Energy response of AlGaAs soft X-ray photon counting detectors

被引:1
作者
Lees, J. E. [1 ]
Bugby, S. L. [1 ]
Boie, L. [2 ]
Scoular, C. [1 ]
机构
[1] Univ Leicester, Dept Phys & Astron, Space Res Ctr, Leicester LE1 7RH, Leics, England
[2] Free Univ Berlin, Dept Phys, D-14195 Berlin, Germany
来源
JOURNAL OF INSTRUMENTATION | 2014年 / 9卷
关键词
Solid state detectors; X-ray detectors; Radiation-hard detectors; PHOTODIODE; NOISE;
D O I
10.1088/1748-0221/9/05/P05003
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
A variety of new types of detectors based on wide band gap materials have been developed for soft X-ray spectroscopy applications (e.g. GaAs, SiC, diamond). In this report we describe the spectroscopic performance of a simple p-i-n diode fabricated on AlGaAs. The energy response of the diode, operating in photon counting mode, at room temperature has been investigated using fluorescence from a number of high purity materials. X-ray spectra over the energy range 5 keV-25 keV show this type of diode can be used for spectroscopy with promising energy resolution (similar to 1.3 keV) at 30 degrees C and excellent linearity.
引用
收藏
页数:13
相关论文
共 25 条
  • [1] [Anonymous], INSTR SHEET 554 800, V1
  • [2] Modelling results of avalanche multiplication in AlGaAs soft X-ray APDs
    Barnett, A. M.
    Lees, J. E.
    Bassford, D. J.
    Ng, J. S.
    Tan, C. H.
    Gomez, R. B.
    [J]. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2011, 626 : 25 - 30
  • [3] Temperature dependence of AlGaAs soft X-ray detectors
    Barnett, A. M.
    Bassford, D. J.
    Lees, J. E.
    Ng, J. S.
    Tan, C. H.
    David, J. P. R.
    [J]. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2010, 621 (1-3) : 453 - 455
  • [4] Barnett A.M., 2012, DETERMINATION ELECT
  • [5] Criteria of choice of the front-end transistor for low-noise preamplification of detector signals at sub-microsecond shaping times for X- and gamma-ray spectroscopy
    Bertuccio, G
    Pullia, A
    DeGeronimo, G
    [J]. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1996, 380 (1-2) : 301 - 307
  • [6] A NOVEL CHARGE SENSITIVE PREAMPLIFIER WITHOUT THE FEEDBACK RESISTOR
    BERTUCCIO, G
    REHAK, P
    XI, DM
    [J]. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1993, 326 (1-2) : 71 - 76
  • [7] Possibility of subelectron noise with room-temperature silicon carbide pixel detectors
    Bertuccio, Giuseppe
    Caccia, Stefano
    Casiraghi, Roberto
    Lanzieri, Claudio
    [J]. IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2006, 53 (04) : 2421 - 2427
  • [8] A graded-gap X-ray detector with charge avalanche multiplication
    Dapkus, L.
    Jasutis, V.
    Juciene, V.
    Pozela, J.
    Pozela, K.
    Silenas, A.
    [J]. LITHUANIAN JOURNAL OF PHYSICS, 2007, 47 (01): : 37 - 40
  • [9] del Rio M.S., 2003, XRAYLIB tables (X-ray fluorescence cross-section)
  • [10] Fraser GW., 1989, XRAY DETECTORS ASTRO