Continuous glucose monitoring by means of fiber-based, mid-infrared laser spectroscopy

被引:33
作者
Lambrecht, A.
Beyer, T.
Hebestreit, K.
Mischler, R.
Petrich, W.
机构
[1] Fraunhofer Inst Phys Measurement Tech IPM, D-79110 Freiburg, Germany
[2] Roche Diagnost GmbH, D-68305 Mannheim, Germany
[3] Heidelberg Univ, Fac Phys & Astron, D-69122 Heidelberg, Germany
关键词
mid-infrared; glucose; spectroscopy; quantum cascade lasers;
D O I
10.1366/000370206777887071
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
The application of mid-infrared laser spectroscopy to the reagent-free quantification of the concentration of glucose was investigated using cryogenically cooled lead salt lasers or, alternatively, quantum cascade lasers operating at room temperature. The concentration of glucose in aqueous solutions was quantified by means of fiber-based attenuated total reflection (ATR) spectroscopy (fiber-optical evanescent field analysis FEFA) as well as fiber-based transmission spectroscopy. Both methods have the potential to be utilized by small fiber sensors, which can be inserted into the subcutaneous tissue in order to continuously measure the local concentration of glucose. In our in vitro experiments. noise-equivalent concentrations as low as 10 mg/dL were achieved. The midterm stability of the measurement schemes was investigated by means of Allan variance analysis. Based on the research presented in this manuscript, an all-room-temperature measurement scheme using quantum cascade lasers, miniaturized fiber-optic sensors, and pyroelectric detectors appears well suited for the continuous monitoring of glucose concentrations at physiological levels.
引用
收藏
页码:729 / 736
页数:8
相关论文
共 35 条
[11]   QUANTUM CASCADE LASER [J].
FAIST, J ;
CAPASSO, F ;
SIVCO, DL ;
SIRTORI, C ;
HUTCHINSON, AL ;
CHO, AY .
SCIENCE, 1994, 264 (5158) :553-556
[12]   High power mid-infrared (lambda greater than or similar to-5 mu m) quantum cascade lasers operating above room temperature [J].
Faist, J ;
Capasso, F ;
Sirtori, C ;
Sivco, DL ;
Baillargeon, JN ;
Hutchinson, AL ;
Chu, SNG ;
Cho, AY .
APPLIED PHYSICS LETTERS, 1996, 68 (26) :3680-3682
[13]   Blood diagnostics using fiberoptic evanescent wave spectroscopy and neural networks analysis [J].
Gotshal, Y ;
Simhi, R ;
Sela, BA ;
Katzir, A .
SENSORS AND ACTUATORS B-CHEMICAL, 1997, 42 (03) :157-161
[14]  
Grisar R, 1998, TECH MESS, V65, P412
[15]   Bio-analytical applications of mid-infrared spectroscopy using silver halide fiber-optic probes [J].
Heise, HM ;
Küpper, L ;
Butvina, LN .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 2002, 57 (10) :1649-1663
[16]   Attenuated total reflection mid-infrared spectroscopy for clinical chemistry applications using silver halide fibers [J].
Heise, HM ;
Küpper, L ;
Butvina, LN .
SENSORS AND ACTUATORS B-CHEMICAL, 1998, 51 (1-3) :84-91
[17]   Assessment of quantum cascade lasers as mid infrared light sources for measurement of aqueous samples [J].
Kölhed, M ;
Haberkorn, M ;
Pustogov, V ;
Mizaikoff, B ;
Frank, J ;
Karlberg, B ;
Lendl, B .
VIBRATIONAL SPECTROSCOPY, 2002, 29 (1-2) :283-289
[18]   MOLECULAR-BEAM EPITAXY OF LATERALLY STRUCTURED LEAD CHALCOGENIDES FOR THE FABRICATION OF BURIED HETEROSTRUCTURE LASERS [J].
LAMBRECHT, A ;
BOTTNER, H ;
AGNE, M ;
KURBEL, R ;
FACH, A ;
HALFORD, B ;
SCHIESSL, U ;
TACKE, M .
SEMICONDUCTOR SCIENCE AND TECHNOLOGY, 1993, 8 (01) :S334-S336
[19]   Gas monitoring in the process industry using diode laser spectroscopy [J].
Linnerud, I ;
Kaspersen, P ;
Jaeger, T .
APPLIED PHYSICS B-LASERS AND OPTICS, 1998, 67 (03) :297-305
[20]   Quantitation of glucose and urea in whole blood by mid-infrared spectroscopy of dry films [J].
Low-Ying, S ;
Shaw, RA ;
Leroux, M ;
Mantsch, HH .
VIBRATIONAL SPECTROSCOPY, 2002, 28 (01) :111-116