Stress shielding after total knee replacement may cause bone resorption in the distal femur

被引:105
作者
vanLenthe, GH
Malefijt, MCD
Huiskes, R
机构
[1] UNIV NIJMEGEN, BIOMECH SECT, NL-6500 HB NIJMEGEN, NETHERLANDS
[2] UNIV NIJMEGEN, INST ORTHOPAED, NL-6500 HB NIJMEGEN, NETHERLANDS
来源
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME | 1997年 / 79B卷 / 01期
关键词
D O I
10.1302/0301-620X.79B1.6808
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Inadequate bone stock is often found in revision surgery of femoral components of total knee replacements, Our aim was to test the hypothesis that these remodelling patterns can be explained by stress shielding, and that prosthetic bonding characteristics affect maintenance of bone mass. We made a three-dimensional finite-element model of an average male femur,with a cemented femoral knee component. This model was integrated with iterative remodelling procedures. Two extreme prosthetic bonding conditions were analysed and gradual changes in bone density were calculated. The long-term bone loss under the femoral knee component resembled clinical findings which confirms the hypothesis that stress shielding can cause distal femoral bone loss. Our study predicts, contrary to clinical findings, that an equilibrium situation is not reached after two years, but that bone resorption may continue, This hidden bone loss may be so drastic that large reconstructions are needed at the time of a revision.
引用
收藏
页码:117 / 122
页数:6
相关论文
共 36 条
[1]  
ANGELIDES M, 1988, T ORTHOP RES SOC, V13, P475
[2]  
BOBYN JD, 1982, CLIN ORTHOP RELAT R, V166, P301
[3]  
Cadambi A, 1994, J Arthroplasty, V9, P579, DOI 10.1016/0883-5403(94)90111-2
[4]  
Cameron H U, 1987, Orthop Rev, V16, P449
[5]   COMPRESSIVE BEHAVIOR OF BONE AS A 2-PHASE POROUS STRUCTURE [J].
CARTER, DR ;
HAYES, WC .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1977, 59 (07) :954-962
[6]  
ERKMAN MJ, 1974, BIOMED ENG, V9, P14
[7]   ADAPTIVE BONE-REMODELING THEORY APPLIED TO PROSTHETIC-DESIGN ANALYSIS [J].
HUISKES, R ;
WEINANS, H ;
GROOTENBOER, HJ ;
DALSTRA, M ;
FUDALA, B ;
SLOOFF, TJ .
JOURNAL OF BIOMECHANICS, 1987, 20 (11-12) :1135-1150
[8]  
HUISKES R, 1995, BRIS MYER Z, P159
[9]  
HUISKES R, 1992, CLIN ORTHOP RELAT R, P124
[10]  
HUISKES R, 1995, CLIN ORTHOP RELAT R, V319, P64