Determination of the detective quantum efficiency of a digital X-ray detector:: Comparison of three evaluations using a common image data set

被引:51
|
作者
Neitzel, U
Günther-Kohfahl, S
Borasi, G
Samei, E
机构
[1] Philips Med Syst, D-22335 Hamburg, Germany
[2] Azienda Osped Santa Maria Nuova, Serv Fis Sanitaria, I-42100 Emilia, RE, Italy
[3] Duke Univ, Dept Radiol, Durham, NC 27710 USA
[4] Duke Univ, Dept Phys, Durham, NC 27710 USA
[5] Duke Univ, Dept Biomed Engn, Durham, NC 27710 USA
关键词
digital x-ray detector; detective quantum efficiency; modulation transfer function; noise power spectrum; standardization;
D O I
10.1118/1.1766421
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The detective quantum efficiency (DQE) of an x-ray digital imaging detector was determined independently by the three participants of this study, using the same data set consisting of edge and flat field images. The aim was to assess the possible variation in DQE originating from established, but slightly different, data processing methods used by different groups. For the case evaluated in this study differences in DQE of up to +/- 15% compared to the mean were found. The differences could be traced back mainly to differences in the modulation transfer function (MTF) and noise power spectrum (NPS) determination. Of special importance is the inclusion of a possible low-frequency drop in MTF and the proper handling of signal offsets for the determination of the NPS. When accounting for these factors the deviation between the evaluations reduced to approximately +/- 5 %. It is expected that the recently published standard on DQE determination will further reduce variations in the data evaluation and thus in the results of DQE measurements. 0 2004 American Association of Physicists in Medicine.
引用
收藏
页码:2205 / 2211
页数:7
相关论文
共 50 条