Remote and wearable ECG devices with diagnostic abilities in adults: A state-of-the-science scoping review

被引:32
作者
Bouzid, Zeineb [1 ]
Al-Zaiti, Salah S. [2 ,3 ,4 ]
Bond, Raymond [5 ]
Sejdic, Ervin [6 ,7 ]
机构
[1] Univ Pittsburgh, Swanson Sch Engn, Dept Elect & Comp Engn, 4420 Bayard St, Pittsburgh, PA 15213 USA
[2] Univ Pittsburgh, Dept Acute & Tertiary Care Nursing, Pittsburgh, PA 15213 USA
[3] Univ Pittsburgh, Dept Emergency Med, Pittsburgh, PA 15213 USA
[4] Univ Pittsburgh, Div Cardiol, Pittsburgh, PA 15213 USA
[5] Ulster Univ, Sch Comp, Belfast, Antrim, North Ireland
[6] Univ Toronto, Fac Appl Sci & Engn, Edward S Rogers Dept Elect & Comp Engn, Toronto, ON, Canada
[7] North York Gen Hosp, Toronto, ON, Canada
基金
美国国家卫生研究院;
关键词
Atrial fibrillation; Device; Diagnosis; Electrocardiogram; Remote; Wearable; HEART-ASSOCIATION ELECTROCARDIOGRAPHY; ATRIAL-FIBRILLATION; CARDIOLOGY-FOUNDATION; ARRHYTHMIAS COMMITTEE; SCIENTIFIC STATEMENT; CLINICAL CARDIOLOGY; APPLE WATCH; RHYTHM; TECHNOLOGY; ACCURACY;
D O I
10.1016/j.hrthm.2022.02.030
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
The electrocardiogram (ECG) records the electrical activity in the heart in real time, providing an important opportunity to detecting various cardiac pathologies. The 12-lead ECG currently serves as the "standard" ECG acquisition technique for diagnostic purposes for many cardiac pathologies other than arrhythmias. However, the technical aspects of acquiring a 12-lead ECG are not easy. and its usage currently is restricted to trained medical personnel, which limits the scope of its usefulness. Remote and wearable ECG devices have attempted to bridge this gap by enabling patients to take their own ECG using a simplified method at the expense of a reduced number of leads, usually a single-lead ECG. In this review, we summarize the studies that investigated the use of remote ECG devices and their clinical utility in diagnosing cardiac pathologies. Eligible studies discussed Food and Drug Administration-cleared, commercially available devices that were validated in an adult population. We summarize technical logistics of signal quality and device reliability, dimensional and functional features, and diagnostic value. Our synthesis shows that reduced-set ECG wearables have huge potential for long-term monitoring, particularly if paired with realtime notification techniques. Such capabilities make them primarily useful for abnormal rhythm detection, and there is sufficient evidence that a remote ECG device can be superior to the traditional 12-lead ECG in diagnosing specific arrhythmias such as atrial fibrillation. However, this review identifies important challenges faced by this technology and highlights the limited availability of clinical research examining their usefulness.
引用
收藏
页码:1192 / 1201
页数:10
相关论文
共 59 条
[21]  
HERMAN MV, 1991, CLIN CARDIOL, V14, P469
[22]   Diagnostic Accuracy of a Smartphone-Operated, Single-Lead Electrocardiography Device for Detection of Rhythm and Conduction Abnormalities in Primary Care [J].
Himmelreich, Jelle C. L. ;
Karregat, Evert P. M. ;
Lucassen, Wirn A. M. ;
van Weert, Henk C. P. M. ;
de Groot, Joris R. ;
Handoko, M. Louis ;
Nijveldt, Robin ;
Harskamp, Ralf E. .
ANNALS OF FAMILY MEDICINE, 2019, 17 (05) :403-411
[23]   Wearable Devices in Clinical Trials: Hype and Hypothesis [J].
Izmailova, Elena S. ;
Wagner, John A. ;
Perakslis, Eric D. .
CLINICAL PHARMACOLOGY & THERAPEUTICS, 2018, 104 (01) :42-52
[24]  
Jaffe AS, 2013, CLIN BIOCHEM, V46, P1, DOI [10.1093/eurheartj/ehs184, 10.1016/j.clinbiochem.2012.10.036, 10.1161/CIR.0b013e31826e1058, 10.1016/j.jacc.2012.08.001, 10.1016/j.gheart.2018.08.004, 10.1016/j.gheart.2012.08.001, 10.5603/KP.2018.0203]
[25]   PREDICTING CARDIOTHORACIC VOLTAGES DURING HIGH-ENERGY SHOCKS - METHODOLOGY AND COMPARISON OF EXPERIMENTAL TO FINITE-ELEMENT MODEL DATA [J].
JORGENSON, DB ;
SCHIMPF, PH ;
SHEN, I ;
JOHNSON, G ;
BARDY, GH ;
HAYNOR, DR ;
KIM, YM .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1995, 42 (06) :559-571
[26]   COMPUTATIONAL STUDIES OF TRANSTHORACIC AND TRANSVENOUS DEFIBRILLATION IN A DETAILED 3-D HUMAN THORAX MODEL [J].
JORGENSON, DB ;
HAYNOR, DR ;
BARDY, GH ;
KIM, YM .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1995, 42 (02) :172-184
[27]   Evaluation of general practitioners' single-lead electrocardiogram interpretation skills: a case-vignette study [J].
Karregat, Evert P. M. ;
Himmelreich, Jelle C. L. ;
Lucassen, Wim A. M. ;
Busschers, Wim B. ;
van Weert, Henk C. P. M. ;
Harskamp, Ralf E. .
FAMILY PRACTICE, 2021, 38 (02) :70-75
[28]   Accurate interpretation of the 12-lead ECG electrode placement: A systematic review [J].
Khunti, Kirti .
HEALTH EDUCATION JOURNAL, 2014, 73 (05) :610-623
[29]   Comparison of electrocardiograms (ECG) waveforms and centralized ECG measurements between a simple 6-lead mobile ECG device and a standard 12-lead ECG [J].
Kleiman, Robert ;
Darpo, Borje ;
Brown, Randy ;
Rudo, Todd ;
Chamoun, Svetlana ;
Albert, David E. ;
Bos, Johan Martijn ;
Ackerman, Michael J. .
ANNALS OF NONINVASIVE ELECTROCARDIOLOGY, 2021, 26 (06)
[30]   Recommendations for the standardization and interpretation of the electrocardiogram - Part I: The electrocardiogram and its technology - A scientific statement from the American Heart Association Electrocardiography and Arrhythmias Committee, Council on Clinical Cardiology; the American College of Cardiology Foundation; and the Heart Rhythm Society [J].
Kligfield, Paul ;
Gettes, Leonard S. ;
Bailey, James J. ;
Childers, Rory ;
Deal, Barbara J. ;
Hancock, E. William ;
van Herpen, Gerard ;
Kors, Jan A. ;
Macfarlane, Peter ;
Mirvis, David M. ;
Pahlm, Olle ;
Rautaharju, Pentti ;
Wagner, Galen S. .