Estimation of effective x-ray tissue attenuation differences for volumetric breast density measurement

被引:0
|
作者
Chen, Biao [1 ]
Ruth, Chris [2 ]
Jing, Zhenxue [1 ]
Ren, Baorui [2 ]
Smith, Andrew [2 ]
Kshirsagar, Ashwini [3 ]
机构
[1] Hologic Inc, 600 Technol Dr, Newark, DE 19702 USA
[2] Hologic Inc, Bedford, MA 01730 USA
[3] Hologic Inc, Santa Clara, CA 95054 USA
来源
MEDICAL IMAGING 2014: PHYSICS OF MEDICAL IMAGING | 2014年 / 9033卷
关键词
Digital Mammography; Breast Density Measurement; Breast Imaging; Cancer Risk;
D O I
10.1117/12.2042920
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
Breast density has been identified to be a risk factor of developing breast cancer and an indicator of lesion diagnostic obstruction due to masking effect. Volumetric density measurement evaluates fibro-glandular volume, breast volume, and breast volume density measures that have potential advantages over area density measurement in risk assessment. One class of volume density computing methods is based on the finding of the relative fibro-glandular tissue attenuation with regards to the reference fat tissue, and the estimation of the effective x-ray tissue attenuation differences between the fibro-glandular and fat tissue is key to volumetric breast density computing. We have modeled the effective attenuation difference as a function of actual x-ray skin entrance spectrum, breast thickness, fibro-glandular tissue thickness distribution, and detector efficiency. Compared to other approaches, our method has threefold advantages: (1) avoids the system calibration-based creation of effective attenuation differences which may introduce tedious calibrations for each imaging system and may not reflect the spectrum change and scatter induced overestimation or underestimation of breast density; (2) obtains the system specific separate and differential attenuation values of fibro-glandular and fat for each mammographic image; and (3) further reduces the impact of breast thickness accuracy to volumetric breast density. A quantitative breast volume phantom with a set of equivalent fibro-glandular thicknesses has been used to evaluate the volume breast density measurement with the proposed method. The experimental results have shown that the method has significantly improved the accuracy of estimating breast density.
引用
收藏
页数:7
相关论文
共 50 条
  • [31] Assessment of reproducibility of volumetric breast density measurement using dual energy digital breast tomosynthesis
    Huang, Hailiang
    Duan, Xiaoyu
    Zhao, Wei
    MEDICAL IMAGING 2021: PHYSICS OF MEDICAL IMAGING, 2021, 11595
  • [32] Comparison of breast density measured by dual energy X-ray absorptiometry with mammographic density among adult women in Hawaii
    Maskarinec, Gertraud
    Morimoto, Yukiko
    Daida, Yihe
    Laidevant, Aurelie
    Malkov, Serghei
    Shepherd, John A.
    Novotny, Rachel
    CANCER EPIDEMIOLOGY, 2011, 35 (02) : 188 - 193
  • [33] Accuracy of fully automated, quantitative, volumetric measurement of the amount of fibroglandular breast tissue using MRI: correlation with anthropomorphic breast phantoms
    Wengert, Georg J.
    Pinker, Katja
    Helbich, Thomas H.
    Vogl, Wolf-Dieter
    Spijker, Sylvia M.
    Bickel, Hubert
    Polanec, Stephan H.
    Baltzer, Pascal A.
    NMR IN BIOMEDICINE, 2017, 30 (06)
  • [34] The advent of anthropomorphic three-dimensional breast phantoms for X-ray imaging
    Bliznakova, Kristina
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2020, 79 : 145 - 161
  • [35] Quantitative characterization of ex vivo breast tissue via x-ray phase-contrast tomography
    Li, Xinbin
    Chen, Zhiqiang
    Zhang, Li
    Zhu, Xiaohua
    Wang, Shengping
    Peng, Weijun
    JOURNAL OF X-RAY SCIENCE AND TECHNOLOGY, 2019, 27 (03) : 503 - 516
  • [36] A Comparative Study of Volumetric and Area-Based Breast Density Estimation in Digital Mammography: Results from a Screening Population
    Kontos, Despina
    Bakic, Predrag R.
    Acciavatti, Raymond J.
    Conant, Emily F.
    Maidment, Andrew D. A.
    DIGITAL MAMMOGRAPHY, 2010, 6136 : 378 - 385
  • [37] Volumetric breast composition analysis: reproducibility of breast percent density and fibroglandular tissue volume measurements in serial mammograms
    Engelken, Florian
    Singh, Jasmin-Maya
    Fallenberg, Eva-Maria
    Bick, Ulrich
    Boettcher, Joachim
    Renz, Diane Miriam
    ACTA RADIOLOGICA, 2014, 55 (01) : 32 - 38
  • [38] Quantitative breast tissue characterization using grating-based x-ray phase-contrast imaging
    Willner, M.
    Herzen, J.
    Grandl, S.
    Auweter, S.
    Mayr, D.
    Hipp, A.
    Chabior, M.
    Sarapata, A.
    Achterhold, K.
    Zanette, I.
    Weitkamp, T.
    Sztrokay, A.
    Hellerhoff, K.
    Reiser, M.
    Pfeiffer, F.
    PHYSICS IN MEDICINE AND BIOLOGY, 2014, 59 (07) : 1557 - 1571
  • [39] Maximum-likelihood estimation of scatter components algorithm for x-ray coherent scatter computed tomography of the breast
    Ghammraoui, Bahaa
    Badal, Andreu
    Popescu, Lucretiu M.
    PHYSICS IN MEDICINE AND BIOLOGY, 2016, 61 (08) : 3164 - 3179
  • [40] Recent advances in X-ray imaging of breast tissue: From two- to three-dimensional imaging
    Heck, L.
    Herzen, J.