A Monte Carlo based lookup table for spectrum analysis of turbid media in the reflectance probe regime

被引:4
作者
Wen, Xiang [1 ,2 ]
Zhong, Xiewei [1 ,2 ]
Yu, Tingting [1 ,2 ]
Zhu, Dan [1 ,2 ]
机构
[1] Huazhong Univ Sci & Technol, Britton Chance Ctr Biomed Photon, Wuhan Natl Lab Optoelect, Wuhan 430074, Peoples R China
[2] Huazhong Univ Sci & Technol, Dept Biomed Engn, Minist Educ, Key Lab Biomed Photon, Wuhan 430074, Peoples R China
关键词
optical fibres; diffuse reflectance; optical spectra; turbid media; Monte Carlo simulation; MODELING DIFFUSE-REFLECTANCE; NEAR-INFRARED SPECTROSCOPY; OPTICAL-PROPERTIES; INVERSE MODEL; SCATTERING; ABSORPTION;
D O I
10.1070/QE2014v044n07ABEH015481
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
Fibre-optic diffuse reflectance spectroscopy offers a method for characterising phantoms of biotissue with specified optical properties. For a commercial reflectance probe (six source fibres surrounding a central collection fibre with an inter-fibre spacing of 480 mm; R400-7, Ocean Optics, USA) we have constructed a Monte Carlo based lookup table to create a function called getR(mu(a), mu(s)'), where mu(a) is the absorption coefficient and mu(s)' is the reduced scattering coefficient. Experimental measurements of reflectance from homogeneous calibrated phantoms with given optical properties are compared with the predicted reflectance from the lookup table. The deviation between experiment and prediction is on average 12.1 %.
引用
收藏
页码:641 / 645
页数:5
相关论文
共 21 条
  • [1] Least-squares support vector machines and near infrared spectroscopy for quantification of common adulterants in powdered milk
    Borin, Alessandra
    Ferrao, Marco Flores
    Mello, Cesar
    Maretto, Danilo Althmann
    Poppi, Ronei Jesus
    [J]. ANALYTICA CHIMICA ACTA, 2006, 579 (01) : 25 - 32
  • [2] Cao J., 2013, ISRN TEXTILES, V2013
  • [3] Hemoglobin oxygen saturations in phantoms and in vivo from measurements of steady-state diffuse reflectance at a single, short source-detector separation
    Finlay, JC
    Foster, TH
    [J]. MEDICAL PHYSICS, 2004, 31 (07) : 1949 - 1959
  • [4] OPTICAL-CONSTANTS OF WATER IN 200-NM TO 200-MUM WAVELENGTH REGION
    HALE, GM
    QUERRY, MR
    [J]. APPLIED OPTICS, 1973, 12 (03) : 555 - 563
  • [5] GRAPHICS PROCESSING UNIT CLUSTER ACCELERATED MONTE CARLO SIMULATION OF PHOTON TRANSPORT IN MULTI-LAYERED TISSUES
    Jiang, Chao
    He, Heng
    Li, Pengcheng
    Luo, Qingming
    [J]. JOURNAL OF INNOVATIVE OPTICAL HEALTH SCIENCES, 2012, 5 (02)
  • [6] Inverse Monte Carlo for estimation of scattering and absorption in liquid optical phantoms
    Karlsson, Hanna
    Fredriksson, Ingemar
    Larsson, Marcus
    Stromberg, Tomas
    [J]. OPTICS EXPRESS, 2012, 20 (11): : 12233 - 12246
  • [7] Calibration of scattering and absorption properties of a liquid diffusive medium at NIR wavelengths. CW method
    Martelli, Fabrizio
    Zaccanti, Giovanni
    [J]. OPTICS EXPRESS, 2007, 15 (02): : 486 - 500
  • [8] 204 years of near infrared technology: 1800-2003
    McClure, WF
    [J]. JOURNAL OF NEAR INFRARED SPECTROSCOPY, 2003, 11 (06) : 487 - 518
  • [9] Optical properties of fat emulsions
    Michels, Rene
    Foschum, Florian
    Kienle, Alwin
    [J]. OPTICS EXPRESS, 2008, 16 (08) : 5907 - 5925
  • [10] Performance of a lookup table-based approach for measuring tissue optical properties with diffuse optical spectroscopy
    Nichols, Brandon S.
    Rajaram, Narasimhan
    Tunnell, James W.
    [J]. JOURNAL OF BIOMEDICAL OPTICS, 2012, 17 (05)