Development and Clinical Implementation of a Universal Bolus to Maintain Spot Size During Delivery of Base of Skull Pencil Beam Scanning Proton Therapy

被引:35
作者
Both, Stefan [1 ]
Shen, Jiajian [1 ,2 ]
Kirk, Maura [1 ]
Lin, Liyong [1 ]
Tang, Shikui [1 ]
Alonso-Basanta, Michelle [1 ]
Lustig, Robert [1 ]
Lin, Haibo [1 ]
Deville, Curtiland [1 ]
Hill-Kayser, Christine [1 ]
Tochner, Zelig [1 ]
McDonough, James [1 ]
机构
[1] Univ Penn, Dept Radiat Oncol, Philadelphia, PA 19104 USA
[2] Mayo Clin, Dept Radiat Oncol, Phoenix, AZ USA
来源
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS | 2014年 / 90卷 / 01期
关键词
D O I
10.1016/j.ijrobp.2014.05.005
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Purpose: To report on a universal bolus (UB) designed to replace the range shifter (RS); the UB allows the treatment of shallow tumors while keeping the pencil beam scanning (PBS) spot size small. Methods and Materials: Ten patients with brain cancers treated from 2010 to 2011 were planned using the PBS technique with bolus and the RS. In-air spot sizes of the pencil beam were measured and compared for 4 conditions (open field, with RS, and with UB at 2- and 8-cm air gap) in isocentric geometry. The UB was applied in our clinic to treat brain tumors, and the plans with UB were compared with the plans with RS. Results: A UB of 5.5 cm water equivalent thickness was found to meet the needs of the majority of patients. By using the UB, the PBS spot sizes are similar with the open beam (P>.1). The heterogeneity index was found to be approximately 10% lower for the UB plans than for the RS plans. The coverage for plans with UB is more conformal than for plans with RS; the largest increase in sparing is usually for peripheral organs at risk. Conclusions: The integrity of the physical properties of the PBS beam can be maintained using a UB that allows for highly conformal PBS treatment design, even in a simple geometry of the fixed beam line when noncoplanar beams are used. (C) 2014 Published by Elsevier Inc.
引用
收藏
页码:79 / 84
页数:6
相关论文
共 11 条
[11]   Measurements of neutron dose equivalent for a proton therapy center using uniform scanning proton beams [J].
Zheng, Yuanshui ;
Liu, Yaxi ;
Zeidan, Omar ;
Schreuder, Andries Niek ;
Keole, Sameer .
MEDICAL PHYSICS, 2012, 39 (06) :3484-3492