Free software for performing physical analysis of systems for digital radiography and mammography

被引:44
作者
Donini, Bruno [1 ]
Rivetti, Stefano [2 ]
Lanconelli, Nico [1 ]
Bertolini, Marco [3 ]
机构
[1] Univ Bologna, Dept Phys & Astron, I-40127 Bologna, Italy
[2] Osped Sassuolo SpA, I-41049 Sassuolo, Italy
[3] Azienda Osped ASMN, Ist Ricovero & Cura Carattere Sci, Med Phys Unit, I-42123 Reggio Emilia, Italy
关键词
modulation transfer function; detective quantum efficiency; digital radiology; digital mammography; PSYCHOPHYSICAL CHARACTERIZATION;
D O I
10.1118/1.4870955
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: In this paper, the authors present a free software for assisting users in achieving the physical characterization of x-ray digital systems and image quality checks. Methods: The program was developed as a plugin of a well-known public-domain suite ImageJ. The software can assist users in calculating various physical parameters such as the response curve (also termed signal transfer property), modulation transfer function (MTF), noise power spectra (NPS), and detective quantum efficiency (DQE). It also includes the computation of some image quality checks: defective pixel analysis, uniformity, dark analysis, and lag. Results: The software was made available in 2009 and has been used during the last couple of years by many users who gave us valuable feedback for improving its usability. It was tested for achieving the physical characterization of several clinical systems for digital radiography and mammography. Various published papers made use of the outcomes of the plugin. Conclusions: This software is potentially beneficial to a variety of users: physicists working in hospitals, staff working in radiological departments, such as medical physicists, physicians, engineers. The plugin, together with a brief user manual, are freely available and can be found online (www.medphys.it/downloads.htm). With our plugin users can estimate all three most important parameters used for physical characterization (MTF, NPS, and also DQE). The plugin can run on any operating system equipped with ImageJ suite. The authors validated the software by comparing MTF and NPS curves on a common set of images with those obtained with other dedicated programs, achieving a very good agreement. (C) 2014 American Association of Physicists in Medicine.
引用
收藏
页数:10
相关论文
共 17 条
[1]  
American Association of Physicists in Medicine, 2006, 93 AAPM
[2]   A comparison of digital radiography systems in terms of effective detective quantum efficiency [J].
Bertolini, Marco ;
Nitrosi, Andrea ;
Rivetti, Stefano ;
Lanconelli, Nico ;
Pattacini, Pierpaolo ;
Ginocchi, Vladimiro ;
Iori, Mauro .
MEDICAL PHYSICS, 2012, 39 (05) :2617-2627
[3]   DQE(F) OF 4 GENERATIONS OF COMPUTED RADIOGRAPHY ACQUISITION DEVICES [J].
DOBBINS, JT ;
ERGUN, DL ;
RUTZ, L ;
HINSHAW, DA ;
BLUME, H ;
CLARK, DC .
MEDICAL PHYSICS, 1995, 22 (10) :1581-1593
[4]  
European Commission, 2012, 162 EUR COMM
[5]   A SIMPLE METHOD FOR DETERMINING THE MODULATION TRANSFER-FUNCTION IN DIGITAL RADIOGRAPHY [J].
FUJITA, H ;
TSAI, DY ;
ITOH, T ;
DOI, K ;
MORISHITA, J ;
UEDA, K ;
OHTSUKA, A .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 1992, 11 (01) :34-39
[6]  
*INT EL COMM, 2003, IEC61267
[7]  
International Electrotechnical Commission, 2007, International Standard IEC 62 220-1-2
[8]  
International Electrotechnical Commission, 2003, IEC622201
[9]   Application of QC_DR Software for Acceptance Testing and Routine Quality Control of Direct Digital Radiography Systems: Initial Experiences using the Italian Association of Physicist in Medicine Quality Control Protocol [J].
Nitrosi, Andrea ;
Bertolini, Marco ;
Borasi, Giovanni ;
Botti, Andrea ;
Barani, Adriana ;
Rivetti, Stefano ;
Pierotti, Luisa .
JOURNAL OF DIGITAL IMAGING, 2009, 22 (06) :656-666
[10]  
Perry N., 2006, EUROPEAN GUIDELINES