Sterilization of heat-resistant instruments with infrared radiation

被引:9
作者
Mata-Portuguez, VH
Pérez, LS
Acosta-Gío, E
机构
[1] Univ Nacl Autonoma Mexico, Sch Dent, Postgrad Div, Microbiol & Infect Control Lab, Mexico City 04510, DF, Mexico
[2] Metropolitan Autonomous Univ Xochimilco, Hlth Attent Dept, Mexico City, DF, Mexico
关键词
D O I
10.1086/502072
中图分类号
R1 [预防医学、卫生学];
学科分类号
1004 ; 120402 ;
摘要
OBJECTIVE: To evaluate the lethality profile of an infrared radiation (IR) prototype sterilizer. METHOD: Simulated use and D value tests were conducted with Bacillus subtilis spores American Type Culture Collection-9372. A spore suspension (1.06 +/- 0.03 x 10(6)) in 5% bovine serum albumin was air dried on stainless steel instruments. IR cycles were completed and the instruments were immersed in tryptic soy broth for 120 hours at 37degreesC. Direct enumeration of survivors was performed to evaluate IR death. Instrument loads contained carriers laden with spores (1.06 +/- 0.3 x 10(6)). The spores were seeded on tryptic soy agar and survivors were counted after 120 hours of incubation at 37degreesC. RESULTS: All instruments exposed to IR (n = 50) were culture negative. In contrast, all unprocessed instruments (n = 30) showed B. subtilis growth. The prototype's D value was 0.56, and the death rate's slope was -1.76 (r = -0.99741; P < .0001). The 10(-6) sterility assurance level was reached after 8 minutes and 40 seconds of exposure, from cold start. CONCLUSIONS: IR destroys B. subtilis spores. Instrument sterilization with IR may provide another technology for infection control.
引用
收藏
页码:393 / 396
页数:4
相关论文
共 3 条
[1]   Diffuse lamellar keratitis related to endotoxins released from sterilizer reservoir biofilms [J].
Holland, SP ;
Mathias, RG ;
Morck, DW ;
Chiu, J ;
Slade, SG .
OPHTHALMOLOGY, 2000, 107 (07) :1227-1233
[2]   Infection control: the role of disinfection and sterilization [J].
Rutala, WA ;
Weber, DJ .
JOURNAL OF HOSPITAL INFECTION, 1999, 43 :S43-S55
[3]  
*US FOOD DRUG ADM, 1993, GUID PREM NOT STER I