A reliable in vitro approach to assess the stability of acetabular implants using digital image correlation

被引:6
作者
Morosato, F. [1 ]
Traina, F. [2 ]
Cristofolini, L. [1 ]
机构
[1] Alma Mater Studiorum Univ Bologna, Sch Engn & Architecture, Dept Ind Engn, Viale Risorgimento 2, I-40136 Bologna, Italy
[2] Rizzoli Orthopaed Inst, Clin Orthopaed & Traumatol 2, Bologna, Italy
关键词
acetabular loosening; biomechanical testing; bone strain; digital image correlation (DIC); hip biomechanics; implant stability; permanent migrations and inducible micromotions; total hip replacement; HIP-ARTHROPLASTY; DIAGNOSTIC PERFORMANCE; INITIAL STABILITY; EARLY MIGRATION; REVISION; COMPONENTS; CUPS; MICROMOTION; PREDICTOR; CEMENT;
D O I
10.1111/str.12318
中图分类号
TB3 [工程材料学];
学科分类号
0805 ; 080502 ;
摘要
The main cause of failure of the hip acetabular component is aseptic loosening. Preclinical test methods currently used to assess the stability of hip acetabular implants rely on crude simplifications. Normally, either one component of motion or bone strains are measured. We developed a test method to measure implant 3D translations and rotations and bone strains using digital image correlation. Hemipelvises were aligned and potted to allow consistent testing. A force was applied in the direction of the load peak during level walking. The force was applied in 100-cycle packages, each load package being 20% larger than the previous one. A digital image correlation system allowed measuring the cup-bone relative 3D displacements (permanent migrations and inducible micromotions) and the strain distribution in the periacetabular bone. To assess the test repeatability, the protocol was applied to six composite hemipelvises implanted with very stable cups. To assess the suitability of the method to detect mobilisation, six loose implants were tested. The method was repeatable: the interspecimen variability was 16 mu m for the bone/cup relative translations, 0.04 degrees for the rotations. The method was capable of tracking extremely loose implants (translations up to 4.5 mm; rotations up to 30 degrees). The strain distribution in the bone was measured, showing the areas of highest strain. We have shown that it is possible to measure the 3D relative translations and rotations of an acetabular cup inside the pelvis and simultaneously to measure the full-field strain distribution in the bone surface. This will allow better preclinical testing of the stability of acetabular implants.
引用
收藏
页数:12
相关论文
共 43 条
[1]   The diagnostic performance of radiographic criteria to detect aseptic acetabular component loosening after revision total hip arthroplasty [J].
Abrahams, J. M. ;
Kim, Y. S. ;
Callary, S. A. ;
De Ieso, C. ;
Costi, K. ;
Howie, D. W. ;
Solomon, L. B. .
BONE & JOINT JOURNAL, 2017, 99B (04) :458-464
[2]  
Adler E, 1992, J Arthroplasty, V7, P295, DOI 10.1016/0883-5403(92)90052-R
[3]  
[Anonymous], 2009, IMAGE CORRELATION SH, DOI DOI 10.1007/978-0-387-78747-3
[4]   Primary stability of two uncemented acetabular components of different geometry: hemispherical or peripherally enhanced? [J].
Antoniades, G. ;
Smith, E. J. ;
Deakin, A. H. ;
Wearing, S. C. ;
Sarungi, M. .
BONE & JOINT RESEARCH, 2013, 2 (12) :264-269
[5]   Comparison of the Primary Stability of a Porous Coated Acetabular Revision Cup With a Standard Cup [J].
Beckmann, Nicholas A. ;
Jaeger, Sebastian ;
Janoszka, Maciej B. ;
Klotz, Matthias C. ;
Bruckner, Thomas ;
Bitsch, Rudi G. .
JOURNAL OF ARTHROPLASTY, 2018, 33 (02) :580-585
[6]  
Birolini A, 2015, SPR SER SOLID STRUCT, V4, P187, DOI 10.1007/978-3-319-06086-6_5
[7]   Characteristics of bone ingrowth and interface mechanics of a new porous tantalum biomaterial [J].
Bobyn, JD ;
Stackpool, GJ ;
Hacking, SA ;
Tanzer, M ;
Krygier, JJ .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1999, 81B (05) :907-914
[8]  
Bordini B., 2017, REGIONAL REGISTER OR, P1
[9]   Primary stability of cementless threaded acetabular cups at first implantation and in the case of revision regarding micromotions as indicators [J].
Buerkner, Andreas ;
Fottner, Andreas ;
Lichtinger, Thomas ;
Teske, Wolfram ;
Vogel, Tobias ;
Jansson, Volkmar ;
Pellengahr, Christoph von Schulze .
BIOMEDICAL ENGINEERING-BIOMEDIZINISCHE TECHNIK, 2012, 57 (03) :169-174
[10]  
Craig J.J., 2005, INTRO ROBOTICS MECH