differences could be traced back mainly to differences in the modulation transfer function (MTF) and noise power spectrum (NPS) determination. Of special importance is the inclusion of a possible low-frequency drop in MTF and the proper handling of signal offsets for the determination of the NPS. When accounting for these factors the deviation between the evaluations reduced to approximately +/- 5 %. It is expected that the recently published standard on DQE determination will further reduce variations in the data evaluation and thus in the results of DQE measurements. 0 2004 American Association of Physicists in Medicine.
引用
收藏
页码:2205 / 2211
页数:7
相关论文
共 50 条
  • [21] A Novel High Quantum Efficiency MV X-Ray Detector for Image Guided Radiotherapy
    Liu, J.
    Teymurazyan, A.
    Xu, Y.
    Pang, G.
    MEDICAL PHYSICS, 2018, 45 (06) : E703 - E703
  • [22] Determination of quantum efficiency of X-ray detectors
    Zelikman M.I.
    Biomedical Engineering, 2001, 35 (4) : 180 - 185
  • [23] Detective quantum efficiency and Swank noise for spectroscopic X-ray imaging detectors
    Tanguay, Jesse
    Richtsmeier, Devon
    Dydula, Christopher
    Iniewski, Kris
    Bazalova-Carter, Magdalena
    MEDICAL PHYSICS, 2021, 48 (08) : 4668 - 4669
  • [24] An experimental framework for assessing the detective quantum efficiency of spectroscopic x-ray detectors
    Zarif Yussefian, Nikta
    Tanguay, Jesse
    MEDICAL PHYSICS, 2023, 50 (03) : 1318 - 1335
  • [25] The effects of x-ray beam hardening on detective quantum efficiency and radiation dose
    Wong, Molly Donovan
    Wu, Xizeng
    Liu, Hong
    JOURNAL OF X-RAY SCIENCE AND TECHNOLOGY, 2011, 19 (04) : 509 - 519
  • [26] MEASUREMENT OF THE DETECTIVE QUANTUM EFFICIENCY OF AN X-RAY SCREEN-FILM SYSTEM
    MARONI, C
    ZAMBONI, R
    GAMBACCINI, M
    RIMONDI, O
    OPTICS COMMUNICATIONS, 1982, 44 (02) : 79 - 84
  • [27] Detective quantum efficiency of a dual-energy photon-counting X-ray detector with charge sharing compensation
    Lee, Junho
    Yoo, Seungjun
    Oh, Seokwon
    Lim, Chang Hwy
    Park, Jong-Won
    Kim, Ho Kyung
    Journal of Instrumentation, 2025, 20 (01)
  • [28] Detective quantum efficiency of an x-ray image intensifier chain as a benchmark for amorphous silicon flat panel detectors
    Koch, A
    Macherel, JM
    Wirth, T
    de Groot, P
    Ducourant, T
    Couder, D
    Moy, JP
    Calais, E
    MEDICAL IMAGING 2001: PHYSICS OF MEDICAL IMAGING, 2001, 4320 : 115 - 120
  • [29] Comparison of experimental and theoretical efficiency of HPGe X-ray detector
    Mohanty, B. P.
    Balouria, P.
    Garg, M. L.
    Nandi, T. K.
    Mittal, V. K.
    Govil, I. M.
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2008, 584 (01): : 186 - 190
  • [30] Detective quantum efficiency of direct, flat panel x-ray imaging detectors for fluoroscopy
    Hunt, DC
    Zhao, W
    Rowlands, JA
    PHYSICS OF MEDICAL IMAGING, 1998, 3336 : 408 - 417